Manuale d’uso / di manutenzione del prodotto 102E del fabbricante Teledyne
Vai alla pagina of 90
ADDENDUM TO MODEL 101E OPERATORS MANUAL (P/N 04740 REV. A) FOR MODEL 102E TOTAL REDUCED SULFUR ANALYZER with MODEL 501 TRS THERMAL CONVERTER © TELEDYNE ADVANCED POLLUTION INSTRUMEN TATION 9480 CARROLL PARK DRI VE SAN DIEGO, CA 92121-52 01 USA Toll-free Phone: 800-324-519 0 Phone: 858-657-980 0 Fax: 858-657-981 6 Email: Api-sales@teledyne.
.
M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) TABLE OF CONTENTS 1. PREFACE ..................................................................................................................... ................. 1 1.1. Reference Numb er ing convention .
M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 8. INSTRUMENT MAINTENANCE ..................................................................................................... 3 9 8.1. Additional and Upd ate d Maintenance Proced ures .....
M102E/M501 TRS PREFACE (Addendum to M101E Manual - P/N 04740 Rev A) LIST OF FIGURES Figure 3-1: M 102E Internal Lay out ................................................................................... 7 Figure 3-2: M501 -TRS Internal L ayout.......
.
M102E/M501 TRS PREFACE (Addendum to M101E Manual - P/N 04740 Rev A) 1. PREFACE NOTE The information contained in t his addendum is pertinent to M102E anal yzers running software revisi on A.2. Some or all of the in formati on may not be applicable t o previous revision of that soft ware.
PREFACE M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 1.1. Reference Numbering convention Unless otherwise specified, chapter, section, figu re and table reference num bers referred to within this text are relative to thi s document. EXAMPLE: “Figure 2- 1” refers to the figure, within this document , labeled as 2-1.
M102E/M501 TRS SPECIFICATIONS, APPROVALS AND WARRANTY (Addendum to M101E Manual - P/N 04740 Rev A) 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. Specifications There are no significant differenc es between the performance specif ications for the M102E and the M102E as listed in Section 2.
SPECIFICATIONS, APPRO VAL S AND WARRANTY M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) • Sample filter: Equipped with PTFE filter element in the in ternal filter assembly. • Sample flow of 650 +/- 65 cc/min. • Vacuum pump (internal or external) capable of 14"Hg Absolute pressure @ 1 slpm or better.
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3. GETTING STARTED 3.1. Unpacking the M102E Unpack the M102E as per the directions ins Se ction 3.1 of the M101E Manual - P/N 04740 Rev A. 1. There are no shipping screws to be removed in the M102 E.
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3.2.1. M501-TRS Ventilation Clearance: Whether the M501-TRS is set up on a be nch or installed into an inst rument rack, be sure to leave sufficient ven tilation clearance. AREA MINIMUM REQUIRED CLEARANCE Back of the instrument 10 cm / 4 inches Sides of the instrument 2.
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3.3. Internal Layouts Figures 3-1 & 3-3 supersede Figure 3-9 of the M101E Manual - P/N 04740 Rev A.
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3.4. Internal Pneumatic Flow of the M102E & the M501-TRS Figure 3-3 shows th e internal pneumatic flow of the M102E in its Stan dard configuration. For information on in struments in which one of t he various zero/span valve opt ions refer to Figures 5- 2 and 5-3.
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3.5. Rear Panel Layout for the M102E & M501-TRS Figures 3-4 & 3-5 supersede Figure 3-2 of the M101E Manual - P/N 04740 Rev A.
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3.6. Initial Setup 3.6.1. Electrical Connections: The electrical connections for the M102E are the sa me as those described in Section 3.1.1 of the M101E Manual - P/N 04740 Rev A except for the test channel analog output: 3.
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) • TO change these settings, see Sections 6. 9.4 and 6.7 of the M101E Manual - P/N 04740 Rev A respectively.
