INFRARED THERMOMETER

Instructions and Guarantee

Salter Infrared Thermometer — INFRARED THERMOMETER diagram (page 1)
Fig. 1

This equipment needs to be installed and put into service in accordance

with the information provided in this manual

TE-250-EU

Page 2 of the Salter Infrared Thermometer user manual — IMPORTANT INFORMATION BEFORE USE
Pagina 2 · clicca per ingrandire
Pagina 2

INTRODUCTION

Utilising infrared technology, this thermometer takes temperatures in seconds by measuring heat generated by the surface skin of the forehead or other objects. This product conforms to the provisions of the EC directive MDD(93/42/EEC). Its advantages include: 1. 6 in 1 functions Human Body/Object/Night Mode Option/Fever Alarm/30 User Memory/Backlight 2. Night Mode Option Switch to Night mode to reduce the interference of buzzer during your baby sleeping. 3. Red LED Light for Fever Alarm This device has the LED light to remind user their readings are over 38°C and plus the beep sound only in Human Body Mode. 4. 30 User Memory 5. Illuminated Backlight Display 6. C/°F Switch-able Function 7. One-second reading 8. Auto power off for power saving 9. Low-battery Indicator 10. Large LCD Display 11. Economic design and convenience: This is a “Non-contact” medical thermometer that enables temperature readings, designed for sanitary conditions, cleanliness, and convenience. Simply move the thermometer close to the subject’s forehead or object at the distance indicated by the device. 12. Instant Measurement By using our unique technology, users can get their precise body temperature instantly and accurate.

IMPORTANT INFORMATION BEFORE USE

When using this product, please be sure to follow all the notes listed below. Any action against these notices may cause injury or affect the accuracy.

Page 3 of the Salter Infrared Thermometer user manual — BATTERY INSTALLATION
Pagina 3 · clicca per ingrandire
Pagina 3

1. Do not disassemble, repair, or remodel the thermometer. 2. Be sure to clean the thermometer lens each time after usage. 3. Avoid direct finger contact with the lens. 4. No modification of this equipment is allowed. 5. It is recommended that user may take 3 temperatures. If they are different, use the highest reading. 6. Do not expose the thermometer to extreme temperature, very high humidity, or direct sunlight. 7. Avoid extreme shock or dropping the device. 8. Before the measurement, users and thermometer should stay in steady state room condition for at least 30 minutes. 9. Avoid measuring temperature for 30 minutes after exercise, bathing, or returning from outdoors. 10. To protect the environment, dispose of empty batteries at appropriate collection sites according to national or local regulations. 11. It is ill-advised to disassemble the thermometer. 12. Please use the thermometer solely for its intended purpose. 13. Carefully hold the device when in use to avoid dropping the device. 14. Allow one minute between successive measurements as slight variations may occur if measurements are taken over a short period of time. Use average temperatures instead. 15. There are no absolute body temperature standards. Keep reliable records of your personal temperature to serve as a reference for judging a fever. 16. Under any circumstances, the temperature taking result is ONLY for reference. Before taking any medical action, please consult your doctor.

BATTERY INSTALLATION

Low battery warning When the battery power becomes low, the low battery symbol will appear on the display. The thermometer can still be used during this time, but the batteries should be replaced as soon as possible. If the batteries run out completely, “Lo” will be displayed along with the low battery symbol. In this case, the batteries will need to be replaced before using the thermometer again.

Page 4 of the Salter Infrared Thermometer user manual — PRODUCT IDENTIFICATION
Pagina 4 · clicca per ingrandire
Pagina 4

Replacing the Battery 1. Gently slide the battery cover back. 2. Carefully remove the old batteries and properly discard. 3. Insert new batteries (Two 1.5V alkaline AAA Size) according to the proper polarity. 4. Slide the battery cover back on.

Salter Infrared Thermometer — PRODUCT IDENTIFICATION diagram (page 4)
Fig. 2
  • Note
  • It is recommended to remove the batteries if the unit will not be used for an extended period of time.
  • Please store batteries away from small children and heat.
  • Do not use rechargeable batteries.
  • Dispose of used batteries in accordance with the applicable legal regulations.
  • Never dispose of batteries in the normal household waste.

