Manuale d’uso / di manutenzione del prodotto 200AH del fabbricante Teledyne
Vai alla pagina of 196
INSTRUCTION MANUAL MODEL 200AH NITROGEN OXIDES ANALYZER ©TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 TOLL-FREE: 800-324-5190 FAX: 858-657-9816 TEL: 858-657-9800 E-MAIL: api-sales@teledyne.
THIS PAGE IS INTENTIONALLY LEFT BLANK.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 i SAFETY MESSAGES Your safety and the safety of others is very impor tant. We have provided many important safety messages in this manual. Please read these messages carefully. A safety message alerts you to potential hazards that could hurt you or others.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 ii THIS PAGE IS INTENTIONALLY LEFT BLANK.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 iii Table of Contents SAFETY M ESSAGES ........................................................................................... I LIST OF T ABLES ............................
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 iv 5.3.4. Range Menu ........................................................................................................ 5-20 5.3.5. Passwor d Enable ......................
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 v 7.12.1. Calibrati on Gasses ............................................................................................ 7-14 7.12.2. Calibrati on Frequency ..................
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 vi APPENDIX A ELECTRICAL SCHEMATICS .........................................
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 vii LIST OF FIGURES F IGURE 2-1: R EMOVAL OF S HIPPING S CREWS & C HECK FOR C ORRECT P OWER ............................ 2-3 F IGURE 2-2: R EAR P ANEL ......................
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 viii LIST OF TABLES T ABLE 2-1: F INAL T EST AND C ALIBRATION V ALUES ................................................................... 2-15 T ABLE 4-1: S AMPLING M ODES OF THE M200AH .
Teledyne API Model 200AH NO X Analy zer Instruction Manual, 06492, Rev. G2 ix T ABLE 9-4: D IAGNOSTIC M ODE - S IGNAL I/O ............................................................................ 9-12 T ABLE 9-5: M ODEL 200AH V ARIABLES ...........
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 x THIS PAGE IS INTENTIONALLY LEFT BLANK.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 1-1 1. HOW TO USE THIS MANUAL The Model 200AH has been designed to produce accurate data, be serviceable, reliable and easy to use. The M200AH's microprocessor continually checks operating param eters such as temperature, flow, and critical voltages.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 1-2 THIS PAGE IS INTENTIONALLY LEFT BLANK.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-1 2. GETTING STARTED 2.1. Unpacking CAUTION Your safety and the safety of others is very important. We have provided many important safety messages in this manual. Please read these messages carefully.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-2 WARNING Analyzer Exhaust – O 3 Scrubber – Pump Pack Danger – Analyzer exhaust contains ozone. Do not defeat the internal zone scrubber. This device must always be present between the analyzer reaction cell and pump.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-3 Figure 2-1: Removal of Shipping Screws & Check for Correct Pow er.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-4 Figure 2-2: Rear Panel.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-5 Figure 2-3: Rear Panel with O 2 Option.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-6 Figure 2-4: Inlet and Exhaust Venting Recommendations.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-7 2.3. Initial Operation 1. Turn on the instrument power. 2. The display should immediately light, showing the com puter’s memory configuration, then the instrument type - M200AH.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-8 7. When the instrument is warm ed up, re-check the TEST functions against Table 2-1. All of the readings should compare closely with those in the table. If they do not see Section 9.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-9 Step 2 - Calibrate the instrument for NO X : Zero/Span Calibration Procedure Step For Units w/o O 2 Option For Units with O 2 Option Comment 1. Input NO X Zero gas Input NO X Zero gas Allow Zero gas to enter the sample port on the rear of the instrument.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-10 Zero/Span Calibration Procedure (Continued) Step For Units w/o O 2 Option For Units with O 2 Option Comment 13. Wait 10 min Wait for reading to stabilize at the zero value.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-11 Step 3 - Review Quality of calibration: Calibration Quality Check Procedure Step For Units w/o O 2 Option For Units with O 2 Option Comment 1. Scroll the TEST function menu until the NO X SLOPE is displayed.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-12 Figure 2-5: Front Panel.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-13 Figure 2-6: Assembly Lay out.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-14 Figure 2-7: Assembly Lay out – O 2 Sensor Option.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-15 Table 2-1: Final Test and Calibration Values TEST Values Observed Value Units Nominal Range Reference Section RANGE ppm 5-5000 5.3.4 O2 RANGE % 5-100 5.3.4 NOX STB ppm 0.0 - 0.