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) FROM ANALYZER Sample gas enters the M501-TRS from the M102E via this port. TO ANALYZER Sample gas leaves the M501-TRS to return to the M102E via thi s port.
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) CAUTION The exhaust from th e instrument needs to be vented outside the immediate area or shelter surrounding the instru ment and conform to all safety requirements using a maximum of 10 meters of 1/4” PT FE tubing.
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Table 3-4: NIST-SRM's Availa ble for Traceability o f H 2 S & S O 2 Calibration Gases NIST-SRM 4 TYPE NOMINAL CONCE.
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) Source of SAMPLE G as V ENT if i nput is pressurize d Zero / S p an Valves – O p tion 50 MODEL 700 Gas Dilution Calibrator.
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Source o f SAMPLE G as VENT if in put is p ressurized MODEL 700 Gas Dilution Calibrator (with O 3 generator option) VENT Cal.
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3.7. Initial Operation 3.7.1. Startup / Warm Up of the M102E Startup proc edures and warm up behavior of the M102E are identical to those described in Sections 3.2.1 and 3.2.2 of the M1 01E Manual - P/N 04740 Rev A .
GETTING STARTED M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3.7.3. Startup / Warm Up of the M501-TRS After electrical and pneumatic conn ections are ma de, turn on the instr ument and pump powe r.
M102E/M501 TRS GETTING STARTED (Addendum to M101E Manual - P/N 04740 Rev A) 3.8. Initial Calibration Initial calibr ation of the M102E should be performed with : • Zero air supplied by a zero air ge.
.
M102E/M501 TRS OPTIONAL HARDWA RE AND SOFTW ARE (Addendum to M101E Manual - P/N 04740 Rev A) 4. OPTIONAL HARDWARE AND SOFTWARE This section includes descriptions of the hardwa re and software options .
OPTIONAL HARDWA RE AND SOFTW ARE M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3 2 1 SAMPLE/CA L V ALVE ZERO/SPAN VALVE 3 2 1 2 3 1 SPAN G AS INLET ZERO A IR INLET VACUUM MANIFOLD FLOW C.
M102E/M501 TRS OPTIONAL HARDWA RE AND SOFTW ARE (Addendum to M101E Manual - P/N 04740 Rev A) ` FLOW SENSOR FLO W / PRESS URE SENSOR P C A SAMPLE PRESSURE SENSOR EXHAUST T O OUTER L AYE R OF KICKE R 3 .
OPTIONAL HARDWA RE AND SOFTW ARE M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 4.3. Additional Manuals 4.3.1. Printed Manuals (P/N 049880000) Additional printed copies of this addendum are avai lable from Teledyne Instrume nts 4.3.2. Addendum on CD (Part number 049880200) This addendum is also available on CD.
M102E/M501 TRS M102E OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) 5. M102E OPERATING INSTRUCTIONS NOTE For th e most pa rt the op eratio n instru ction fo r the M102 Eare the s .
M102E OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) The output, labeled A3 is special. It can be set by th e user (see Section 6.9.10 of the M101E Manual - P/N 04740 Rev A) to output many of the parameters accessible through the <TST TST> keys of the units Sampl e Display.
M102E/M501 TRS M102E OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) 5.2.2. M102E Test Channel Output The following table supercedes Table 6- 14 of the M101E Manual - P/N 04740 Rev.
M102E OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 5.4. Remote Operation of the Analyzer 5.4.1. Control Inputs The description of the control inputs in Section 6.12.1.2 of the M101 E Manual - P/N 04740 Rev A in correctly shows an external lo w span cal input.
M102E/M501 TRS M102E OPERATING INSTRUCTIONS (Addendum to M101E Manual - P/N 04740 Rev A) CONTRO L IN - + 5 VDC Power Supply LOW SPA N ZERO SPAN A B C D E F U + Figure 5-3: Control Inputs wi th external 5 V power supply 5.4.2. Using the M102E with a Hessen Protocol Network 5.
