Manuale d’uso / di manutenzione del prodotto 1522 del fabbricante Fluke
Vai alla pagina of 60
1522 Handheld Thermometer Readout User’ s Guide Rev. 5B2202 Hart Scientific.
Rev. 5B2202 Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Subject to change without notice.
T able of Contents 1 Bef ore Y ou Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 W ARNINGS .
6.5 Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 6.6 INFO-CON Connector . . . . . . . . . . . . . . . . . . . . . . . 29 7 Display Functions . . . . . . . . . . . . . . . . . . . . . . . . . 31 7.1 Blank Mode . . . . . . . .
8.2.1 Primary Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 8.2.2 Calibration Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 9 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 9.
iv Figures and T ables T able 1 International Electrical Symbols . . . . . . . . . . . . . . . . . . . . . 1 Figure 1 Front V iew . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Figure 2 T op and Side V iew . . . . . . . . . . . . .
1 Bef ore Y ou Start 1.1 Symbols Used T able 1 lists the International Electri cal Symbols. Some or all of these symbols may be used on the instrument or in this manu al.
Symbol Description Canadian Standards Associati on O VERV OL T AGE (Installat ion) CA TEGORY II, P ollution Degr ee 2 per IEC1010-1 re - f ers to the le vel of Impulse Wit hstand V oltage protect ion provided. Equipment of O VERV OL T AGE CA TEGOR Y II is energy-consuming equipment to be suppli ed from the fix ed installat ion.
• The A C adapter has circuits with high voltage inside that could present danger of electrical shock or fire if exposed. If the A C adapter is damaged in any way or becomes hot, discontinue its use immediately , disconnect it from any A C supply , and hav e it replaced.
• Keep the probe wires clean and aw ay from fluids. 1.3 A uthoriz ed Service Centers Please contact one of the follo wing authorized Service Centers to coordinate service on your Hart product: Fluke Corporation, Hart Scientif ic Division 799 E. Utah V alle y Driv e American Fork, UT 84003-9775 USA Phone: +1.
Fluke South East Asia Pte Ltd. Fluke ASEAN Regional Of fice Service Center 60 Alexandra T errace #03-16 The Comtech (Lobby D) 118502 SINGAPORE Phone: +65 6799-5588 T elefax: +65 6799-5588 E-mail: antng@singa.
2 Intr oduction The Hart 1522 is a low-cost, high-accura c y digital thermometer readout de - signed to be used with PR Ts and thermistors. The unique combination of fea - tures makes this instrument suitable fo r a wide v ariety of application s in industry .
3 Specifications and En vir onmental Conditions 3.1 Specifications Resistance Range 0 Ω to 500 k Ω Resistance Accuracy, PRT, one year † 0 Ω to 25 Ω : 0.002 Ω 25 Ω to 400 Ω : 0.008% (80 ppm) of reading Resistance Accuracy, thermistor, one year † 0 Ω t o2k Ω :0 .
Digital Filter Exponential with adj ustable time constant (1 to 60 seconds) Remote Communications RS-232 s erial, infrared Display LCD, 6-digit x 7-segment wi th 16-character alphanumeric Operating Te.
This warrant y extends onl y to the original purchaser and shall not apply to an y product which, in Hart’ s sole opinion, has been subject to misuse, al teration, abuse or abnormal c onditions of operation or handling. Software is w arranted to operate in accordance with its programmed instruc - tions on appropriate Hart products.
4 Quic k Start This section briefly explains the basics of setting up and operating your ther - mometer readout. 4.1 Unpac king Carefully unpack the 1522 and accessories and inspect them to make sure all components are present and in satisf actory condition.
4.4 Connect the Probe The PR T or thermistor probe connects to the top of the 1522 using Hart’ s unique INFO-CON probe connector . The probe connector is inserted into the top of the instrument wi th the ridged to the back and f its snugly and locked into place when it is fully inserted.
want to measure the te mperature. DO NO T f orce the probe or otherwi se allo w it to be bent, stressed, or o verheat ed. It can be easily damaged if misused. F or further suggestions on handl ing the probe and using the 1522 and probe to measure temperature accurately , see Section 6.
5 P ar ts and Contr ols The functions of the v arious features of the 1522 are described below . 5.1 Fr ont View The front of the 1522 features the LCD display and control b uttons.
Display -The display sho ws the current temperature (or resist ance) measure - ment on the large numeri c upper portion of the display . It can also show a v ari - ety of information on the small er alphanumeric lo wer portion such as minimum, maximum, delta(x), an d other instru ment settings.
5.2 T op and Side View The top and side of the 1522 feature the probe connector , DC power input, se - rial port, and infrared windo w . Probe Connector - At the t op of the thermometer readout is the opening where the probe connector is inserted. The probe must be connected using an INFO-CON probe connector to measure temperature (see Section 6.
