Manuale d’uso / di manutenzione del prodotto PM5300 del fabbricante Precision Trading
Vai alla pagina of 16
USER MANUAL AIR-O XYGEN BLENDER (DIS S and NIST Connections) Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS CAUTION F ederal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive T el: (+001) 610-262-6090 Northampton, P A 18067 USA F ax: (+001) 610-262-6080 www .
1 Blender Air-Oxygen CONTENTS RECEIVING / INSPECTION ......................................................... 2 INTENDED USE .......................................................................... 2 READ ALL INSTRUCTIONS BEFORE USING .............
2 Blender Air-Oxygen RECEIVING / INSPECTION Remove the Precision Medical, Inc. Air -Oxygen Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider .
3 Blender Air-Oxygen SAFETY INFORMA TION - W ARNINGS AND CAUTIONS DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury . WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury .
4 Blender Air-Oxygen WARNING An Oxygen Analyzer/Monitor must be used to verify oxygen concentration. Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15 lpm for the High Flow Blender , and 3 lpm for the Low Flow Blender .
5 Blender Air-Oxygen SPECIFICA TIONS Model PM5200 High Flow PM5300 Low Flow Primary Outlet Flow Range 15 - 120 lpm 3 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 . 4 b a r ) w i t h N O B L E E D Auxiliary Outlet Flow Range 2 - 100 lpm 0 - 30 lpm W i t h b o t h s u p p l y p r e s s u r e s a t 5 0 p s i ( 3 .
6 Blender Air-Oxygen SPECIFICA TIONS continued Dimensions: (without fittings) Depth: 4.9 in (12.5 cm) Width: 2.3 in (5.7 cm) Height: 4.1 in (10.4 cm) W eight: 2.
7 Blender Air-Oxygen DIAGRAMS CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. Depending on model, your fittings may differ from these diagrams.
8 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION A Oxygen Concentration Dial A di al use d fo r sel ect in g ox yg en con ce ntr ati on s be tw een 21%-100%. The FiO 2 scale is used for reference only . This Dial does not rotate 360°. The dial starts at 21% and ends at 100%.
9 Blender Air-Oxygen COMPONENT DESCRIPTION ITEM DESCRIPTION F Air Inlet Fitting A male DISS or NIST air fitting with one way valve that is used to connect an air supply hose . G Alarm An audible alarm that sounds due to an excessive pressure drop or deletion of either gas supply .
10 Blender Air-Oxygen ALARM TEST 1. The Air -Oxygen Blender is installed and the flowmeter is turned on. Disconnect or turn off the air supply line to the Air -Oxygen Blender . 2. The Blender should alarm with a loud whistle noise. The whistle indicates the alarm is operating correctly .
11 Blender Air-Oxygen 3. T o activate the bleed, turn the knurled collar until the bleed pin is engaged. Then slide the collar back until it contacts the cover . 4. Confirm the concentration of air/oxygen with an Oxygen Analyzer/Monitor . 5. Confirm the flow of air and/or oxygen mixture to the patient.
12 Blender Air-Oxygen 1. Disconnect all gas connections and equipment before cleaning. 2. Clean exterior surfaces with a cloth dampened with mild detergent and water .
13 Blender Air-Oxygen TROUBLESHOOTING If the Air -Oxygen Blender fails to function, consult the T roubleshooting Guide below . If problem cannot be solved by using T roubleshooting Guide refer to the Air -Oxygen Blender Service Manual (P/N 504827) available on the Internet; www .
14 Blender Air-Oxygen LIMITED W ARRANTY AND LIMIT A TION OF LIABILITY Precision Medical, Inc. warrants that the Blender , (the Product), will be free of defects in workmanship and/or material for the .
DECLARA TION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held Drive, Northampton, P A 18067, USA CONT ACT : Quality Manager Phone: 610-262-6090 Authorized European Representative: Emergo E.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Precision Trading PM5300 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Precision Trading PM5300 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Precision Trading PM5300 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Precision Trading PM5300 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Precision Trading PM5300, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Precision Trading PM5300.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Precision Trading PM5300. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Precision Trading PM5300 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.