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 2-16 Table 2-1: Final Test and Calibration Values (Continued) Parameter Observed Value Units Nominal Range Reference Section NO Span Conc ppm 0.5 - 5000 Table 7-3 NO x Span Conc ppm 0.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 3-1 3. SPECIFICATIONS, WARRANTY 3.1. Specifications NOX Specifications O 2 Option Operating Modes NO/NO x switching m ode, NO only mode, NO x only m ode.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 3-2 3.2. Warranty WARRANTY POLICY (02024c) Prior to shipment, Teledyne API equipment is thoroughly inspected and tested. Should equipment failure occur, Teledyne API assures its customers that prom pt service and support will be available.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-1 4. THE M200AH NOX ANALYZER 4.1. Principle of Operation 4.1.1. NO X Measurement The Teledyne API Model 200AH Analyzer is de.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-2 concentration is measured and stored internally. The valve is then switched, routing the sample gas through the converter, the signal measured is the NO X concentration which is also stored in the computer.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-3 Figure 4-1: Paramagnetic Sensor Refer to Figure 4-1. The measuring system is “null-balanced”. First, the ‘zero’ position of the suspension assembly, as measured in nitrogen.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-4 Figure 4-2: Block Diagram.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-5 4.2. Operation Summary 4.2.1. NO X Sensor Module, Reaction Cell, Detector The sensor module is where light from the chem ilumenescent reaction is generated and detected. It is the most complicated and critical sub-assem bly in the entire analyzer.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-6 4.2.3. Pneumatic Sensor Board The sensor board consists of 2 pressure sensors and a flow sensor. One pressure sensor measures the pressure in the reaction cell. The reaction cell is maintained at about one-quarter of atmospheric pressure.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-7 4.2.5. V/F Board The computer CPU board comm unicates via 2 major hardware assemblies. These are the V/F board and the front panel display/keyboard. The V/F board communicates with the CPU via the STD-100 bus.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-8 Table 4-2: Front Panel Status LED's LED State Meaning Green Off On Monitoring Blinking NOT monitoring, DAS disabled M.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-9 An external pump comes as standard equipm ent. The Pump Pack includes a vacuum pum p and ozone scrubber. It is supplied with 0.25" tube fitting to connect to the exhaust fitting on the M200AH rear panel.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-10 Figure 4-3: External Pump Pack.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-11 4.2.9. Ozone Generator Because of the instability of ozone, it is necessary to generate this gas inside the analyzer. The ozone generation module consists of a switching power supply that drives a high voltage transformer and silent discharge tube.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 4-12 THIS PAGE IS INTENTIONALLY LEFT BLANK.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-1 5. SOFTWARE FEATURES The M200AH control software has two major operating modes. The SAMPLE m ode is the normal mode when the instrum ent is taking data. The software menu that covers the SAMPLE mode is diagramm ed in Figure 5-1.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-2 Figure 5-1: Sample Menu Tree.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-3 Figure 5-2: Setup Menu Tree.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-4 Figure 5-3: Setup Menu Tree- Units with O 2 Sensor.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-5 Figure 5-4: Setup Menu Tree – Units with O 2 Sensor.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-6 5.1.1. Sample Menu Table 5-1: M200AH Sample Menu Structure Level 1 Level 2 Level 3 Level 4 Level 5 Description Reference Section TEST TST> Test functions 5.2.1, Table 9-1 CAL Zero/Span calibration w/ gas through sam ple port 5.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-7 5.1.2. Set-Up Menu Table 5-2: M200AH Setup Menu Level 1 Level 2 Level 3 Level 4 Description Reference Section CFG C FG is primari ly used for showing special configuration opti ons and factory special software.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-8 Table 5-2: M200AH Setup Menu (Continued) Level 1 Level 2 Level 3 Level 4 Description Reference Section ZERO-LO- HI Do a zero, low, and high span calibrate 5.3.2, 6.2 LO Do a low span calibrate 5.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-9 Table 5-3: M200AH Setup Menu Level 1 Level 2 Level 3 Level 4 Level 5 Description Reference Section DAS Data Acquisition menu 5.3.3 VIEW Select which DAS data collector to view 5.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-10 Table 5-4: M200AH Setup Menu Level 1 Level 2 Level 3 Level 4 Level 5 Description Reference Section PASS Password protection 5.3.5 ON/OFF Enable password protection of calibration and setup m enus.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-11 5.2. Sample Mode 5.2.1. Test Functions NOTE In any of the following TEST functions, if a value of XXXX is displayed, that indicates an off scale and therefore meaningless reading.