5.3 Bac k View See Figure 5 on page 25. Stand - The stand at the back of the 1522 can be fl ipped down t o prop up the instrument for better viewing. Battery Compartment - Behind the stand is the compartment that contains the battery pack.
5.5 Accessories The 1522's accessories and their features are described here. AC A d a p t e r - The A C adapter recharges the internal battery pack and may also be used to supply po wer to operate the 1522 whi le the battery i s being charged (see Section 6.
Ordering Infor mation Model Type Range Size Cal Uncertainty and Repeatability † Typical Drift † (1 year) 5613 Pt-100 Ohm, 4-wir e –200 to 300°C 3/16” x 6” (4.76 x 152 mm) ±0.025°C ±0.01°C 5612 Pt-100 Ohm, 4-wir e –200 to 450°C 3/16” x 9” (4.
computer and written to a file quickly . These items must be p urchased sepa - rately . Log Wa r e Softwar e -T h eL o g Wa r e software are data acquisition software packages that can be used with your t hermometer to collect and graph data.
6 General Operation This section e xplains the details of the operat ion of the 1522 with its compo - nents and accessories. 6.1 Battery The 1522 has a b uilt-in nickel-metal-hydride battery pack that can po wer the instrument for about 18 hours before needing to be rechar ged.
tery automaticall y when it’ s fully char ged. It normally tak es about four hours to fully charge the battery . The instrumen t can be operated while the battery is b e - ing charged. CA UTION: DO NOT leave the A C adapter plugged in for more than 24 consecutive hours or , the battery life could be degr aded.
adapter supp lied by Hart Scientific. The DC power source plugs in to the DC power i nput on the right side of the 1522. W ARNING: The AC adapter contains high voltage cir cuits that could pr esent danger of electric shock or f ire if e xposed.
second. The display unit associ ated with the measurement is sho wn in smaller type to the right of the measurement. The possible dis play units are degrees Celsius (C), resistance in oh ms ( Ω ), de grees Fa hrenheit (F), Ke lvin (K), and degrees Rankine (R).
T emperature is generally sensed at the tip of the probe. T o get an accurate tem - perature measurement the probe sheath should be well immersed, with ade - quate depth and fit, into the mediu m to be measured.
C1 and P1 terminals and the other single wire to t he C2 and P2 terminals (which are shorted toget her). 1522 Handheld Thermometer Readout User’s Guide 30 GND P1 C2 P2 C1 J1 J2 Hart Scientific Four-.
7 Displa y Functions The insturment operates in an y of sev eral modes which determine what infor - mation is visible in th e lo w er alphanumeric portion of the disp lay . The MODE button sequences through the menu functions. This b utton can be pressed re - peatedly until a desired mode is set.
Each of these operating mod es is described in the sections that follo w . 7.1 Blank Mode This mode is identified by “Blank” that temporarily app ears on the display .
In this mode nothing appears on the alphanumeric porti on of the display . Use this mode if you are only interested in the latest measurements and want to sim - plify the appearance of the display . The message “Prb cal expired”, “M tr cal expired”, “ Recharge Needed”, or “Probe is locked” may be displayed.
The labels can be pre-loaded using the W indo ws® HyperT erminal or o ther se - rial communications program and the command LBnn[=nnnnnnnn] as de - scribed in Section 8.
7.4.1 Clear A utolog? This function clears t he memory of all autologged dat a. Once the 1522 has logged 10,000 readings, further automati c logging is pre vented until the mem - ory has been cleared. Pressing the CLR button cancels the “Clr Autolog?” function and adv ances to the ne x t function.
Press the ENT butt on to stop autologgi ng and to adv ance to the ne xt function. 7.4.5 Dump Data? This function allo ws the user to download t he autologged data using the seri al port or print the do wnloaded dats to a serial printer .
button stores the selection. If th e MODE b utton or CLR button is pressed with - out pressing ENT the unit remains unchanged. Once the unit is select ed, subse - quent measurements will appear in that unit. After the Min/ Max and Delta(x) modes are initiated, their values are automatically con verted when changing units between °C, K, °F , or R.
When editing any parameter , be sure to press the ENT b utton through the enti re line of alphanumeric characters to ensure that the edi ts are stored. Pressing any button other than the ENT button adv ances to the next function wit hout storing the new v alue.
b4 [ITS-90 only] - Set the b4 or b5 probe coef ficient R(0) [CVD only] - Set the R0 probe coefficient Alpha [CVD only] - Set the alpha probe coefficient Delta [CVD only] - Set the delta probe coeffici.
pressed, the message “In valid Passcode” i s displayed and the instrument is re - turned to the display mod e that follows the function that was acti ve when the CAL Mode was ent ered. Each indi vidual instrument has its o w n unique passcode. The passcode number is gi ve n on a notice pro vided with this i nstrument.
7.9.8 Prb T ype This function specif ies the type of probe an d its characterization. It allows the instrument to use the appropriate al gorithm to calculate te mperature from the measured resistance of the probe.
