Before using the instru ment, be sure to read th is manual thoroughly . Also, read the manuals of other instruments in this system. Keep the manual where it is easily accessible.
Please Note 1. In no event will Canon be liable for direct or indirect consequential dam age arising out of the use of this product. Canon will not be liable for loss of image data due to any reason. 2. Please have this product inst alled by Canon repr esentative or distributor .
Contents 1. Overview ................. ..................................................................... 1 1.1 About MLT (F) ............................... ........................................................ 1 1.2 Available Functions of ML T (F) .
.
Not ation Conventions Notation 1 In this manual, the name of bu tton is enclosed with a pair of brackets [ ], and the buttons that are selected in succession are connected with > to indicate the order .
.
– 1 – 1. Overview 1.1 About ML T (F) NOTE: ML T (F) is supported by CXDI System Software V ersion 4.3 or later . The CXDI Image Processing Sof tware ML T (F) (called “ML T (F)” in this manual).
1.Overview – 2 – 1.2 A vailable Functions of ML T (F) This manual describes the operating procedures of the following three image pro cessing func- tions that are available with ML T (F). For deta ils and information on other image processing functions, please refer to Operation Manual and Setup Guide.
Continued on the following page >> – 3 – 2. Edit Exposure Mode Settings The examples used in this chapter are the sc reens of CXDI System So ft ware versi on 6.3. 2.1 Procedures to Edit Exposure Mode 1 Display the Edit Ex posure Mode screen On the Exposure screen, touch the followings: [SYSTEM] > [EDIT EXPOSURE MODE] .
2.Edit Exposure Mode Settings – 4 – 4 Select the Exposure Mode Make sure that [EDIT P ARA.] at the top-left is highlighted, and select the desired Expo- sure Mode. The EDIT P ARA. screen appears. EDIT P ARA. Screen (first page) 5 Display the next p age T ouch [NEXT P AGE] .
Continued on the following page >> 2.Edit Exposure M ode Settings – 5 – 6 Select ML T (F) FREQUENCY If ML T (F) FREQUENCY is not curren tly selected (in this case, the DWT ENHANCE. , DWT DEP , and DWT NOISE R. tabs are not disp layed), touch the Frequency Pro- cessing Mode button to open the window , then touch [ML T (F) FRE QUECY] .
2.Edit Exposure Mode Settings – 6 – 9 Save the p arameters T ouch [OK] . The window disp layi ng the following message appears: Save the chang ed paramete r . Then, touch [OK] . When Password is r equire d If the password is specified, a different messa ge is displayed in the window as follow s: Administrator’s authority is needed.
Continued on the following page >> 2.Edit Exposure M ode Settings – 7 – 2.2 Setting DWT ENHANCE. T ab 1 Set the DWT Edge Enhancement effective T ouch [EDGE ENHANCE.
2.Edit Exposure Mode Settings – 8 – 3 Display the FREQUENCY BASE window T ouch the FREQUENCY BASE but ton to open the window . FREQUENCY BASE Window The types of Frequency Bases are classified int.
– 9 – 2.Edit Exposure M ode Settings 5 Display the DENSITY BASE window T ouch the DENSITY BASE button to open the window . DENSITY BASE Window If granularity is too coarse on the Low-Density Field.
2.Edit Exposure Mode Settings – 10 – 2.3 Setting DWT DEP T ab 2.3.1 Setting DWT DEP for High-Density Field 1 Set the DWT DEP for the High-Density Field effective T ouch [HIGH DENS.
Continued on the following page >> 2.Edit Exposure M ode Settings – 1 1 – 4 Set the DEP EFFECT p arameter of the High-Density Field T ouch the or button, and set the value within the range between 0 to 10. The value on the left is the setting saved p rev iously , and the value on th e right is the cur- rent setting.
2.Edit Exposure Mode Settings – 12 – 2 Set the ST ARTING DENSITY parameter of the Low-Density Fie ld T ouch the or button, and set the value within the range between 0 to 30. If the paramet ers of the High-Densi ty Field are ef fective, this value cannot be set lower than the value of ST ARTING DENSITY of the High-Density Field ( → P.
– 13 – 2.Edit Exposure M ode Settings 2.4 Setting DWT NOISE R. T ab 1 Set the DWT Noise Reduction effective T ouch [NOISE REDUCTION] to display the parameters. The appearance of the button switches over between the hig hlig ht (in whit e outline characters ag ainst a colored back- ground) and normal display each time you touch th e button.
– 14 – 3. QA Mode Settings The examples used in this chapter are the sc reens of CXDI System So ft ware versi on 6.3. 3.1 Procedures to Change Parameters Image processing of the exposure images wi.
