Manuale d’uso / di manutenzione del prodotto MBR-704GR del fabbricante Sanyo
Vai alla pagina of 24
MBR-704G MBR-704GR Blood Bank Refrigerator INSTRUCTION MANUAL MBR - 704GR.
1 Note: 1. No part of this manual may be reprodu ced in any form without the expressed w ritten permission of SANYO. 2. The contents of this manual are subject to change without notice. 3. Please contact SANYO if any point in this manual is unclear or if there are any inaccuracies.
2 It is imperative that the user compli es with this manual as it contains important safety advice. Items and procedures are described so that you can use this unit correctly and safely. If the precautions advised are followed, this will prevent possible injury to the user and any other person.
3 PRECAUTIONS FOR SAFE OPERATION Do not use the unit outdoors. Current leakage or electric shock may result if the unit is exposed to rain water. Only qualified engineers or serv ice personnel should install the unit. The installation by unqualified personnel may cause electric shock or fire.
4 PRECAUTIONS FOR SAFE OPERATION Disconnect the pow er supply to the unit prior to any repair or maintenance of the unit in order to prevent electric shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance.
5 PRECAUTIONS FOR SAFE OPERATION Fix the shelves securely . Incomplete installation may cause injury or damage. When removing the plug from the power supply outlet, grip the pow er supply plug, not the cord. Pulling the cord may result in electr ic shock or fire by short circuit.
6 CAUTIONS FOR USAGE 1. If the unit is unplugged or the power to the unit is interrupted, do not restart the unit for at least 5 minutes. This protects the compressor. 2. This inner cabinet is refrigerated by the forced ci rculation of cooled air inside the chamber.
7 REFRIGERATOR COMPONENTS 21 <Back side> 20 22 1 5 7 2 12 3 14 13 MBR-704GR 6 9 10 8 18 23 4 15 11 17 16 19.
8 REFRIGERATOR COMPONENTS 1. Front cover: Open this cover when connecting the remote alarm or replacing glow starter for fluorescent lamp. 2. Control panel: Panel can be opened when upper right corner of the lower part cover is pushed. Refer to page 10.
9 REFRIGERATOR COMPONENTS Inside of front cover 1. Glow starter (with cover): This is for the fluorescent light. It is recommended that the glow starter is also replaced when the fluorescent light is replaced. Refer to page 19. 2. Remote alarm terminal: This is used to connect the unit to an ex terior alarm to notify users of any malfunction.
10 REFRIGERATOR COMPONENTS Control panel and keypad 1. Buzzer stop key (BUZZER): To silence the audible alarm, press this key. 2. Temperature display key (UPPER/LOWER): By pressing this key, the displ.
11 INSTALLATION Installation site T o operate this unit properly and to obtain maximum performance, inst all the unit in a location with the following conditions: 1. A location not subjected to direct sunlight Installation in a location subjected to dire ct sunlight may lead to inadequate cooling.
12 INSTALLATION Installation 1. Remove the packaging materials and tapes Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the outside panels are dirty, clean them with a neutral detergent and wipe off with a soft cloth washed in clean water.
13 INSTALLATION Preparation of monitor bottle Prior to the operation of the unit, fill the monitor bottles (upper and lower) with 10% glycerol (or other solution such as ethylene glycol) using the following procedure: 1. Remove the top and bottom drawer (or shelf).
14 START-UP OF UNIT Follow the procedures for the initia l and consequent operations of the unit. 1. Connect the power cord to the dedicated outlet with appropriate rating. 2. On start-up, the alarm buzzer sometimes operates. In this case, stop the buzzer by pressing the buzzer stop key (BUZZER).
15 OPERATING INSTRUCTIONS Control panel Temperature display The unit is set at the factory to obtai n the proper temperature automatically (4 o C ± 1 o C). The high temperature alarm is preset at 6 o C. The low temperature alarm is preset at 2 o C. T a b l e 1 .
16 OPERATING INSTRUCTIONS Temperature sensor calibration procedure The temperature sensors are attached to the monitor bottles. Follow the following instructions for proper calibration (0 adjustments). 1. Remove the monitor bottle cover as shown in the figure.
17 ALARMS & SAFETY FUNCTIONS This unit has the alarms and safety functions shown in Table 2, and also self diagnostic functions. Table 2 Alarms and safety functions Alarm & Safety Situation Indication Buzzer Safety operation High temperature alarm If the upper or lower temperature sensor senses a temperature ≧ 6 o C.
18 ROUTINE MAINTENANCE WARNING Alw ay s disconnect the power supply to the unit prior to any repair or maintenance of the unit in order to prevent electr ic shock or injury. Ensure you do not inhale or consume medication or aerosols from around the unit at the time of maintenance.
19 ROUTINE MAINTENANCE Replacement of fluorescent lamp The fluorescent lamp is placed horizontally in t he upper front of the chamber. Follow the procedure below to replace the lamp. 1. Turn off the fluorescent light switch and disconnect the power supply plug.
20 TROUBLE SHOOTING If the unit malfunctions, check out the following before calling for service. If nothing operates even when sw itched on 1. The power is failed. 2. The circuit breaker is inactivated. 3. The unit is not connected to the power supply.
21 SPECIFICATIONS Name Blood Bank Refrigerator Model MBR-704G MBR-704GR External dimensions W770 x D830 x H1955 (mm) Internal dimensions W650 x D697 x H1500 (mm) Effective capacity 625 L 617 L Exterio.
22 CAUTION Please fill in this form before servicing. Hand over this form to the service engineer to keep for his and your safety. Safety check sheet 1. Refrigerator contents : Risk of infection: Risk of toxicity: Risk from radioactive sources: (List all potentially hazardous materials that have been stored in this unit.
7FB6P101294003 (5 April 2005) Recycl ed paper CEM SANYO Electric Biomedical Co., Ltd. 10-15, Hongo 3-Chome Bunkyoku, T okyo 1 1 3-8434 Jap an Printed in Japan.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Sanyo MBR-704GR è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Sanyo MBR-704GR - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Sanyo MBR-704GR imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Sanyo MBR-704GR ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Sanyo MBR-704GR, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Sanyo MBR-704GR.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Sanyo MBR-704GR. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Sanyo MBR-704GR insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.