Pr ocess Calibrator T echChek™ 830-KP.
1 Y our New T echChek 830-KP Lighten your load … take the T echChek 830-KP to every site. It ’ s like bringing a cartload of test equipment from the shop to the control room or the fi eld.
3 GENERAL TURN-ON Each time you turn on the T echChek 830-KP it runs through a self check then returns to the most recently selected settings. SOURCE - The three QUIK-CHEK outputs will be the same as previously stored. Each time a different function is selected, the associated three QUIK-CHEK outputs will be recalled.
5 GENERAL CONFIGURING TEMPERA TURE SCALES The thermocouple and RTD ranges may be con fi gured for full time use of ° C, full time use of ° F or selectable ° C and ° F operation. This con fi guration is part of the DEF AUL T SETTINGS below . AUTO-OFF T echChek 830-KP can be set up to turn itself off after 30 minutes of inactivity .
GENERAL SELECTING FUNCTIONS Press a function key to choose among mA, V , T/C, Ω , R TD, FREQ and PRESSURE. SELECTING RANGES Press the RANGE/TYPE pushbutton to select the desired range and scale.
9 CALIBRA TE MILLIAMP INPUTS mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow) Choose this function to provide an output from 0.00 to 24.00 milliamps. The compliance voltage is a nominal 24 VDC to provide the driving power to your milliamp receivers. 1) Disconnect one or both input wires from the device to be calibrated.
11 CALIBRA TE VOL T AGE INPUTS V, m V Choose this function to provide an output from 0.00 mV to 1 10.00 mV and from 0.00 to 10.25 VDC. Current compliance up to 20 mA to provide the driving power to your voltage receivers.
13 CONTROLLER (TYPICAL) ADDITIONAL 4-20 mA OR 1-5V INSTRUMENTS RECORDER COMPUTER LOGGER ETC, 1-5V DC 250 Ω Typical 2-Wire T ransmitter +IN- REF +OUT- CHECK 1-5 VOL T INPUTS WITHOUT DISCONNECTING WIR.
15 CALIBRA TE RESIST ANCE INPUTS Choose this function to simulate a resistance into a variety of instruments. 1) Disconnect one or both input wires from the device to be calibrated 2) Press the T/C/RTD/ Ω pushbutton until the Ω symbol appears on the display .
17 OPERA TING INSTRUCTIONS Instrument with R TD Input Controller T emperature Transmitter T emperature Indicator T emperature T rip or Alarm Receiver Input Ohms or 2-Wire R TD Controller T ransmitter .
19 READ FUNCTIONS Select read by pressing the DISPLA Y/SOURCE/READ pushbutton until the word READ appears on the LCD display . The READ functions measure the desired signal. Multiple scales are available for some functions. MIN/MAX T o read the Maximum or Minimum INPUT since READ mode was entered, simply press the MAX or MIN push-button.
21 READ DC and AC VOL T AGE OUTPUTS mV , V , Vhi and V ac Choose this function to measure from 0.00 to 1 10.00 millivolts (mV), 0.00 to 10.25 DC V olts (V). Use the high voltage connection to read from 0.0 to 200.0 VDC (Vhi). See the next page for V olts AC.
23 22 MEASURE THERMOCOUPLE SENSORS Choose this function to read a thermocouple. The input of the T echChek 830-KP is automatically cold junction compensated. 1) Disconnect the thermocouple from any instrument. 2) Press the T/C/RTD/ Ω pushbutton until the word TYPE & any T/C appear on the display .
25 CHECK CONTINUITY Choose this function to check continuity . A tone will sound and a sound symbol will appear on the display when the resistance between the leads is less than 100 Ohms. 1) Plug the leads into the T echChek 830-KP as shown below . 2) Press the T/C/RTD/ Ω pushbutton until the Ω symbol appears on the display .
27 BLA CK BLA CK RED 3 2 1 R TD Element OPERA TING INSTRUCTIONS MEASURE RTD SENSORS 26 MEASURE RTD SENSORS Choose this function to read an RTD. Three wires must be use for both 2 and three wire RTDs. 1) Disconnect the RTD from the instrument being calibrated.
29 Frequency Output Signal Flowmeter Flow Sensor V ariable Speed Drive Controller T ransmitter Frequency Output Signal Flow Meter Line Frequency V ariable Speed Drives Connection for signals with amplitudes below 10.
31 +IN- REF T ypical 2-Wire T ransmitter OPERA TING INSTRUCTIONS CALIBRA TE 2-WIRE TRANSMITTERS (T/C / mA Shown) CALIBRA TE 4-WIRE TRANSMITTERS +IN- REF T ypical 4-Wire T ransmitter Loop Power Supply .
33 Pressure Source PROCESS CALIBRAT OR QUIK-CHEK READ PRESSURE Choose this function to measure pressure with a ModPak 91 module holder and one of the many QuikCal 90 pressure modules.
35 PROCESS CALIBRATOR QUIK-CHEK 34 OPERA TING INSTRUCTIONS CALIBRA TE PRESSURE TRANSMITTERS READ PRESSURE AND P-XMTR mA, P-XMTR %, P-XMTR DP% Choose this function to simultaneously supply power to a 2-Wire pressure transmitter while supplying the input signal to the transmitter and displaying the 4-20 mA output of the transmitter.
37 SPECIFICA TIONS POWER & MEASURE 2-WIRE TRANSMITTERS RANGES & ACCURACY : Same as for MILLIAMP SOURCE OUTPUT CURRENT : up to 24.00 mA TYPICAL DRIVE CAP ABILITY : 1200 Ohms @ 20.00 mA COMPLIANCE VOL T AGE: nominal 25 VDC @ 20 mA COMMON MODE ERROR: 0.
39 SPECIFICA TIONS SOURCE RTD & OHMS RTD TYPES: Pt 100 Ω for 1.3850 (DIN/IEC 751 & New JIS), 1.3902 (Burns), 1.3926 (US Lab) & 1.3916 (Old JIS 1604C-1981) Ni 120 Ω , Cu 10 Ω & Cu 50 Ω RTD RESOLUTION: 1 ° C or 1 ° F RANGE OHMS: 0.
41 THERMOCOUPLE SPECIFICA TIONS T/C TYPE N R S ° C ° F RANGE ACCURACY RANGE ACCURACY 300 to 1300 ± 2.1 572 to 2372 ± 3.0 100 to 299 ±2.3 212 to 571 ± 3.3 -50 to 99 ± 2.6 -58 to 21 1 ±4.0 -200 to -49 ± 5.0 -328 to -57 ± 8.2 1750 to 1768 ± 4.
43 QuikCal 90 MODULE SPECIFICA TIONS OPERA TING TEMPERA TURE: -10 ° C TO 50 ° C (13 ° F TO 122 ° F) STORAGE TEMPERA TURE: -40 ° C TO 85 ° C (-40 ° F TO 185 ° F) WEIGHT : 0.
44 W ARRANTY Altek products are warranted to be free from defects in material and workmanship (excluding fuses, batteries and leads) for a period of three years from the date of shipment. W arranty repairs can be obtained by returning the equipment prepaid to our factory .
PN 1800470 Rev A April 2002 ©2002 Altek Industries, Inc. Speci fi cations subject to change without notice. All rights reserved. Printed in U.S.A..
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Altek 830-KP è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Altek 830-KP - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Altek 830-KP imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Altek 830-KP ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Altek 830-KP, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Altek 830-KP.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Altek 830-KP. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Altek 830-KP insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.