Continued on the following page >> 3.QA Mode Se ttings – 15 – 2 Select ML T (F) FREQUENCY If ML T (F) FREQUENCY is not curren tly selected (in this case, the DWT ENHANCE. , DWT DEP , and DWT NOISE R. tabs are not disp layed), touch the Frequency Pro- cessing Mode button to open the window , then touch [ML T (F) FRE QUECY] .
3.QA Mode Se ttings – 16 – 5 Save the new p arameters T o save new parameters as the default va lu es of the Exposure Mode (except opening from Study List), touch [SA VE P ARAM] . The w indow displaying the foll owing message appears: Are you sure you wa nt to change the default generator p arameters? Then, touch [YES] .
Continued on the following page >> 3.QA Mode Se ttings – 17 – 3.2 Changing DWT ENHANCE. T ab 1 Set the DWT Edge Enhancement effective T ouch [DWT EDGE EN HANCE.
3.QA Mode Se ttings – 18 – 3 Display the MORE window T ouch [MORE] to open the window . MORE Window 4 Display the FREQUENCY BASE window T ouch the FREQUENCY BASE but ton to open th e window . For details of the FREQUENCY BASE parameter , refer to the step 3 in 2.
Continued on the following page >> 3.QA Mode Se ttings – 19 – 6 Display the DENSITY BASE window T ouch the DENSITY BASE button to open the window . For details of the DENSITY BASE param e ter, refer to the step 5 in 2.2 Setting DWT ENHANCE. T ab ( → P.
3.QA Mode Se ttings – 20 – 7 Select the DENSITY BASE p arameter T ouch the desired Density Bas e button. The parameter is selected for the Density Base, and the DENSITY BASE window is closed. 8 Set the settings valid, and close the MORE window T ouch [OK] .
– 21 – 3.QA Mode Se ttings . Switching the d isplay of the graph T ouch the graph area to switch over the Freque ncy Base Graph (bar graph) and Density Base graph (line graph). Actual processing shall not be performed unless the DWT Edge Enhancement is effective, although the graph i s displayed on the screen.
3.QA Mode Se ttings – 22 – 3.3 Changing DWT DEP T ab 3.3.1 Changing DWT DEP for High-Density Field 1 Set the DWT DEP for the High-Density Field effective T ouch [HIGH DENS.
– 23 – 3.QA Mode Se ttings 3 Display the MORE window T ouch [MORE] to open the window . MORE Window 4 Set the HIGH WEIGHT p arameter T ouch the or button, and set the value within the range between 0 to 20. The value on the left is the setting saved p rev iously , and the value on th e right is the cur- rent setting.
3.QA Mode Se ttings – 24 – 3.3.2 Changing DWT DEP for Low-Density Field 1 Set the DWT DEP for the Low-Density Field effective T ouch [LOW DENS.] to display the parameters.
– 25 – 3.QA Mode Se ttings 3 Display the MORE window T ouch [MORE] to open the window . MORE Window 4 Set the LOW WEIGHT p arameter T ouch the or button, and set the value within the range between 0 to 20. The value on the left is the setting saved p rev iously , and the value on th e right is the cur- rent setting.
3.QA Mode Se ttings – 26 – 3.4 Changing DWT Noise R. T ab 1 Set the DWT Noise Reduction effective T ouch [NOISE REDUCTION] to display the parameters. The appearance of the button switches over between the hig hlig ht (in whit e outline characters ag ainst a colored back- ground) and normal display each time you touch th e button.
Continued on the following page >> – 27 – 4. QA Parameters When ML T(F) frequency processing is select ed, the QA parameters (%IP_P ARAM%) in Anno- tation are displayed in the format below . n stands for the numeric value, ch aracter or symbol displayed on Annotation .
4.QA Parameters – 28 – ST ARING DENSITY (High-D ensit y and Low-Density Fields) DEP WEIGHT (High- Density and Low-Density Fields) 5 DWT Noise Reduction The parameters are displayed as follows: Nn ( EFFECT ) , n ( PROCESSING UPPER LIMIT ) If the DWT Noise Reduction is not effective, nothing is displayed.
– 29 – 4.QA Parameters 9 Suppress Grid Processing One of the following parameters is displayed: Z0 (No grid stripe is detected), Z1 (Grid stripes are suppressed), Z* (Processing is not determined yet) If the Grid Reduction Processing is no t effective, nothing is displayed.
.
.
.
.
CANON INC. Medical Equipment Group 30-2, Shimomaruko 3-chome, Ohta-ku, T okyo, Japan T elephone: (81)-3-3758-2 1 1 1 CANON U.S.A., INC. CANON MEDICAL SYSTEMS 15955 Alton Parkway , Irvine, CA 92618-3616, U.S.A. T elephone: (1)-949-753-4 160 CANON EUROP A N.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Canon MLT (F) è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Canon MLT (F) - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Canon MLT (F) imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Canon MLT (F) ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Canon MLT (F), ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Canon MLT (F).
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Canon MLT (F). Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Canon MLT (F) insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.