Manuale d’uso / di manutenzione del prodotto 5081T-FF/FI del fabbricante Emerson
Vai alla pagina of 24
Model 5081T -FF/FI Instruction Sheet PN 51A-5081T -FF/FI/rev .A November 2005 Y our purchase from Rosemount Analytical, Inc. has result- ed in one of the finest instruments available for your p ar- ticular application. These instrument s have been designed, and tested to meet many national and interna- tional standards.
PHYSICAL SPECIFICA TIONS HOUSING: Epoxy-polyester painted over low-copper alu- minum. Neoprene O-rings on cover . 160.5 mm x 175.3 mm x 161.3 mm (6.3 in.
3 3 MODEL 5081T -FF/FI SPECIFICA TIONS TRANSMITTER DISPLA Y DURING CALIBRA TION AND PROGRAMMING (FIGURE 1) 1. Continuous display of conductivity , % concentration, or custom reading. 2. Units: µS/cm, mS/cm, ppm, %, or none. 3. Current menu appears here.
4 U1 7 :! <[9:1 -E< / ! 922<U '"*#=22 /1[[! /9P<U <Q2U 4-SG U<[9:1 2D>[=T '""" 25.0C 12.00mA CAL key PROG key DIAG key HOLD key Model 5081T -FF/FI Pr.
MODEL 5081T -FF/FI DEF AUL T SETTINGS V ARIABLE NAME MNEMONIC F ACTOR Y SETTINGS CUSTOMER SETTINGS Program Menu T emperature UH OR Auto temperature compensation UDVUQ on ___________ Manual temperature UODP 25.
6 MODEL 5081T -FF/FI WIRING WIRING Wire sensor as shown below in Figure 2. Keep sensor wiring separate from power wiring. For best EMI/RFI protection, use shielded output signal cable in an earth-grounded metal conduit. Refer to the sensor instruction manual for more det ails.
7 7 MODEL 5081T -FF/FI WIRING WIRING THROUGH A JUNCTION BOX The sensor can be wired to the analyzer through a remote junction box (PN 23550-00). Wire the extension cable and sensor cable point-to-point. Refer to the sensor instruction manual for more details.
8 MODEL 5081T -FF/FI INST ALLA TION INST ALLA TION UNP ACKING AND INSPECTION Inspect the shipping container . If it is damaged, cont act the shipper immediately for instructions. Save the box. If there is no apparent damage, unp ack the container . Be sure all items shown on the packing list are present.
9.
10.
11.
1 1 2 2.
13.
1 1 4 4 FM A 5/6/04 J. FLOCK 5/6/04 J. FLOCK .140 ±. 0 0 5 1.30 ±. 0 2 .650 ±. 0 15 .125 2.180 ±. 0 0 5 2.56 ±. 0 2 .120 ±. 0 15 Ø .125 R .25 5-6-04 89 33 4 4 FINISH:SILKSCREEN BLACK EPOX Y PAINT (BAKED). 5081-T-FI-67 MODEL 5081-T-FI HARDNES S BRINEL L 190 .
1 1 5 5 5/6/04 J. FLOCK 5/6/04 J. FLOCK 70°C MA X CLASS III, DI V 1 CLASS II, DIV 1, GPS E-G CLASS I, DIV 1, GPS B-D HAZARDOU S AREA 3 R T D C O M 6 R C V S H L D 7 R C V C O M 8 R C V 9 D R V S H L .
1 1 6 6 FM B La, Lt OR Lo; OR Lc/Rc (La/Ra OR Lo/Ro) AND Li/Ri (L a/Ra OR Lo /Ro) 1. INSTALLATION SHOULD BE IN ACCORDANC E W ITH ANSI /ISA RP12.06 "INSTALLATION OF INTRIN SICALLY SAFE SYSTEMS FOR.
1 1 7 7 THIS DOCUMENT IS CERTIFIED BY REV REV REV REV REV REV REVISIONS NOT PERMITTED W/O AGENCY APPROVAL DIMENSIONS ARE IN INC HES REMOVE BURRS & SHA RP EDGES .
1 1 8 8 Rosemount Analytical, Uniloc Divisi on 2400 Barran ca Pkwy Irvine, CA 92 606 UNLESS OTHERWISE SPECIFIED DIMENSIO NS ARE IN INCHES REMOVE BURRS & S HARP EDGES .02 0 MAX MACHINED F ILLET RADII .0 20 MAX FINISH MATERIAL PROJECT ENGR A PVD SIZE DWG NO.
1 1 9 9 9241514-01/ A DIME NSIONS ARE IN I NCHES REMOVE BURRS & SHA RP EDGES .020 MAX MACHINED FILLE T RADII .020 MA X - ITEM PART NO DESCRIPTION QT Y SCALE SIZE REV DW G NO BILL OF M AT ERIAL MAT.
2 2 0 0 Rosemount Analytical, Uniloc Division 2400 Ba rranca Pk wy Irvine, CA 9 2 606 UNLESS OTHERWISE SPECIFIED DIMENSIONS ARE IN INCHES REMOVE BURRS & SHARP EDGES .020 MAX MACHINED FILLET RADII .020 MAX FINISH MATER IAL PROJECT ENGR APV D SIZE DWG NO.
2 2 1 1 OVERVIEW This section gives general procedures for routine maintenance of the 5081-T transmitter . The transmitter needs almost no routine maintenance. TRANSMITTER MAINTENANCE Periodically clean the transmitter window with household ammonia or glass cleaner .
2 2 2 2 Replacement Part s for Model 5081-T -FF/FI T ransmitter Location in Shipping drawing PN Description Weight 1 23992-07 PCB stack consisting of the CPU (p art 3) and analog (part 4) boards, 1 lb/0.
2 2 3 3 QUICK-ST ART 1. On the Remote, press PROG , NEXT , ENTER. 2. Use the arrow buttons to select COnduc (conductivity), nAOH (Sodium Hydroxide 0-15%), HCL (Hydrochloric Acid 0-16%), H2SO4L (Sulfuric Acid 0-30%), H2SO4H (Sulfuric Acid 95-99.99%), or CuSt (custom curve) mode.
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Emerson 5081T-FF/FI è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Emerson 5081T-FF/FI - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Emerson 5081T-FF/FI imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Emerson 5081T-FF/FI ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Emerson 5081T-FF/FI, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Emerson 5081T-FF/FI.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Emerson 5081T-FF/FI. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Emerson 5081T-FF/FI insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.