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-12 Signal Stability (STABIL, NOX STB) The instrument noise is computed by calculating th e standard deviation of the last 10 m inutes of concentration data. If NO/ NO X switching mode is selected, the noise reading is that of the NO X channel.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-13 DC Power Supply Voltage (DCPS) The DCPS voltage is a composite of the 5 and 15 VDC voltages in the Power Supply Module. This is meant to be a quick indicator to show if the PSM is working correctly.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-14 If the analyzer is equipped with the molybdenum converter option the TEST function will report MOLY TEMP, the temperature setpoint will be 315 10°C.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-15 5.2.2. CAL, CALS, CALZ, Calibration Functions The calibration and zero/span checking of the M200AH analyzer is treated in detail in Section 7, Table 7-1 summarizes types of calibration.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-16 5.2.2.5. NO, NO X and O 2 Cal Concentration Before the M200AH can be spanned, it is necessary to enter the expected span concentrations for NO and NO X . In single range mode there is one value each for the NO, NO X and O 2 expected span values.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-17 Because the span calibration calculation includes zero calibration data and zero calibration includes span calibration data, the zero and span calibrations can be performed in any order and the instrument will be accurately calibrated either way.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-18 5.3.3. Data Acquisition System (DAS) The M200A contains a flexible and powerful built in data acquisition system (DAS) that enables the analyzer to store concentration data as well as diagnostic parameters in its battery backed memory.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-19 The function of each of the default Data Channels is described below: CONC: Samples NO X , NO and NO 2 (NO X , NO and O 2.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-20 5.3.4. Range Menu For NO X measurements, the instrum ent operates on any full scale range from 5 to 5,000 ppm . The range is the concentration value that equals the maximum voltage or current output on the rear panel of the instrument.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-21 To set up Auto Range press SETUP-RNGE-MODE-AUTO, then press ENTR. To set the values press SETUP-RNGE-SET. The M200AH will prompt you for LO, then HI which is the lower and upper ranges of Auto Range.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-22 3. SELECT RANGE The range selection is the same with dilution as with normal m onitoring.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-23 5.3.4.7. Recorder Offset If necessary, a recorder offset can be introduced into the analog output voltages. This can be used to bias the output voltage or current 10%. It is intended for older design chart recorders that cannot show slightly negative readings.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-24 5.3.8. Communications Menu The COMM menu allows the RS-232 BAUD rate to be set. To set the BAUD rate press SETUP- MORE-COMM-BAUD, select the appropriate BAUD rate, then press ENTR.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-25 5.4. M200AH Operating Modes The M200AH has two main operating modes. The instrum ent is normally in SAMPLE mode when it is taking data or doing a calibration. The second main mode is the SETUP m ode.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-26 This is the most comm on mode of operation. It provides continuous output of NO, NO X and NO 2 gas concentrations. To select switching m ode press SETUP-MORE-VARS-ENTR and press NEXT to scroll to the MEASURE-MODE variable.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-27 5.5. 4-20 mA Current Loop The analog signal output is available on the rear panel as a voltage or 4-20 mA current output. Optionally the current loop outputs can be isolate d.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-28 5.7. RS-232 Interface The RS-232 communications protocol allows the instrum ent to be connected to a wide variety of computer based equipment. The interface provides two basic functions in the M200AH.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-29 Table 5-9: RS-232 Port Setup - Front Panel Decimal Value Description 1 Turns on quiet mode (m essages suppressed) 2 Place.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-30 940331 is the default password. The password can be changed to any number from 0 to 999999 by the variable RS232_PASS. To change the password enter the command V RS232_PASS=940331 which sets the password to the factory default.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-31 Table 5-11: RS-232 Terminal Mode Editing Keys Key Function CR (carriage return) Execute command BS (backspace) Backspace .
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-32 Table 5-12: RS-232 Command Summary Commands Definition ? Print help screen T SET ALL Enable display of all test variables.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-33 Table 5-13: RS-232 Command Summary Terminal Mode Editing Keys Definition V MODE Print current analyzer mode BS Backspace .