PRODUCT IDENTIFICATION

Salter Infrared Thermometer — PRODUCT IDENTIFICATION diagram (page 4)
Fig. 3

Red LED Light for Fever Alarm (Only in Human Mode)

Power / scan & measure Button

Show 6 more labels / short text on this page
LCD display · Measurement sensor · Buzzer · Memory Recall button · Human/Object Button · Battery compartment lid
Page 5 of the Salter Infrared Thermometer user manual — SWITCHING BETWEEN CELCIUS OR FAHRENHEIT
Pagina 5 · clicca per ingrandire
Pagina 5

DESCRIPTION OF LCD DISPLAY

Night mode Human scan mode

Smiling face Measurement

Celcius scale Fahrenheit scale

Measurement ≥38.0°C(≥ 100.4°F)

SWITCHING BETWEEN CELCIUS OR FAHRENHEIT

Your thermometer can display results in either degrees Celsius (°C) or degrees Fahrenheit (°F). To switch between Celsius and Fahrenheit, while the unit is on press and hold both the Human/Object button and the Memory button for approximately 3 seconds. This will change the mode to either °C or °F. Once the thermometer beeps, at which time the mode has been changed and you may release the buttons.

Switching between 4 Kinds of Scan Mode 1. Under power on status, you can press the Human/Object button to switch different scan mode. There are 4 kinds of mode which including Human, Object, Human/Night, and Object/Night mode (in order). 2. The beep sounds will be closed when your choice in Human/Night mode, and Object/Night mode, and the Moon symbol will appears on the LCD in both Night mode.

Show 10 more labels / short text on this page
Scan in progress · Memory symbol · and sets of · records · Object scan mode · Temperature in degrees · No Fever · Frowning face · Low battery warning · Fever
Page 6 of the Salter Infrared Thermometer user manual — TIPS FOR MEASURING HUMAN TEMPERATURE
Pagina 6 · clicca per ingrandire
Pagina 6

Object mode & night mode Note Each press will comes with a beep sound to ensure the setting is activated. (Except both Night modes)

Human mode Object mode Human mode & night mode

TIPS FOR MEASURING HUMAN TEMPERATURE

  • Bear in mind that the thermometer needs to have been in the room in which the measurement is taken for at least 30 minutes before use.
  • Attempting to take temperature readings from sites on the body other than the forehead may produce inaccurate results.
  • The patient should remain still while the reading is being taken.
  • Infrared forehead temperature readings are equivalent to oral temperature readings. In all of these cases, please consult your doctor.
  • Readings taken while asleep should not be compared directly to readings taken while awake, as body temperature while asleep is typically lower.
  • Do not take body temperature readings within 30 minutes of being outdoors, exercising or bathing.
Salter Infrared Thermometer — TIPS FOR MEASURING HUMAN TEMPERATURE diagram (page 6)
Fig. 4
1. Press the Power button to turn the thermometer on. The unit will run a self-test and all symbols on the display will momentarily appear. 2. Ensure that the thermometer is in Human mode; the Human symbol will be on the display. To alternate between modes press and release the Human/Object Mode button until you see the desired measurement symbol on the display. 3. Press and hold the “Scan & measure button”, aiming the
Page 7 of the Salter Infrared Thermometer user manual — MEMORY FUNCTION
Pagina 7 · clicca per ingrandire
Pagina 7

Measurement Sensor at the person’s forehead, holding the thermometer approximately 0.5~3 cm. But don’t touch the forehead. 4. Release the “Scan & measure button”. In about a second, you will hear a short beep means this temperature reading has been completed and accompany with a back-light. 5. If the temperature measurement is below 38°C, a “Smiling Face” will be appear next to the reading. If the reading is 38°C or above, a “Frowning Face” will be displayed and the RED LED light up. Fever Alarm - Only in Human Mode and Human/Night Mode 6. After about 30 seconds after use, the thermometer will automatically beep and shut off.

0.5~3 cm

MEASURING OBJECT TEMPERATURE

1. Press the Power button to turn the thermometer on. All symbols on the display will momentarily appear. 2. Ensure that the thermometer is in Object mode; the Object symbol will be on the display. To alternate between modes press and release the Human/ Object Mode button until you see the desired measurement symbol on the display. 3. Press and hold the “Scan & measure button”, aiming the Measurement Sensor at the object, holding the thermometer approximately 0.5~3 cm. 4. Release the “Scan & measure button” and the temperature reading will be displayed. 5. After about 30 seconds after use, the thermometer will automatically beep and shut off.

0.5~3 cm

MEMORY FUNCTION

You can recall up to 30 stored measurements in memory to share with your doctor or trained healthcare professional.

