Teledyne Analytical Instruments QUICKS T AR T GUIDE M O D E L 3 0 0 0 T A T ra ce Oxygen Analyzer.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 3 GETTING S T ARTED This Quickstart Guide is designed to get you s et up and ope rating your Teledyne An a lytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 4 • Left & Right Arrows Select between functi ons currently displayed on the VF D screen. • Up & Down Arrows Increment or decrement values of functions currently displayed.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 5 • Analog Output C onnecti on: Terminals for current and volt age a n alog output signals. • Digital Input Terminals: Termi n als for connecti ng d igital input signals. • Alarm Connections: 2 concentration alarms and 1 system alarm.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 6 SETUP AND INS T ALL Mount the Ana lyzer The Model 3000T A is for indoor use in a general purpose are a. It is NOT for hazardous environme nts of any type. It is design ed fo r f lush p anel m ount ing.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 7 4. If using remote calibra tion feature: • Connect zero gas to t he Z ERO IN port. • Connect span ga s to the SP AN IN port. 5. Otherwise, c a li b ra t io n ga se s m u st b e T e e' d i nt o t h e s a m p l e i n le t wi t h a p p r o p ri a t e v a l ve s .
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 8 • Connect primary input power: Insert the power cord into the power cord rece ptacle. • Install the fuse: Fuses are not installed at the fac tory. Be sure to install the proper fuse as par t of installati on.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 9 • Connect Range ID Contacts: There a re four ded icated Range ID relay contacts on the rear panel. The first three ranges are assigned to relays in ascending order. Range 1 ID— Low range Range 2 ID— Medium ra nge Range 3 ID— High range Range 4 ID—Air Cal Ra nge (25%).
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 10 9. With t h e O-ring in place, align the guide pin with the hole on the cell holder.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 11 BASIC OPER A TION Once the a naly zer has been installed, it can be configured for your application. Prior to using the instrument you m ust : • Calibrate t he instrument • Define the analysis r anges and choose whether to use autoranging or a fixed range.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 12 A vailab le Functions The six function butt ons are used to control the analyzer. Some functi ons have submenus that allow additional choices as s hown below. Any function ca n be sel ected at any time by pressing the appropriate button (unless password restrictions appl y).
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 13 Password Protection If a password is assigned, then setting the following system parameters can be done only after t he passw ord is en.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 14 TRAK/HLD Auto — Cal PSWD Logou t M ore • Use the ◄► arrow keys to scroll the blin king over to PSWD , • Press Enter to select the password function.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 15 Auto Mode Zero Calibration • Press Zero to enter the zero function mode. • Use the ▲▼ arrow keys to toggle between AUT O and MAN zero settling. Stop when AUT O appears blinking. Zero : Settling: AUTO <ENT> To B e gin • Press Enter to begin zeroing.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 16 Span Calibration Make sure an appropriate span gas is connecte d to the sample port. Auto Mode Spanning: • Press Span to enter the span function. • Use the arrow keys to toggle between AUTO and MAN span settling.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 17 Calib. Holding time Cal hold: 5 min • Press Enter key to continue. Span Val : 000008.00 <ENT>Span <UP>Mod # • Press ▲ (<UP>) to permit modification ( Mod # ) of span value.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 18 • Latc hing? Ltch–Y/N (Yes/No). • To change any of th e alarm parameters, use the ◄► arrow keys to move the blinkin g over to the specific par ameter. • Use the ▲▼ arrow keys to change t he va lue of the parameter.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 19 • Use the ▲▼ arrow keys to switch from AUTO t o FX/L O , FX/MED , or F X /HI to set the instrument on the desired fixe d range (low, medium , or h igh).
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Teledyne 3000TA è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Teledyne 3000TA - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Teledyne 3000TA imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Teledyne 3000TA ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Teledyne 3000TA, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Teledyne 3000TA.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Teledyne 3000TA. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Teledyne 3000TA insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.