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-34 Table 5-14: RS-232 Interface Command Types First Character Message Type C Calibration status D Diagnostic R DAS report T Test measurement V Variable W Warning There are 6 different types of messages output by the M200AH.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-35 5.7.3. TEST Commands and Messages Table 5-15: RS-232 Test Messages Name Message Description NOXRANGE 1 NOX RNG=xxxxx ppm Indep. range for NO x channel NORANGE (1) NO RNG=xxxxx ppm (3) Indep.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-36 The T command lists TEST messages. Examples of the T com mand are: T LIST Lists test messages currently enabled with T SE.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-37 Whenever a warning message is reported on the analyzer display, if the RS-232 interface is in the normal mode(i.e. not in quiet m ode) the warning message is also sent to the RS-232 interface.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-38 Whenever the analyzer starts or finishes a calibration, it issues a status report to the RS-232 interface. If the RS-232 interface is in the norma l mode, these reports will be sent.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-39 The following is a summary of the Diagnostic com mands. Table 5-19: RS-232 Diagnostic Command Summary Command Description D LIST Prints all I/O signal values. See Table 9-4 for Sig I/O definitions.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-40 To print records from a DAS data channel enter: D REPORT “name” RECORDS=nnn COMPACT|VERBOSE Examples of reports are: .
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-41 All of the default Data Channels sample more than one param eter, for these channels, each parameter is printed on a separate line.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 5-42 Where: name = name of the variable value = current value of variable warning_lo = lower limit warning (displayed if appli.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 6-1 6. OPTIONAL HARDWARE AND SOFTWARE Optional equipment offered with the M200AH includes: 1. Rack mount with slides 2. Rack mount without slides, ears only 3. Rack mount for external pump 4.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 6-2 6.2. Zero/Span Valves Option The Zero/Span Valve option consists of a manifold of stainless steel valves plus drive electronics connected to the CPU. The Option consists of two variations.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 6-3 6.2.1. Autocal - Setup Zero/Span Valves The Zero/Span valves can be set up to operate automatically. Table 6-2 in this section describes Autocal setup. Table 6-2: Setup Automatic Zero/Span Checking or Calibration Step Action Comment 1.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 6-4 Table 6-3: Action of MODE Field in AutoCal MODE CALIBRATE = ON CALIBRATE = OFF ZERO Zero calibrates LO or HI (depending on RANGE-TO-CAL), SINGLE or IND ranges. Zero checks LO or HI (depending on RANGE-TO-CAL), SINGLE or IND range.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 6-5 6.4. Isolated 4-20 mA Current Loop Option The M200AH comes equipped with 4-20mA current loop. The electrical ground of this output is part of the instrument ground. In certain applications the instrument ground m ust be isolated from the external ground.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 6-6 The M501 contains 2 converters, a high temperat ure stainless steel and a molybdenum converter. While this comb ination produces very high conversion efficiencies, the life of the moly must be considered.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-1 7. CALIBRATION AND ZERO/SPAN CHECKS There are several ways to check and adjust the calibration of the M200AH. These different methods are summ arized in the following Table 7-1.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-2 Table 7-1: Types of Zero/Span Check and Calibration Section Type of Cal or Check Description 7.1 Manual Z/S Check or Calibration with span gas in the sample port This calibration method uses calibration gas coming in through the sample port.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-3 7.1. Manual NO X Zero/Span Check or Cal With Zero/Span Gas in the Sample Port The calibration of the instrument can be checked or adjusted using gas coming in the sam ple port.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-4 Enter the expected NO x and NO span gas concentrations: Table 7-3: Enter Expected Span Gas Concentrations Procedure Step Number Action Comment 1. Check “Before Calibration” Check “Before Calibration” Section 7 before proceeding.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-5 Table 7-4: Manual Span Calibration Procedure - Span Gas Thru Sample Port Step Number Action Co mment 1. Press CAL-CONC-NO This key sequence causes the M200AH to prompt for the expected NO concentration.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-6 7.2. Manual O 2 Zero/Span Check or Cal With Zero/Span Gas in the Sample Port The calibration of the O 2 Sensor can be checked or adjusted using gas coming in the sam ple port.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-7 Table 7-6: Manual O 2 Span Calibration Procedure - Span Gas Thru Sample Port Step Number Action Co mment 1. Press CAL-O2-CONC This key sequence causes the M200AH to prompt for the expected O 2 concentration.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-8 Table 7-7: Manual Zero Check Procedure - Z/S Valves Step Number Action Co mment 1. Check “Before Calibration” Check “Before Calibration” Section 7 before proceeding.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-9 Table 7-8: Manual Span Check Procedure - Z/S Valves Step Number Action Co mment 1. Press CALS The M200AH enters the sp an calibrate mode and switches the Z/S valves to admit NO span gas.