Coefficien ts Cer tificate V alu e R(.01) Rtpw a a5, a6, a7, a8, a9, a10, or a11 b b5, b6, b7, b8 or b9 cc 6 o r c 7 dd a4 a4 or a5 b4 b4 or b5 Consider the follo wing e xamples. Example 1: A PR T was calibrated to ITS-90 and its calibration certificate states v alues for coeff icients R(273.
Coefficien ts Certificate Va lue R(.01) R(273.16K) aa 5 bb 5 c0 . 0 d0 . 0 a4 a5 b4 b5 7.9.10 IEC751 The IEC751 probe type is for standard PR Ts (R TDs) conforming to the IEC751 or DIN-43760 standards. This type of probe is generally lo w cost and has a wide operating range (–180°C to 420°C or higher), b ut has limited accurac y (0.
7.9.13 THERM The THERM type is for th ermistors that hav e u nique calibration coeff icients or do not match the YSI-400 speci ficati on. This type of probe general ly has mod - erate cost, a limited operating range (–40°C to 150°C), excellent accurac y (0.
measurement error due to the resistance of the wires. W ith four wires the in - strument can completely cancel the error . The internal measurement circuit is configured di fferentl y depending on the number of wires. Note: The instrument tries to calculate a measurement e ven if the wire settings are incorrect.
each date segment and the ENT b utton to move to the ne xt date segment on the right. Press ENT when the date is correctly entered. Press CLR to move to the next funct ion without changing the date. 7.9.19 MDue This function sh o ws the date the instrument is du e for calibration and allows it to be set.
8 Comm unications Interface Remote communications allows an e x ternal de vice, such as a computer, to com - municate with the i nstrument to obtain measurement data and control its opera - tion. Communication is accomplished with v arious commands issued to the instrument through the RS-232 serial port.
The protocol for serial communicatio ns is 8 data bits, 1 stop bit, and no parity . Use no flow contr ol. The baud rate is f ixed at 2400, the linefeed to ON (all car - riage returns are follo wed by a linefeed (ASCII decimal 10)), and the duple x to HALF disabling ech o.
Description Command Example Response Format Response Example Range Set autolog mode on/off . Note: Exit Autolog menu to use this command. Aft er alog =ON is sent, the but tons are disabled until t he command alog=OFF is sent. alog=[ON/OFF] alog=ON ON or OFF Clear autolog memory aclr aclr Initiate downl oad of autologged data.
Description Command Example Response Format Response Example Range n Numeric data supplied by user 9 Numeric data retur ned to user x Character data r eturned to user 8.2.2 Calibration Commands The following group of commands directly or indirectly af fect the accurac y of the instrument.
Description Command Example Response Format Response Example Range Read probe wires w[ires] wi wi: 9 wi: 4 Set probe wire s w[ires]= 2/3/4 wi=4 2,3,4 Read probe calibrati on date pc[al] pcal pc: yyyy-.
9 Calibration The 1522 should be calibrated a t regul ar interv als to ensure t hat it continues to measure with proper accuracy . Calibration should only be done by qualified, authorized personnel.
4. V erify the accuracy at 0 Ω , 4k Ω ,1 0k Ω ,4 0k Ω , 100 k Ω , and 500 k Ω . 5. Set the meter calibration date to the present date (see Section 7.
10 Maintenance • This instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product dev elopment. W ith proper care the instrument should require v ery little maintenance.
11 T r oubleshooting In case you run into dif ficulty while operat ing the 1522, this sect ion provides some suggestions that may help you solve the probl em. Belo w are sev eral situa - tions that may arise follo wed by possible causes of the problem and suggested actions you might take .
Problem Solution While attempti ng to measure re - sistance the display shows an i n - correct value. Poor or incorrect connecti on of the probe. A common mista ke is to connect the wires of t he probe to the wrong t erminals. Check the wir - ing carefully ( see Figure 7).
a) Set up the terminal program to use the correct COM port and operate at 2400 baud, 8 data bits, 1 stop bit, no parity, no flo w control. b) If you want to sav e the downloaded data to a file, conf igure your ter - minal program to capture the incoming text and save it to a te xt file.
sion. For best results , place the IR dongle and the IR windo w on the 1522 at an angle of 0° and within 0.5 meters. 11.4 CE Comments 11.4.1 EMC Directive Hart Scientific’ s equipment has been tested to meet the European Electromag- netic Compatibility Directiv e (EMC Dir ectiv e, 89/336/EEC).
Electric F ast T ransit (EFT , Burst, IEC 61000-4-4). If the instrument is sub - jected to EFT conditi ons at 2kV , the instrument may require the user to c ycle the power to return to normal operation.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Fluke 1522 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Fluke 1522 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Fluke 1522 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Fluke 1522 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Fluke 1522, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Fluke 1522.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Fluke 1522. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Fluke 1522 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.