Page 8 of the Salter Infrared Thermometer user manual — MEASURING OBJECT TEMPERATURE
Pagina 8 · clicca per ingrandire
Pagina 8

Memory Recall button. 1. When the device is on, press once briefly on the “MEM button”, then pass it again to show the last measurement accompanied by symbol. 2. The symbol or symbol will appear with each measurement stored in memory to indicate whether a person or object temperature was taken. 3. Each press of the same button recalls a previous measurement, so then all the way to . Continue to press the Memory Recall button button to view the next previously stored measurement. 4. Any new measurement will be recorded and the oldest memory deleted without you having to do anything.

Memory Deletion 1. Under power on status, you may keep press MEM Button for more than 3 seconds to delete all the readings. 2. You will see “--” on display and “sound four short beeps”, that mean all the memories are cleared. 3. Automatically on the 31th measurement: when the 30 memories have been used up, any new measurement will be recorded with and the oldest memory deleted without you having to do anything.

All the readings will be cleared no matter record in Human mode or Object mode.

Salter Infrared Thermometer — MEASURING OBJECT TEMPERATURE diagram (page 8)
Fig. 5

Lens/ Measurement Sensor Gently clean with an alcohol swab. Do not use water to wash the thermometer lens directly.

Thermometer Clean with a soft, dry cloth. Do not use water to rinse the device.

Page 9 of the Salter Infrared Thermometer user manual — ERROR CODES
Pagina 9 · clicca per ingrandire
Pagina 9

APPLIED STANDARDS

  • This product conforms to the provisions of the EC directive MDD(93/ 42/ EEC). The following standards apply to design and/or manufacture of the products:
  • ASTM E1965-98 Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature.
  • ISO 14971 Medical devices-Application of risk management to medical devices. The Classification according to IEC/EN60601-1 sub-clauses 5:

ERROR CODES

When a malfunction or incorrect temperature measurement occurs, an error message will appear as described below.

LCD DisplayCauseSolution The temperature measured is higher than Human thermometer mode: 43°C (109.4°F) Object temperature mode: 100°C (212.0°F)Operate the thermometer only between the specifeid temperature ranges. If necessary, clean the sensor tip. In the event of a repeated error message, contact your retailer or Customer Services. The temperature measured is lower than Human thermometer mode: 34°C (93.2°F) Object temperature mode: 0°C (32.0°F) The operating temperature is not in the range 16°C~40°C (60.8°F~104°F)Operate the thermometer only between the specifeid temperature ranges.
Page 10 of the Salter Infrared Thermometer user manual — EXPLANATION OF SYMBOLS
Pagina 10 · clicca per ingrandire
Pagina 10

TECHNICAL SPECIFICATION

  • Measuring range: Human Body: 34°C ~43°C (93.2°F ~ 109.4°F) Object: 0°C~100°C (32.0°F~212.0°F)
  • Calibration Accuracy: Human Body: ±0.2°C(±0.4°F): from 34°C ~43°C (93.2°F ~ 109.4°F) Object: <LESS THAN EQUAL TO>40°C : ± 2°C > 40°C : ± 5%
  • Power supply: 2 x 1.5V AAA size alkaline batteries
  • Dimensions: approx. 128.5mm×48.83mm×38.85mm (L×W×H)
  • Weight : approx. 80g (with batteries)
  • Operation Distance : 0.5~3 cm

EXPLANATION OF SYMBOLS

SGS No.

Type BF equipment (Sensor)

Disposal information: Should you wish to dispose of the article, do so in accordance with current regulations. Details are available from your local authority

Manufacturer’s name and address

Refer to instructions · European Authorized Representative

Page 11 of the Salter Infrared Thermometer user manual — EMC TABLES
Pagina 11 · clicca per ingrandire
Pagina 11