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-10 7.5. Use of Zero/Span Valves with Remote Contact Closure The Zero/Span valve option can be operated using Remote Contact Closures provided on the rear panel. See Figure 2-2 for connector location and pinout.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-11 7.6. NO Only Mode Calibration For details on the operation of this mode, check Section 5.4. In NO only mode, the span check or span calibration is operated using the same procedures covered in Sections 7.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-12 Table 7-10: Calibration Requirements for AutoRange or Remote Range No matter which calibration method you choose from this chapter, all of the following parameters must be set to properly calibrate the instrum ent on AutoRange or Remote range.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-13 Table 7-11: Calibration Quality Check Step Number Action Comment 1. Scroll the TEST function menu until the NO X SLOPE is displayed. The SLOPE value for NO X should be 1.0 0.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-14 Table 7-12: Converter Efficiency - Automatic Calculation Step Number Action Comment 1. Preparation Make sure a calibra tion is completed before setting up Converter efficiency.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-15 NOTE It is important to note that if you are using blended calibration gasses, other components of the blend can interfere with the NO concentration value. This can affect calibration accuracy and linearity.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 7-16 THIS PAGE IS INTENTIONALLY LEFT BLANK.
Teledyne API Model 200AH NO X Analy zer Instruction Manual, 06492, Rev. G2 8-1 8. MAINTENANCE 8.1. Maintenance Schedule NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-2 8.2. Replacing the Sample Particulate Filter The particulate filter should be inspected often fo r signs of plugging or contamination. It is also common for dirt particles to absorb NO 2 , thus causing the reading to be low.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-3 Figure 8-1: Replacing the Particulate Filter.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-4 8.3. Replacing the Converter The Converter is located in the cen ter of the instrument, refer to Figure 2-6 for its location. The heater, thermocouple, ozone scrubber and converter are designed to be replaced as a single unit.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-5 Figure 8-2: Converter Assembly.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-6 8.4. Cleaning the Reaction Cell The reaction cell should be cleaned whenever troubleshooting points to it as the cause of the trouble. A dirty cell and window will cause excessive noise, unstable span or zero, or low response.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-7 10. Reassemble in the reverse order of disassem bly, being sure to seat the paper gasket in the reaction cell before installing the window. After inserting the window, drop the o-ring in before inserting the sleeve.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-8 Figure 8-4: Sample/Bypass Flow Control Assembly.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-9 8.5. Pneumatic Line Inspection Particulate matter in the pneumatic lines will affect both flow rate and response tim e. It is important that the pneumatic system be periodically inspected and thoroughly cleaned if necessary.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-10 Figure 8-5: Pneumatic Diagram.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-11 Figure 8-6: Pneumatic Diagram with Zero/Span Valves.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-12 Figure 8-7: Pneumatic Diagram with Oxygen Sensor.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-13 8.6. Leak Check Procedure If a leak checker is not available, it is possible to leak check the instrument using the M200AH's pump plus a shut-off valve. 1. Turn off instrument power and pump power.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 8-14 8.8. Prom Replacement Procedure Preparation: If any setup changes such as RANGE, AUTOCAL, ON/OFF etc. have been made, record the changes because all settings should be checked after the PROM is changed.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-1 9. TROUBLESHOOTING AND ADJUSTMENTS NOTE The operations outlined in this chapter are to be performed by qualified maintenance personnel only. This section of the manual contains information on diagnosing and repairing instrum ent performance problems.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-2 4. Pay special attention to the slopes, the slopes are the software equivalent of the span pot on an analog instrument. If the slopes are not 1.0 0.3 the gain has changed, usually from a change in cell pressure (vacuum) - compare to value in Table 2-1.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-3 The Table 9-1 indicates possible fault conditions that could cause the TEST functions to be outside the acceptable range. Table 9-1: Test Functions Test Function Factory Set-Up Comment RANGE 500 PPM This is the Range of the instrum ent.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-4 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment SAMPLE FLW 290 cc/min 30 Thi s is the instrum ent flow. It is com puted using the up stream and down stream pressures across the sam ple and bypass flow orifices.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-5 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment PMT 0-5000 mV This is the instantaneous output of the PMT. During normal operation the value varies widely as the M200AH switches from NO to NO x modes.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-6 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment NORM PMT 0-5000 mV Normalized PM T voltage. Takes raw PMT voltage and appl ies temperature and pressure corrections.