M102E OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 5.4.2.2. Setting Hessen Protocol Status Flags The following table supercedes Table 6- 29 of the M101E Manual - .
M102E/M501 TRS M501-TRS OPERATING INSTRUCTI ONS (Addendum to M101E Manual - P/N 04740 Rev A) 6. M501-TRS OPERATING INSTRUCTIONS CAUTION ! DO NOT OPERATE WITHO UT THE COVER OF THE M 501TS CONVERTER INSTALLED. OVEN TEMPERATURE WILL NOT REGUL ATE PROPERLY WITHOUT THE COVER IN PLACE.
M501-TRS OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Table 6-1: M501-TRS Temperatur e Controls and Definiti ons NAME FUNCTION Main Display A 4-digit, 7 segment LED display on wh ich the current value of the PV and the SV as well as error codes and programming parameters and data are all displayed.
M102E/M501 TRS M501-TRS OPERATING INSTRUCTI ONS (Addendum to M101E Manual - P/N 04740 Rev A) 6.3. To Manually Adjust the Converter Oven Temperature: CAUTION ! DO NOT SET THE TEM PERATURE HIGHER THAN 1050 OC 1. Set the main display to show the current value of t he set variable by pressing the PV/SV mode key.
M501-TRS OPERATING INSTRUCTIONS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 6.4. Autotune the Temperature Controller: The M501-TRS controller includes an auto tune feat ure which allow.
M102E/M501 TRS M501-TRS OPERATING INSTRUCTI ONS (Addendum to M101E Manual - P/N 04740 Rev A) 6.4.2. Aborting the Autotune Process: 1. Press the DATA key once. The display will show (1). 2. Press the 1’s UP key once. The will begin blinkin g 3. Press the DOWN key on ce.
.
M102E/M501 TRS CALIBRATION PROCEDURES (Addendum to M101E Manual - P/N 04740 Rev A) 05514 Rev A1 37 7. CALIBRATION PROCEDURES 7.1. M102E Calibration Calibration of the M102E should be performed according to the proce dures described in Chapters 7 & 8 of the M101E Manual - P/N 04740 Rev A.
.
M102E/M501 TRS INSTRUMENT MAINTENANCE (Addendum to M101E Manual - P/N 04740 Rev A) 05514 Rev A1 39 8. INSTRUMENT MAINTENANCE The following table superc edes Table 9-1 of the M101E Manual - P/N 04740 R.
INSTRUMENT MAINTENANCE M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 8.1. Additional and Updated Maintenance Procedures The following procedures need to be performed re gularly as part of the standard maintenance of the Model 102E. 8.1.1.
M102E/M501 TRS INSTRUMENT MAINTENANCE (Addendum to M101E Manual - P/N 04740 Rev A) 8.1.1.2. Checking the Func tion of the SO 2 Scrubber To check to see if your SO 2 scrubber is operating properly perform th e following test: 1. Set the analyzer for TRS gas measurement mode (see 6.
.
M102E/M501 TRS THEORY OF OPERATION (Addendum to M101E Manual - P/N 04740 Rev A) 9. THEORY OF OPERATION The M102E is a modified M101E which, when used in conjunction with a M501-TRS determines the conc.
THEORY OF OPERATION M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) nm. Th e SO 2 molecules absorbs some of energy from th e UV light causing one of the electrons of each of the affected molecules to move to a higher energy orbital state.
M102E/M501 TRS THEORY OF OPERATION (Addendum to M101E Manual - P/N 04740 Rev A) The amount of det ectable UV g iven off by the decay of the SO 2 * is affected by the rate at wh ich this reaction occurs ( k ). ( ) * SO k F 2 = (Equation 9-5) Where: F = the amount of fluorescent li ght given off.
THEORY OF OPERATION M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 9.2. The UV Light Path The following in formation is in addition t o that contained in Section 10.