EMC TABLES

Guidance and manufacturer’s declaration-electromagnetic emissions The Ear Thermometer is intended for use in the electromagnetic environment specifeid below. The customer or the user of the Thermometer should assure that it is used in such an environment. Emissions testComplianceElectromagnetic environment guidance RF emissions CISPR 11Group 1The Thermometer uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emissions CISPR 11Class BThe Thermometer is suitable for use in all establishments other that domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2Not applicable Voltage fulctuations Flicker emissions IEC 61000-3-3Not applicable
Page 12 of the Salter Infrared Thermometer user manual — EMC TABLES
Pagina 12 · clicca per ingrandire
Pagina 12
Guidance and manufacturer’s declaration-electromagnetic immunity The Thermometer is intended for use in the electromagnetic environment specifeid below. The customer or the user of the Thermometer should assure that it is used in such an environment. Immunity testIEC 60601 test levelCompliance levelElectromagnetic environmentguidance Electrostatic discharge (ESD) IEC 61000-4-2±6kV contact ±8kV air±6kV contact ±8kV airFloor should be wood, concrete or ceramic tile. If folors are covered with shythetic material, the relative humidity should be at least 30%. Power frequency (50/60 Hz) magnetic feild IEC 61000-4-83 A/m3A/mPower frequency magnetic feilds should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Page 13 of the Salter Infrared Thermometer user manual — EMC TABLES
Pagina 13 · clicca per ingrandire
Pagina 13

Guidance and manufacturer’s declaration-electromagnetic immunity

The Thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the Thermometer should assure that it is used in such an environment.

Immunity test IEC 60601 test level

Compliance level Electromagnetic environment-guidance

Conducted RF IEC 61000-4-6

3Vrms 150kHz to 80 MHz

Portable and mobile RF communications equipment should be used no closer to any part of the Ear Thermometer, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance. d=1.2 d=1.2 80 MHz to 800 MHZ d=2.3 800 MHz to 2,5 Ghz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,” should be less than the compliance level in each frequency range.” Interference may occur in the vicinity of equipment marked with the following symbol:

Radiated RF IEC 61000-4-3

3V/m 80MHz to 2.5 GHz

Field strengths from fixed transmitters, such as base stations from radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretic call with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Ear Thermometer is used exceeds the applicable RF compliance level above, the Ear Thermometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Ear Thermometer. Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.

3 Vrms · 3 V/m

Page 14 of the Salter Infrared Thermometer user manual — EMC TABLES
Pagina 14 · clicca per ingrandire
Pagina 14

Recommended separation distances between portable and mobile RF communications equipment and the Thermometer

The Thermometer is intended for use in an electromagnetic environment in which radiated RF distances are controlled. The customer or the user of the Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Thermometer as recommended below according to the maximum output power of the communications equipment.

Rated maximum output power of

Separation distance according to frequency of transmitter / m

150 kHz to 80 MHZ d=1.2

80 MHz to 800 MHZ d=1.2

800 MHz to 2.5 GHz d=2.3

0.01 0.12 0.12 0.23

0.1 0.38 0.38 0.73

1 1.2 1.2 2.3

10 3.8 3.8 7.3

100 12 12 23

2 YEAR GUARANTEE FKA Brands Ltd guarantees this product from defect in material and workmanship for a period of 2 years from the date of purchase, except as noted below. This FKA Brands Ltd product guarantee does not cover damage caused by misuse or abuse; accident; the attachment of any unauthorised accessory; alteration to the product; or any other conditions whatsoever that are beyond the control of FKA Brands Ltd. This guarantee is effective only if the product is purchased and operated in the UK / EU. A product that requires modification or adaptation to enable it to operate in any country other than the country for which it was designed, manufactured, approved and / or authorised, or repair of products damaged by these modifications is not covered under this guarantee. FKA Brands Ltd shall not be responsible for any type of incidental, consequential or special damages. To obtain guarantee service on your product, return the product post-paid to your local service centre along with your dated sales receipt (as proof of purchase). Upon receipt, FKA Brands Ltd will repair or replace, as appropriate, your product and return it to you, post-paid. Guarantee is solely through HoMedics Service Centre. Service of this product by anyone other than the Salter Service Centre voids the guarantee. This guarantee does not affect your statutory rights. For your local Salter Service Centre, go to www.salterhousewares.co.uk/servicecentres

transmitter / W

Fai una domanda su Salter Infrared Thermometer

I possessori del prodotto e il nostro team di solito rispondono entro 24 ore, e la tua domanda sarà utile anche ad altri.

Add your email to get notified when it’s answered — no spam, ever.

0 questions
Nessuna domanda ancora per il dispositivo Salter Infrared Thermometer

Invia la prima domanda: risponderanno altri possessori o il nostro team.

Documento originale

Salter Infrared Thermometer manuale in PDF

80 pagine·1.4 MB·Termometro·PDF
Clicca su qualsiasi pagina in alto per aprirla qui
Pagina 1 / …
100%
Caricamento pagine…