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-7 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment O2 TEMP (Option) 50°C 2 Temperature of t he oven that houses the optional O 2 sensor and the sample flow ori fice.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-8 9.1.2. Fault Diagnosis with WARNING Messages The M200AH monitors several internal values for alarm conditions. If the condition for an alarm is met, the alarm is displayed on the front panel and the warning is transm itted out the RS-232 port.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-9 Table 9-2: Front Panel Warning Messages Message Description SYSTEM RESET Analyzer was reset/powered on. This warning occurs every time the instrument is powered up, as in a power failure.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-10 Table 9-2: Front Panel Warning Messages (Continued) Message Description BLOCK TEMP WARNING The sample flow orifi ce is located in a separate temperature controlled block near the rear of the instrument.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-11 Table 9-3: Summary of Diagnostic Modes DIAG Mode Description SIGNAL I/O Gives access to the digital and analog inputs and outputs on the V/F board. The status or value of all of the signals can be seen.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-12 9.1.3.1. Signal I/O Diagnostics Table 9-4: Diagnostic Mode - Signal I/O No. Signal Control Description 0 DSP_BROWNOUT NO Display brownout is used to keep the display from getting corrupted during low line voltage conditions.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-13 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Control Description 11 OPTIC_TEST YES Turns on optic test bit in preamp. Should be used for troubleshooting Preamp logic lines.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-14 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Control Description 22 ST_FLOW_ALARM YES Status Bit - Flow.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-15 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Control Description 33 PMT_TEMP NO PMT cold block temperature. Typically 4300 mV for 7°C. 34 DCPS_VOLTAGE NO DC power supply composite voltage output.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-16 9.1.3.2. Electric Test This function injects a constant voltage between the preamplifier and the buff er amplifier on the preamp board. Electric test checks part of the preamp, the V/F and com puter for proper functioning.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-17 9.1.3.4. Ozone Gen Power This diagnostic manually turns the power off and on to the ozone generator. When the M200AH is powered up from a cold start the ozone generator is not imm ediately started.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-18 Table 9-5: Model 200AH Variables No. Name Units Default Value Value Range Description 0 DAS_HOLDOFF min 15 0.5 - 20 Time, in minutes, that data is not stored in the DAS after calibration or power-up.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-19 9.1.5. Test Channel Analog Output Many of the TEST functions have an analog voltage associated with them. As a diagnostic aid it is possible to route the various test voltages out the 4th analog output port.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-20 Table 9-6: Test Channel Readings (Continued) TEST Channel Minimum* Maximum* Description RCELL TEMP 0 C 70 C Reaction Cell temperature is set to 50 C. At the setpoint, a typical reading is 3600 mV.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-21 PMT Calibration Procedure: NOTE In this procedure a range of 100 ppm and a span gas concentration of 80 ppm is used as an example. Other values can be used. 1. On the Preamp board, set S2 to 6, set S1 to 8.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-22 10. Adjust S2, the HVPS coarse adjustment, on the preamp board to the setting that produces a signal that is closest to 640 mV. Adjust S1, the HVPS fine adjustm ent, to the setting that produces a signal that is closest to 640 mV.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-23 Figure 9-1: Span Calibration Voltage.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-24 9.2. Performance Problems When the response from a span check is outside the control lim its, the cause for the drift should be determined, and corrective action should be taken.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-25 9.2.2. Flow Check 1. Check TEST function RCEL - this is the absolute pressure in the reaction cell. It should be 3 - 10 in-Hg-A. 2. Check that pump is running. Observe pump for proper operation.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-26 2. Check instrument electronics. A. Do ELEC TEST procedure in DIAGNOSTIC menu Section 9.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-27 3. HVPS noisy - see HVPS test procedure. See Section 9.3.8.5. 4. Defective electronic components on preamp board. - use optic test and electric test to check electronics, optics and observe noise.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-28 9.2.7. Unstable Zero Common causes are: 1. Leak in pneumatic system. 2. Light leak - check the sensor module with strong light. 3. Sample lines or sample filter dirty - clean or replace.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-29 A. A leak that admits gas containing NO. B. Contaminated reaction cell. Remove and clean cell. C. Wet (i.e. undried ambient air) air into the ozone generator. Check the Perm aPure drier and associated plumbing for leaks and correct operation.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-30 7. Ozone concentration too low: A. Wet air in generator - need to disassemble, clean, dry. B. Electrical short circuit. C. Transformer partial failure. 8. Sample inlet vent line too short - should be at least 15".
