Manuale d’uso / di manutenzione del prodotto 201A del fabbricante Teledyne
Vai alla pagina of 31
MANUAL ADDENDUM MODEL 201A AMMONIA ANALYZER TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (T-API) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 TOLL-FREE: 800-324-5190 FAX: 858-657-9816 TEL: 858-657-9800 E-MAIL: api-sales@teledyne.com WEB SITE: www.te le dy ne -ap i.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 SAFETY MESSAGES Your safety and the safety of others is very im portant. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 TABLE OF CONTENTS SAFETY MESSAGES ....................................................................................................... 1 TABLE OF CONTENTS ...................
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 LIST OF FIGURES F IGURE 1 - M201A A MMONIA A NALYZER – B LOCK D IAGRAM ................................................................. 21 F IGURE 2 - M201A E XTERNAL P NEUMATIC C ONNECTION D IAGRAM WITH ECO (E XTERNAL C ALIBRATOR O PTION ) .
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 LIST OF TABLES T ABLE 2.1 - Z ERO C ALIBRATION P ROCEDURE - Z ERO G AS T HROUGH THE S AMPLE P ORT ......................... 11 T ABLE 2.2 - E NTER THE E XPECTED S PAN G AS C ONCENTRATIONS P ROCEDURE .
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 5.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 1.0 M201A AMMONIA ANALYZER This manual is to be used in conjunction with the M200A NOx Analyzer manual p/n 02246 supplied with this instrument.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 converts some of the ammonia to NO, t hus the TN channel data contains a small ammonia signal. Secondly, the ammonia conv erter does not convert all of the ammonia or ambient NO2 to NO.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Do NOT use copper tubing or fittings designed for household plumbing. Use ONLY Chromatography grade (cleaned, passivated) stainless steel tubing Use ONLY Glass tubing for sample inlet manifold.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 1.2 M201A Analyzer Specifications Ranges 0-50 to 0-2000 ppb in 1 ppb increments Range Modes Single, Independent, AutoRange Noise at Zero 1 0.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 2.0 CALIBRATION PROCEDURE Calibration of the M201A is done by first calibrating the TN and TNx channels with NO span gas diluted with zero air. In the second phase of calibration, ammonia span gas is introduced to measure the overall system e fficiency for NH3 conversion.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 2.1 - Zero Calibration Procedure - Zero Gas Through the Sample Port Step No. Action Comment 1. Press CAL The M201A enters the calibrate mode from sample mode. Input zero gas through the sample port.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 2.3 - Span Calibration Procedure - Span Gas thru Sample Port Step No. Action Comment 1. Press CAL The M201A enters the calibrate mode. Input NO span gas through the sample port.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 2.1.1 Zero/Span Calibration with Z/S Valves or ECO Option The zero gas enters through the zero gas inle t port on the rear panel. Make sure that the umbilical cord that makes the cont rol connections between the analyzer and the converter is connected (see Figure 2 below).
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 2.5 - Enter the Expected Span Gas Concentrations Procedure Step No. Action Comment 1. Press CAL- CONC- TNX This key sequence causes the M201A to prompt for the expected TNX concentration.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 2.6 - CALS - Span Cal Procedure - Gas from rear panel span gas inlet Step No. Action Comment 1. Press CALS The M201A enters the calibrate mode. The Z/S valves are activated to route span gas into the analyzer.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 2.1.3 NH3 Calibration Since the system efficiency is less than 100%, it is necessary to determine a conversion efficiency factor for ammonia and enter it into the system. 1. Set the converter efficiency to 1.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 The instrument is now fully calibrated. 2.1.4 Channel Calibration Quality After the ammonia channel is calibrated via the conv.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 3.0 MAINTENANCE SCHEDULE 3.1 M201A Maintenance Schedule The maintenance requirements of the M201A are essentially like a standard M200A NOx analyzer. Please refer to Table 8-1 in the M200A Operator Manual for the M200A maintenance schedule.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Table 3.2 – M501NH Converter Rebuild Parts List Part No. Description KIT000191 Converter Rebuild Kit KIT000139 Ceramic Bushings HE0000007 Ceramic Heater 220W 60 VAC Procedure: 1.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 3.3 AutoZero Flow Check Since the M201A is an ammonia analyzer, the flow through the AutoZero orifice is especially important. Check the AutoZero flow as follows: 1. This procedure should be performed with the sample pump running.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 1 - M201A Ammonia Analyzer – Block Diagram 21.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 MODULE AIR ZERO FROM SPAN 02522 UMBILICAL CORD TO VACCUM PUMP POWER UMBILICAL TO CONV CONV FROM UMBILICAL SS TUBING GAS INLET .
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 3 - M201A Pneumatic Diagram Basic 23.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 4 - M201A Pneumatic Diagram Z/S Option 24.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 5 - M201A Pneumatic Diagram External Calibrator Option (ECO) 25.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 Figure 6 – M501NH Converter Cartridge 26.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 27.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 4.0 M201A SPARE PARTS LIST 000940400 Orifice, 4 mil, 80 cc, Rx Cell 000940600 Orifice, 10 mil, 500 cc, Rx Cell 002270100 Gasket (Rx Cell) Qty.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 M201A Spare Parts List (cont.) 016810200 O3 Generator Assembly for M200A/200AH Low Output 017230000 Dryer Material, Drierite 018720000 Welded Moly Conv.
Teledyne API Model 201A NH 3 Analy zer Instruction Manual, 04689, Rev. B1 M201A Spare Parts List (cont.) KIT000129 Welded Moly, Cartridge only KIT000191 Rebuild, NH3 INNER TUBE & CATALYST OR000000.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Teledyne 201A è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Teledyne 201A - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Teledyne 201A imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Teledyne 201A ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Teledyne 201A, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Teledyne 201A.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Teledyne 201A. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Teledyne 201A insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.