MODEL HPM 4/6 VACUUM GAUGE INSTRUCTION MANUAL TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 2 of 12 Manual Print History The print history shown below lists the p rinting dates of all revisions and a ddenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent u pdates.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 3 of 12 Table of Contents 1.0 QU IC K S TA RT I NS TR U CT ION S....................................................................................................... ............ 4 2.0 SPE CI FIC AT IO NS .
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 4 of 12 1. Unpack unit and insure battery is installed 2. Connect thermocouple tube. 3. Slide switch to DV4 or DV6 position according to tube type 4. Indicated pressure is displayed on LCD display. 5. Slide switch to center/off position when not in use.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 5 of 12 Operating Range .................................................................... DV-4 - 0.01 - 20 Torr (.01 to 20 mbar) ..........................................................................
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 6 of 12 3.1. Receiving Inspection Unpack unit and inspect all items for obvious si gns of damage. Immediately advise Tel edyne Hastings and the carrier of any suspe cted damage.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 7 of 12 Operation of the Hastings gauge sensor (DV-4 or DV-6) is based on a low voltage AC bridge that heats a noble metal thermopile. A change in pressure in the gauge tube alters the molecular collision rate and therefore the thermal conducti on of the gas or gas mixture surrounding the thermopile.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 8 of 12 c. Pressure under range “-0.00 to -0.05” (neg ative display values) indicating that the instrument requir es low-pressure adjustment. Ref er to Section 4. 3. Always place slide switch to the off position when not in use.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 9 of 12 5.1. Calibrat ion in a Vacuum Syste m 1 Install a DV-4 or DV-6 sensor in your vac uum system and ev acuate vacuum chamber to a pressure less than 1.00E-04 T orr. NOTE: Refer to Fig 3 for locations of calib ration pots.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 10 of 12 1 DV-4 low pressure adjust ments: a. Connect instrument gauge cable to a DB-16D Re ference Tub e and position slide switch to DV-4. b. Remove battery compartment cover, and then adj ust DV-4 calibration pot until display reads pressure value listed on reference tube label.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 11 of 12 6.1. Authorized Maintena nce The HPM 4/6 Power Supply/Display does not require maintenance other than routine battery replacement and calibration. Note: A low battery will be indicat ed by “LOBAT” appearing in lower left corner of LCD display.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 12 of 12 7.1. Wa rranty Repair P olicy Hastings Instruments warrants this product for a period of one year from the date o f shipment to be free from defects in material and wor kmanshi p.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Teledyne HPM 4/6 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Teledyne HPM 4/6 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Teledyne HPM 4/6 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Teledyne HPM 4/6 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Teledyne HPM 4/6, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Teledyne HPM 4/6.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Teledyne HPM 4/6. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Teledyne HPM 4/6 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.