M102E/M501 TRS THEORY OF OPERATION (Addendum to M101E Manual - P/N 04740 Rev A) 9.4. Electronic Operation 9.4.1. Sensor Module This Section replaces Section 10.
THEORY OF OPERATION M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 9.4.1.1. Sample Chamber The main electronic components of the sample chamber are the refe rence detector(see Section 10.2.2 of the M101E Manual - P/N 04740 Rev A ); th e UV Lamp (see Se ction 10.
M102E/M501 TRS THEORY OF OPERATION (Addendum to M101E Manual - P/N 04740 Rev A) 9.4.2. M501-TRS electronics Electronically the M501-TRS is a si mple device.
THEORY OF OPERATION M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 9.4.2.2. Temperature Alarms and Alarm Output The M501-TRS’ controller has two user settable alar m points: High Alarm and Low Alarm.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 10. TROUBLESHOOTING & REPAIR This section includes v arious troubleshooting and repair in formation that is.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 10.1.2. Fault Diagnosis with Test Functions The Following table supercedes Table 11.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 10.2. M501-TRS Trouble shooting 10.2.1. TRS Converter Not Heating: Problems with heating the If the TRS co nverter oven can have several causes.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 10.4. Subsystem Checkout In addition to the informatio n contained in Section 11.5 of the M101E Manual - P/N 04740 Rev A, the following diagnosti c procedures are useful for troubleshooting and diagnosi ng problems with your M102E and M501-TRS.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 10. If TRS Æ SO 2 converter efficiency i s below 90% check for one of the foll owing possible causes: • Pneumatic leak • Perform a leak check (with the M501-TRS connected) as described in Section 11.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Lamp Positioning – The UV output level of the lamp is not even across the entire length of the lamp. Some portions of th e lamp shine slightly more brightly than others.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) UV Lamp B racket Mounting Screws Shutter Housi ng Shutter Assy UV Filter Retain er & Lens Housing UV Lamp P.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 9. Finger tighten the th umbscrew. NOTE: DO NOT over-tighten the thumbscrew.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 5. Carefully remove the UV filter. 6. Install the UV filter. Handle carefully and never t ouch the filter’s surface. The UV filter’s wider ( ring ) side should be facing out.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 3. Remove the reaction cell assembly. 4. Remove the two connectors on the PMT housin g end plate facing towa rds the front panel. 5. Remove the end plate itself (4 scre ws with plastic washers).
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) 15. Re-insert the PMT/HVPS subassembly. • Don’t forget the gasket between HVPS and PMT . • Use new plastic screws to mount the PM T assembly on the PMT cold block.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 6. Locate the Preamp board (see Figure 3-1). 7. Locate the Following Components On the Preamp board (see Figure.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) If a reporting range other than 50 0 ppb is u sed in this pro cedure : Use a span gas equal to 80% of the report ing range and adjust the PMT to a target NORM PMT value of twice the ppb value of the span gas.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) 2. Allow it to cool to room temperatur e. This may take up to 30 minutes. 3. Remove the instruments top cover (s ee Section 3.2 of this addendum) 4. Remove the four screws holdi ng the converter cover in place.
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) DO NOT manually alter the PID parameters of the M 501-TRS temperatur e Controller unless directed to do so by Teledyne Instruments customer service per sonnel.
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) Table 10-3 – Temperature C ontrolle r – Primar y Parameter Settings DISPLAY PARAMETER NAME SET TO COMMENTS .
M102E/M501 TRS TROUBLESHOOTING & REPAIR (Addendum to M101E Manual - P/N 04740 Rev A) Table 10-4 – Temperature C ontrolle r – Primar y Parameter Settings DISPLAY PARAMETER NAME SET TO COMMENTS .
TROUBLESHOOTING & REPAIR M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) User Notes: 68 05514 Rev A1.
