Ser vic e Manual DPO 4000 Ser ie s Digital Phosp hor Oscilloscopes 071-1844-00 Th is d ocu men t ap plies to firmw are versio n 1.0 0 and a bove. W a rnin g Th e serv icing instru ctio ns are fo r u se by qu alified per sonne l onl y . T o avoi d per sona l i njur y , do not perfo rm any serv icin g un less yo u are qu alified to do so .
Copyright © T ekt ronix, Inc . All rights re served. Lice nsed software products are owned by T ekt ronix or its subsidiari es or suppliers, a nd are prot ect ed by nat ional copyright laws and int ernat ional trea ty provisions. T ektroni x products are covere d by U.
W arranty 16 T ektroni x warrants tha t the produc t will be free from defe cts in m ate rial s and workmanship for a pe riod of thre e (3) years from the date of origi nal purcha se from an authoriz ed T ektronix distri butor .
.
DPO4000 S erie s S ervic e Manual i T able of Contents Gen eral S afety S umma ry iii ................................... Serv ice S afety S u mmary v .................................... Environmental Considerat ion s vii ............................
T able of Contents ii DPO4000 S erie s S ervic e Manual List of T ables T a bl e 4 - - 1: Extern al in sp ectio n check l ist 4 - - 3 ...................... T a bl e 4 - - 2: Intern al in sp ectio n check l ist 4 - - 4 ...................... List of Figures Figure 2- - 1: DPO4000 serie s block diagram 2- - 1 .
DPO4000 S erie s S ervic e Manual iii General Safety Summar y Rev iew th e follo win g safety p recauti ons t o avoi d inj ury and p revent d amage to this product or any pr oducts connec ted to it. T o avoi d poten tial h azards, use th is pr oduct only as spec ified.
General S afety S um mary iv DPO4000 S erie s S ervic e Manual These te rms may appe ar in this manua l: W ARNING. W arning statements identify conditions or prac tices that could r esult in injury or loss of life . CAUTION. Caution stateme nts identify conditions or prac tices that could r esult in damage to this pr oduct or othe r pr operty .
DPO4000 S erie s S ervic e Manual v Service S afety Summary Only qualifie d personnel should pe rform se rvice pr ocedure s. R ea d this Servic e Safety Summary and t he General Safety Summary before perform ing an y service proce dures. Do Not Service Alone.
Service Saf ety S ummar y vi DPO4000 S erie s S ervic e Manual.
DPO4000 S erie s S ervic e Manual vii Environmental C onsiderations This section pr ovides informa tion about the envir onmental impac t of the product. Observe the f ollowing guidelines when r ecycling a n instrument or c omponent: Equipment Recycli ng.
Environmenta l Considerations viii DPO4000 Series Service Ma nual.
DPO4000 S erie s S ervic e Manual ix Preface This Service manua l provides infor mation to troubleshoot, disasse mble, and repla ce parts on the DPO4000 Series Digital Phosphor Oscilloscope s. Manual Convent ions This manual use s certa in conventions that you should be come fa miliar with befor e attempting servic e.
Preface x DPO4000 S erie s S ervic e Manual.
Operat ing I nfor mati on.
.
DPO4000 S erie s S ervic e Manual 1- 1 Operating Information For informa tion on installing and oper ating your DPO4000 Series Digita l Phosphor Oscilloscope, please r efer to the DPO4000 Se ries Digital Phosphor Oscilloscope s User Manual . This manual i s available on the W eb at www .
Opera ting Informa tion 1- 2 DPO4000 S erie s S ervic e Manual.
Theory of Oper ati on.
.
DPO4000 S erie s S ervic e Manual 2- 1 Theory of Operation This chapter de scribes the ele ctrica l opera tion of the oscilloscope to the module level.
Theo ry of O perati on 2- 2 DPO4000 S erie s S ervic e Manual Power Suppl y The power supply conve rts AC line voltage to +12V to powe r all inter nal circ uits. I/O Bo a rd The I/O boa rd contains USB ports, Compact Flash port, an Ethe rnet conne ctor , T ekLink, XGA V ideo, and a T rigge r Out por t.
Theo ry of O perati on DPO4000 S erie s S ervic e Manual 2- 3 The proce ssor system contains a 440EP P ower PC microproce ssor that controls the entire instr ument. The processor system also conta ins FLAS H ROM, system RA M, an d int erfaces to US B p orts an d a Com pact flas h slo t.
Theo ry of O perati on 2- 4 DPO4000 S erie s S ervic e Manual.
Adjust ment Procedur es.
.
DPO4000 S erie s S ervic e Manual 3- 1 Adjustment Pr ocedures This chapte r contains the fa ctory adjustme nt procedur es for the DPO4000 serie s oscilloscope s. Only qualified per sonnel should perf orm adjustment pr ocedure s. Read t he Service Safe ty Summary and t he General Safety Summary before perform ing an y servi ce procedu res.
Adjustment Proce dures 3- 2 DPO4000 S erie s S ervic e Manual Over view of the Adjust ment Pr ocess Before pe rforming adjustme nt proce dures, you must warm up the oscilloscope at least t en min utes i n an ambi ent tem peratu re between 2 0 ° C and 30 ° C.