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-31 9.3. Subsystem Troubleshooting and Adjustments 9.3.1. Computer, Display, Key board The purpose of this section is to determine if the computer subsystem electronics hardware are working properly.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-32 Testing and operation of the CPU RS-232 port is described in Section 9.3.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-33 Figure 9-2: CPU Board Jumper Settings.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-34 9.3.1.3. Front Panel Keyboard The keyboard consists of 8 keys and 3 LED's. Key strokes are sent to the SBC40 computers' parallel port. The computer software detects the key strokes via CPU interrupts.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-35 Figure 9-3: RS-232 Pin Assignments.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-36 9.3.2.1. RS-232 Port If the RS-232 port is not working, check the following: Physical Wiring Refer to Figure 9-3 for the wiring diagram of the DB-9 plug on the M200A rear panel.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-37 Check that the voltage level on Pin 8 of the Analyzer is between +5 V and +15 V. This pin should be connected (through the cable) to Pin 4 of the modem. Now set the baud rate of the Analyzer to the speed required by the modem and it should work.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-38 A. Turn on the instrument. B. Enter the SETUP-MORE-DIAG menu. C. Select the ANALOG OUTPUT test. This causes the M200AH to output a 5 step voltage pattern to the 4 analog outputs on the rear panel.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-39 9.3.3.1. ADC/DAC Calibration Procedure Due to the stability of modern electronics, this procedure should only need to be performed if a major sub-assembly is exchanged or when the display voltage does not m atch the input voltage or current to the V/F card.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-40 10 V output or non-isolated current loop Switch State Comment 3 ON Use for non-isolated current loop or 10 V output 4 OFF.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-41 C. The M200AH will display a voltage near 1% of the voltage range set in the above procedure. Adjust R27 until the displayed voltage matches the DVM voltage, then press ENTR.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-42 Figure 9-4: V/F Board Jumper Settings.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-43 9.3.4. Status/Temp Board The Status/Temp Board is a multif unction board that: 1. Converts the resistance readings of the thermistors to a voltage. 2. Amplifies and scales the thermocouple voltage for the V/F card.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-44 The thermocouple amplifier for the converter tem perature has a gain adjustment.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-45 9.3.4.4. 4-20 mA Current Output There are two types of current loop outputs available for the M200AH, non-isolated (standard equipment) and isolated (optional). These outputs may be used in place of the analog voltage signals that appear at the rear panel connector.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-46 Table 9-9: Power Supply Module Subassemblies Module Description Linear Power Supply Board The linear power supply board takes multiple voltage inputs from the power transformer and produces +5, +15, -15, +12 VDC outputs.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-47 Figure 9-5: Power Supply Module Layout.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-48 Figure 9-6: Electrical Block Diagram.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-49 PSM Diagnostic procedures: The Linear Power Supply board can be tested by checking the DCPS - TEST function on the front panel.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-50 Common faults in the ozone generator are: 1. A leak or some other failure in the drier will let am bient air into the generator. There is enough water vapor in room air to cause the generator to make nitric acid aerosol.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-51 9.3.6.1. The PermaPure Drier The PermaPure drier is constructed of 2 concentric tubes. The inner tube is a special material that has an affinity for water vapor. The outer annulus is evacuated by the instrument pump.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-52 Figure 9-7: Ozone Generator Subsystem.