M102E/M501 TRS APPENDIX A - Version Specific Software Document ation (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX A - Version Specific Software Documentation APPENDIX A-1: Model 102E Software Menu Trees NOTE The menu tree struct ure for the M102E is nearly identical to th at of the M101E.
APPENDIX A-1: M102E Software Menu Trees, Revisio n A.2 M102E/M501 TRS (Addendum to M101E Manual) APPENDIX A-1: M102 E Software Menu Tre es, Revision A.2 ZERO SPAN CONC SAMPLE <TST TST> RANGE STABIL PRES SAMP FL PMT NORM PMT UV LAMP LAMP RATIO STR.
APPENDIX A-1: M102E Software Menu Trees, Revisio n A.2 M102E/M501 TRS (Addendum to M101E Manual) LOW HIGH ZERO SPAN CONC SAMPLE <TST TST> SETUP COMM DIAG VARS (Secondary Setup Menu) CFG DAS RANG PASS CL K MORE ( Primary Se tup Menu) Only appear if reporting range is s et fo r AUTO range mode.
APPENDIX A-1: M102E Software Menu Trees, Revisio n A.2 M102E/M501 TRS (Addendum to M101E Manual) . YES NO SAMPL E FLO W WAR NING BENCH T EMP WARNING SOURCE WAR N I NG BOX TEMP WARNING WHEEL TEM P WAR .
X A-1: M102E Software Menu Trees, Revision A.2 M102E/M501 TRS (Addendum to M101E Manual) 515 Rev A.2 A-5 32 INTERNAL ANALOG ↓ VOLTAGE SIGNALS 56 (see Ap pendix A) CFG ACAL 1 CLK PASS RNGE DAS MORE C.
APPENDIX A-2: Setup Variables For Se rial I/O, R evision A.2 M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX A-2: Setup V ariables For Se ria l I/O, Revision A.
M102E/M501 TRS APPENDIX A-3: War n i ngs and Test Functions, Revi sion A.2 (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX A-3: Warnings and Test Functions, R evision A.
APPENDIX A-4: M102E Signal I/ O Definitions, Revision A.2 M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX A-4: M102E Signal I/O Definiti ons, Revision A.
M102E/M501 TRS APPENDIX A-5: M102E iDAS Functions, Revision A.2 (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX A-5: M102E iDAS Func tions, Revision A.
APPENDIX A-5: M102E iDAS Functions, Revision A.2 M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) A-10 05515 Rev A.2.
M102E/M501 TRS APPENDIX B - M102E Spare Parts List (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX B - M102E Spare Parts List NOTE Spare parts for the M102E are t he same as those for the M101E .
APPENDIX B - M102E Spare Parts List M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) B-2 05516 Rev A.
M102E/M501 TRS Warranty/Repair TELEDYN E Addendum to Appendix C M101E Manual (P/N 04740) Questionnaire Model 102E I NSTRUMENTS Advanced Poll uti on I nst r umentat i o n A Teledyne Techno logi es Com .
M102E/M501 TRS Warranty/Repair TELEDYN E Addendum to Appendix C M101E Manual (P/N 04740) Questionnaire Model 102E I NSTRUMENTS Advanced Poll uti on I nst r umentat i o n A Teledyne Techno logi es Com .
M102E/M501 TRS APPENDIX D - ELECTRONIC SCHEMATICS (Addendum to M101E Manual - P/N 04740 Rev A) APPENDIX D - ELECTRONIC SCHEMATICS The following drawing(s) are relevant to the M501-TRS. All drawings relevant t o the M102E are listed in Appendix D of the M101E Manual.
APPENDIX D - ELECTRONIC SCHEMATICS M102E/M501 TRS (Addendum to M101E Manual - P/N 04740 Rev A) User Notes D-2 05518 Rev A.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Teledyne 102E è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Teledyne 102E - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Teledyne 102E imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Teledyne 102E ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Teledyne 102E, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Teledyne 102E.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Teledyne 102E. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Teledyne 102E insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.