No. MIIXR0002EAC DICOM CONFORMANCE STATEMENT MEDIA STORAGE FOR TOSHIBA DIGITAL FLUOROGRAPHY SYSTEM MODEL DFP-2000A with XIDF-037A and XIDF-038A or SRS-1000A (MIIXR0002EAC) 2001.
No. MIIXR0002EAC IMPORTANT! (1) No part of this manual may be copied or reprinted, in whole or in part, without written permission. (2) The contents of this manual are subject to change without prior notice and without our legal obligation.
No. MIIXR0002EAC Table of Contents Page 1 INTRODUCT ION ................................................................................................................ 1 1.1 R EFERENCES ................................................................
No. MIIXR0002EAC 8.1.1 Media Storage Directory IOD Modules ................................................................ 9 8.1.2 Media Storage XA IOD Modules .......................................................................... 9 8.1.3 XA IOD Modules .
No. MIIXR0002EAC 1 Introduction This document is a DICOM Conformance Statement for Toshiba's Digital Fluorography Systems. It is intended to provide the reader with the knowledge of how to integrate this product within DICOM compliant equipments.
No. MIIXR0002EAC 1.3 Acronyms, Abbreviations and Sy mbols • ACC American College of Cardiology • ACR American College of Radiology • ASCII American Standard Code for Information Interchange • .
No. MIIXR0002EAC 2 Implement ation Model 2.1 Application Data Flow Diagram Transfer of X-r a y Examinat ion DICOM Recording CD-R Medium Review of X-r a y Examinat ion DICOM Reader Legend Applicati on Entity Local Real-W orld Activi ty File-set Creator (FS C) File-set Reader (FS R) Local Relationship Figure 1 2.
No. MIIXR0002EAC 2.2.2 DICOM Reader The DICOM Reader supports the following tasks: • Reads the DICOMDIR file that represent s the contents of t he data as recorded. • Displays the ordered list of patient and image identifying information or icon images.
No. MIIXR0002EAC 2.3.2 Operation 2.3.2.1 Transfer of X-ray Examination The operation for manual transfer of X-ray examination is described below: STEP-1: Select the images or the patient to be transferred. STEP-2: Request transfer. STEP-3: Insert the blank CD-R media.
No. MIIXR0002EAC 3 AE Sp ecifications 3.1 DICOM Recording Specification The DICOM Recording AE provides Standard Conformance to the DICOM Media Storage Service, File Format Class and t he Media Storage Application Profile.
No. MIIXR0002EAC 3.2.1 File Meta Information for the Application Entity The Application Entity Title is not relev ant because this DICOM Reader AE only reads CD-R media. 3.2.2 Real-World Activity 3.2.2.1 Real-World Activity - Review of X-ray Examination DICOM Reader AE acts as a FSR when chosen by the operator.
No. MIIXR0002EAC 5 Extensions, S pecializations, and Privatizations of SOP Classes and T ransfer Synt axes Not applicable to this product. 6 Configuration For the Digital Fluorography System, the configuration can be set.
No. MIIXR0002EAC 8 X-Ray Angiographic Information Object Definition 8.1 Entity Module Definitions The information modules for the X-Ray Angiographic devices are defined below. 8.1.1 Media Storage Directory IOD Modules Table 5 Object Module / key Reference Usage 1 DICOM File Meta Information 8.
No. MIIXR0002EAC 8.1.3 XA IOD Modules Table 7 Information Entity Module Reference Usage 1 Patient Patient Module 8.5.1 M Study General Study Module 8.5.2 M Study Patient Study Module 8.5.3 U Series General Series Module 8.5.4 M Equipment General Equi pment Module 8.
No. MIIXR0002EAC 8.2 DICOM File Meta Information 8.2.1 DICOM File Meta Information of Directory IOD Table 8 Attribute Name Tag Type Attribute Description File Preamble No Tag 1 Alw ays set DICOM Prefi.
No. MIIXR0002EAC 8.3.2 Directory Information Module Table 11 Attribute Name Tag Type Attribute Description Offset of the First Directory Record of the Root Directory Entity (0004, 1200) 1 Always set O.
No. MIIXR0002EAC 8.4.2 STUDY keys Table 13 Attribute Name Tag T ype Attribute Description Study Date (0008, 0020) 1 Alway s set Study Time (0008, 0030) 1 Always set Study Description (0008, 1030) 2 Le.
No. MIIXR0002EAC 8.4.4 IMAGE keys Table 15 Attribute Name Tag Type A ttribute Description Instance Number (0020, 0013) 1 Alway s set Image Type (0008, 0008) 1 Alway s set Image Date (0008, 0023) 3 Alw.
No. MIIXR0002EAC 8.5 XA Information Object Definisions 8.5.1 Patient Module Table 16 Attribute Name Tag Type Attribute Description Patient's Name (0010, 0010) 2 Always set except for urgent patie.
No. MIIXR0002EAC 8.5.4 General Series Module Table 19 Attribute Name Tag Type Attribute Description Modality (0008, 0060) 1 Always set ("XA") Series Instance UID (0020, 000E) 1 Alway s set S.
No. MIIXR0002EAC 8.5.7 Image Pixel Module Table 22 Attribute Name Tag Type Attribute Description Samples per Pixel (0028, 0002) 1 Always set (0x0001) Photometric Interpretation (0028, 0004) 1 Always s.
No. MIIXR0002EAC 8.5.11 X-ray Image Module Table 26 Attribute Name Tag Type Attribute Description Frame Increment Pointer (0028, 0009) 1C Always set (0x00181063) Image Type (0008, 0008) 1 Alw ays set .
No. MIIXR0002EAC 8.5.13 XA Positioner Module Table 28 Attribute Name Tag Type Attribute Description Distance Source to Detector (0018, 1110) 3 Always set [mm] Estimated Radiographic Magnification Fact.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Toshiba XIDF-037A è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Toshiba XIDF-037A - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Toshiba XIDF-037A imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Toshiba XIDF-037A ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Toshiba XIDF-037A, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Toshiba XIDF-037A.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Toshiba XIDF-037A. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Toshiba XIDF-037A insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.