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-53 9.3.7. Flow/Pressure Sensor The flow/pressure sensor board consists of 2 pr essure sensors and a flow sensor. See Figure 9-8 for a diagram of this board. From these three sensors four values are com puted and displayed on the front panel TEST functions.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-54 To adjust the SAMPLE PRESSURE: 1. Go to DIAG mode by pressing SETUP-MORE-DI AG, then select SIGNAL I/O, and press ENTR. Select SAMPLE_PRESS by using NEXT-PREV keys. 2. Adjust R1 to read 4100 mV.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-55 Figure 9-8: Flow/Pressure Sensor.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-56 Figure 9-9: NO X Sensor Module.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-57 Figure 9-10: NO X Sensor Module.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-58 9.3.8. NO X Sensor Module Figure 9-9 and Figure 9-10 show the assembly of the NO X sensor module. 9.3.8.1. PMT The PMT detects the light emitted by the reaction of NO with ozone.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-59 9.3.8.4. PMT Cooler The PMT cooler uses a Peltier cooler supplied with DC current from the switching power supply in the Power Supply Module. An overall view is shown in Figure 9-11.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-60 Figure 9-11: PMT Cooler Subsystem.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-61 9.3.8.5. High Voltage P.S. The HVPS is located in the interior of the Sensor Module, and is plugged into the PMT tube. It requires 2 voltage inputs. The first is +15 VDC, which powers the supply.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-62 Figure 9-12: High Voltage Power Supply.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-63 9.3.9. Z/S Valves The Z/S Valves is an option in the instrument. Before troubleshooting this sub-assembly, check that the options were ordered, and that they are enabled in the software.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 9-64 It is also possible to manually toggle each of the valves in the DIAGNOSTIC mode. Refer to Section 9.1.3 for information on using the DIAG mode . Also refer to Figure 8-4, Figure 8-5 for a pneumatic diagram of the system .
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 A-1 APPENDIX A ELECTRICAL SCHEMATICS Table A-1: Electrical Schematics Part No.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 A-2 Table A-1: Electrical Schematics (Continued) Part No. Name 02034 A-Series Valve Driver Assembly 02035 A-Series Valve Drive.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 B-1 APPENDIX B SPARE PARTS AND EXPENDABLES NOTE Use of replacement parts other than those supplied by API may result in non-compliance w ith European standard EN 61010-1. Table B-1: Spare Parts and Expendables Part No.
Teledyne API Model 200AH NO X Analyzer Instruction Manual, 06492, Rev. G2 B-2 THIS PAGE IS INTENTIONALLY LEFT BLANK.
SPARE PARTS LIST M200AH Part Number Description 000940100 Orifice, 3 mil, 60 cc (IZS) 000940500 Orifice, 7 mil, 250 cc (844-126010) 000940600 Orifice, 10 mil, 500 cc, Rx Cell (844-012607) 002160000 Ozone Scrubber (Charcoal) 002270100 Gasket (Rx Cell) Qty.
SPARE PARTS LIST M200AH 017130000 Model 200AH Level 1 Spare Parts Kit (for 10 units) with MINI-HICON 017130100 Model 200AH Level 1 Spare Parts Kit (for 10 units) with MOLY 017140000 AKIT, EXP KIT, M200AH/EH 018080000 Desiccant Baggie Kit - Qty.
LEVEL 1 SPARES KIT (WITH MINI-HICON CONVERTER) M200AH Part Number Description 000940100 CD, ORIFICE, .003 GREEN 000940500 CD, ORIFICE, .007 ORANGE 000940600 CD, ORIFICE, .
LEVEL 1 SPARES KIT (WITH MOLYBDENUM CONVERTER) M200AH Part Number Description 000940100 CD, ORIFICE, .003 GREEN 000940500 CD, ORIFICE, .007 ORANGE 000940600 CD, ORIFICE, .
EXPENDABLES KIT M200AH Part Number Description 002270100 AKIT, GASKETS, WINDOW, (12) 009690300 AKIT, TFE FLTR ELEMENT, 47MM, 1UM (30) FL0000001 FILTER, SS FL0000003 FILTER, DFU (KB) HW0000020 SPRING OR0000001 ORING, 2-006VT * 01714A - M200AH, EXP Kit.
THIS PAGE IS INTENTIONALLY LEFT BLANK.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Teledyne 200AH è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Teledyne 200AH - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Teledyne 200AH imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Teledyne 200AH ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Teledyne 200AH, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Teledyne 200AH.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Teledyne 200AH. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Teledyne 200AH insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.