Adjustment Proce dures DPO4000 S erie s S ervic e Manual 3- 3 Factor y Adjustm ent Procedur e T o perfor m the fac tory adjustment pr ocedure , do these steps: 1. Connect the oscilloscope to an AC power source . NOTE . Y ou must connect the oscillosc ope and the test equipment to the same AC power cir cuit.
Adjustment Proce dures 3- 4 DPO4000 S erie s S ervic e Manual 8. Push th e Factory lower b ezel bu tton . 9. Push th e OK Do Factory Calibrat ion s ide bezel b utt on to s tart th e adjustment pr ocess. 10. Connect signals from the DC voltage source a s instructed by the oscilloscope display .
Maint enance.
.
DPO4000 S erie s S ervic e Manual 4- 1 Maintenance This section c ontains the infor mation neede d to do periodic a nd corre ctive maintena nce on the osc illoscope as we ll as repa ckaging instruc tions if you need to retur n the oscilloscope to T ektronix for ser vice.
Maintena nce 4- 2 DPO4000 S erie s S ervic e Manual Prev enti ve main tenan ce consi sts o f visu ally i nsp ectin g and clean ing t he oscilloscope and using ge neral c are whe n operating it. How often you do ma intenance depe nds on the sever ity of the envir onment in which the oscilloscope is used.
Maintena nce DPO4000 S erie s S ervic e Manual 4- 3 CAUTION. A void the use of c hemical cle aning agents, w hich might damage the plastics used in this osc illoscope. Use only de ionized water when cleaning the menu buttons or fr ont-pane l buttons. Use a 75% isopr opyl alc ohol solution as a cleane r and rinse with deionized water .
Maintena nce 4- 4 DPO4000 S erie s S ervic e Manual T o access t he in sid e of the o sc illoscope f or inspection a nd cleaning, re fer to the Remo val Procedu r es in this section. Inspec t the internal portions of the oscilloscope f or damage and wea r , using T able 4- - 2 as a guide .
Maintena nce DPO4000 S erie s S ervic e Manual 4- 5 Cleaning Procedure — Interior . T o c lean the osc illoscope inter ior , do the following steps: 1.
Maintena nce 4- 6 DPO4000 S erie s S ervic e Manual Over view of Removal Pr ocedures Refer to Figures 4- - 1 and 4- - 2 for the loc ation of modules tha t you are r emoving or installing. If you are disasse mbling the instrument for clea ning, refe r to the Inspec tion and Cleaning proce dure on page 4 - - 1 for instruc tions.
Maintena nce DPO4000 S erie s S ervic e Manual 4- 7 Rear case Front cover Front foot Figure 4 - 1: Locator for tr im and cabinet rem oval.
Maintena nce 4- 8 DPO4000 S erie s S ervic e Manual Front Panel assembly Main board I/O board assembly Side fans assembly Power supply assembly Inverter board assembly Display assembly Front Case asse.
Maintena nce DPO4000 S erie s S ervic e Manual 4- 9 CAUTION. T o avoid damage to the fr ont panel Standby /On sw itch asse mbly , do not set the Display module assembly on a w ork surface. Sliding the oscilloscope over the edge of the work surface could br eak off the On/Standby switch assembly .
Maintena nce 4- 1 0 DPO4000 Series Servic e Manual Follow these steps to r emove the I /O board. Y ou need to have pre viously remove d the rea r case and powe r supply module. Use T OR X T -15 sc rewdrive r tip. S ee Figure 5- - 4 on page 5 - - 10. 1.
Maintena nce DPO4000 S erie s S ervic e Manual 4- 1 1 Follow these steps to remove the front ca se assembly . Y ou must have previously remove d the R ear Case Asse mbly . S ee Figure 5- - 2 on page 5 - - 6. 1. Pull off 14 knobs (2 la rge , 5 medium, 6 small, and 1 rotar y).
Maintena nce 4- 1 2 DPO4000 Series Servic e Manual T roubl eshooti ng This section c ontains informa tion and proce dures to help you isola te a def ective module in a DPO4000 series oscillosc ope.
Maintena nce DPO4000 S erie s S ervic e Manual 4- 1 3 T roubl eshooti ng Procedur e Figures 4- - 3 through 4- - 5 are trouble shooting procedur e flowcha rts. Use them to troubleshoot a n instrument fa ilure. Begin with Figure 4- - 3. NOTE . Befor e performing the tr oubleshooting pr ocedur e, r emove the instrument cabinet (se e page 4 - - 9 ).
Maintena nce 4- 1 4 DPO4000 Series Servic e Manual ¯ Reconnect loose cables. ¯ Retu rn to top o f Fig ure 4 - - 3. No Ye s Are there +12 volts on the connector pins ? ¯ Measure for +12 V DC ( ± 0.
Maintena nce DPO4000 S erie s S ervic e Manual 4- 1 5 ¯ Push UTILITY . ¯ Push System t o select Self T est. ¯ Push OK Run Self T est. ¯ Determine f ailure from Self T est Results. ¯ Use table to select module t o replace. From the oscilloscope power-up diagnostics message display .
Maintena nce 4- 1 6 DPO4000 Series Servic e Manual Unpacking and Repacki ng Inst ruct ions This section c ontains the infor mation neede d to unpack the osc illoscope and repa ck it for shipment or storage. The oscilloscope a nd its standard accessori es are careful ly p acked at t he facto ry in a shipping ca rton.
Mechanical Par ts List.
.
DPO4000 S erie s S ervic e Manual 5- 1 Mechanical Parts List Thi s chap ter con tain s a li st o f the rep laceabl e m odules for the DPO4000 series oscilloscope .
Mechan ical P arts Li st 5- 2 DPO4000 S erie s S ervic e Manual Modules ca n be servic ed by selec ting one of the f ollowing three options. Contact your local T ektronix ser vice cente r or repre sentative for r epair assista nce. Module Exchange. In some ca ses you may excha nge your module for a rema n- ufac tured module.
Mechan ical P arts Li st DPO4000 S erie s S ervic e Manual 5- 3 The table titled Ma nufactur ers Cross Inde x shows codes, names, and a ddresses of manuf acture rs or vendor s of components liste d in the parts list.
Mechan ical P arts Li st 5- 4 DPO4000 S erie s S ervic e Manual Figures 5- - 1 through 5- - 5 on the following pa ges show the module- level exploded views of the DPO4000 series oscillosc ope. The adj acent pag e is t he li st of compone nts for that e xploded view , indexed by the numbe rs in the figur e.
Mechan ical P arts Li st DPO4000 S erie s S ervic e Manual 5- 5 Replaceable Parts List Fig. & inde x number T ektron ix part number Serial no. effective Serial no.
Mechan ical P arts Li st 5- 6 DPO4000 S erie s S ervic e Manual 1 2 3 4 5 6 7 Flex Circuit J13 13 8 9 10 11 12 13 14 15 16 17 18 19 20 Figure 5 - 2: Exploded view: fr ont panel and displ ay , 4-channe.
Mechan ical P arts Li st DPO4000 S erie s S ervic e Manual 5- 7 Replaceable Parts List Fig. & inde x number T ektron ix part number Serial no. effective Serial no.
Mechan ical P arts Li st 5- 8 DPO4000 S erie s S ervic e Manual 1 2 3 4 5 6 7 Figure 5 - 3: Exploded view: front panel 2-channel instrum ents.
Mechan ical P arts Li st DPO4000 S erie s S ervic e Manual 5- 9 Replaceable Parts List Fig. & inde x number T ektron ix part number Serial no. effective Serial no. dis cont’ d Qty Name & description Mfr. code M fr . par t num ber 5- - 3 Front Panel - - 1 335 - - 1 189- - 00 1 LABEL.
Mechan ical P arts Li st 5- 1 0 DPO4000 Series Servic e Manual Main board I/O board assemly Main board J310 J780 J850 1 2 3 4 5 6 7 8 9 10 2 2 2 Figure 5 - 4: Exploded view: power suppl y and I/O boar.
Mechan ical P arts Li st DPO4000 S erie s S ervic e Manual 5- 1 1 Replaceable Parts List Fig. & inde x number T ektron ix part number Serial no. effective Serial no.
Mechan ical P arts Li st 5- 1 2 DPO4000 Series Servic e Manual I/O board assemly Main board J610 Display assembly to Invert er board assembly Main board Inverter board assembly Main board CN2 CN1 J930.
Mechan ical P arts Li st DPO4000 S erie s S ervic e Manual 5- 1 3 Replaceable Parts List Fig. & inde x number T ektron ix part number Serial no. effective Serial no.
Mechan ical P arts Li st 5- 1 4 DPO4000 Series Servic e Manual Replaceable Parts List Fig. & inde x number T ektron ix part num ber Serial no. effective Serial no. dis cont’ d Qty Name & description Mfr. code M fr . par t num ber ST ANDARD ACC ESSORIES -- -- -- A c c e s s o r i e s 161- - 0104 - - 05 1 CA ASSY ,PWR: 3,1.
Index.
.
DPO4000 S erie s S ervic e Manual Index - 1 Index A acquis ition s ys tem, descr iption, 2- - 2 adjustm ent proce dure, 3 - - 2 factor y , 3 - - 3 adjus tments , 3 - - 1 appli cat ion modul es, ix C c.
Index Index - 2 DPO4000 Series Servic e Manual R rear cas e removal, 4- - 9 remova l procedure s, 4- - 6, 4 - - 8 display m odule, 4- - 1 1 fans, 4- - 10 fro nt case, 4- - 10 front panel , 4- - 1 1 I/.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Tektronix 071-1844-00 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Tektronix 071-1844-00 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Tektronix 071-1844-00 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Tektronix 071-1844-00 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Tektronix 071-1844-00, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Tektronix 071-1844-00.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Tektronix 071-1844-00. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Tektronix 071-1844-00 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.