Manuale d’uso / di manutenzione del prodotto SUPERCHARGER 60 R3.0 60 R3.0 del fabbricante Precision Trading
Vai alla pagina of 62
Operator Manual, Supercharg er 60 R3.0 Page 1 of 62 TABLE OF CONTENTS 1. INTRODUCTION AND SYSTEM DESCRIPTION .................................................................. 5 2. CONDENSED OPERATING INSTRUCTIONS .....................................
Operator Manual, Supercharg er 60 R3.0 Page 2 of 62 4.20 DS5A - FULL: .................................................................................................................. ........................ 21 4.21 DS5B - DISCH: ...................
Operator Manual, Supercharg er 60 R3.0 Page 3 of 62 8.8 DISCH ARGE VOLTAGE CUT-OFF. ................................................................................................... 38 8.9 FULL DISCHA RGE. ...........................................
Operator Manual, Supercharg er 60 R3.0 Page 4 of 62 TABLE OF FIGURES Figure 1 - Bl ock Diagram ...................................................................................................................................... 7 Figure 2 - Fr ont Panel .
Operator Manual, Supercharg er 60 R3.0 Page 5 of 62 1. INTRODUCTION AND SYSTEM DESCRIPTION Congratulations! You now own the finest batte ry Charger-Analyzer available today.
Operator Manual, Supercharg er 60 R3.0 Page 6 of 62 The TIMER provides all the required timing func tions related to the charge and discharge durations. Digital time and speed selectors provide the external inputs, while the four digit readout of the CLOCK provides the elapsed time display.
Operator Manual, Supercharg er 60 R3.0 Page 7 of 62 Figure 1 - Block Diagram.
Operator Manual, Supercharg er 60 R3.0 Page 8 of 62 2. CONDENSED OPERATING INSTRUCTIONS CAUTION: Disconnect power and batteries befor e performing any internal maintenance. Failure to observe this caution may result in serious damage to the unit and injury to the operator.
Operator Manual, Supercharg er 60 R3.0 Page 9 of 62 2.2 Constant Voltage 2.2.1. Battery Voltage 2.2.1.1. Enter the nominal battery voltage using the cell selector (see chart on page 19). 2.2.1.2. Program the charge (m aximum) current in the Topping Current selector.
Operator Manual, Supercharg er 60 R3.0 Page 10 of 62 2.3 Peak Voltage Charge 2.3.1. Nominal Battery Voltage 2.3.1.1. Enter the nominal battery voltage using the cell selector (see chart on page xxx ). 2.3.1.2. Select transfer from Main to Topping on Peak or stop on Peak.
Operator Manual, Supercharg er 60 R3.0 Page 11 of 62 2.4 Discharge NOTE 1: Do not place batteries on the Temp-Plate during discharge. Due to the normal heating of the battery a false Overtemp may be generated later during the charge process. NOTE 2: Do not attempt to discharge with the voltage mode selector in the Float or Peak voltage modes.
Operator Manual, Supercharg er 60 R3.0 Page 12 of 62 3. SPECIFICATIONS 3.1 CHARGE 3.1.1. CONTROL: SCR, air cooled. 3.1.2. MAXIMUM CURRENT: 50 AMPS. 3.1.3. MINIMUM INCREMENT: 0.1 AMP. 3.1.4. ACCURACY and STABILITY of settings: ±1% of reading, ±0.1 AMP.
Operator Manual, Supercharg er 60 R3.0 Page 13 of 62 3.1.5. MODES: 3.1.5.1. Main and topping for Cons tant Current 3.1.5.2. Constant Voltage 3.1.5.3. Peak Voltage. 3.1.6. TIMER: 3.1.6.1. 1 to 9 hours for MAIN charge time and 1 to 6 0 hours for TOTAL time.
Operator Manual, Supercharg er 60 R3.0 Page 14 of 62 3.1.12.1. 3 AMPS idling plus ½ the value of the charging curre nt; e.g. when charging 40 AMPS the line current is 23 AMPS (Note: nominal values, line voltage and frequency dependent). 3.2 DISCHARGE 3.
Operator Manual, Supercharg er 60 R3.0 Page 15 of 62 3.2.6. TERMINAL VOLTAGE: 3.2.6.1. Programmable, in accordance with the number of cells selected, at the equivalent of 1 VOLT/cell.
Operator Manual, Supercharg er 60 R3.0 Page 16 of 62 3.3.2.4. NOTE: The colon flashes at the rate of one cycle per second regardless of the selected mode.
Operator Manual, Supercharg er 60 R3.0 Page 17 of 62 3.5 DIGITAL PANEL METERS 3.5.1. VOLTMETER: 3.5.1.1. Accuracy (system): ±0.25% of r eading, ±0.1V in the 200V scale, ±0.01V in the 20V scale, ±0.005V in the 2V scale. 3.5.1.2. Input impedance: 1 M-OHM.
Operator Manual, Supercharg er 60 R3.0 Page 18 of 62 4. CONTROLS AND DISPLAYS 4.1 M1 - Ammeter 4.1.1. 0 to 199.9ADC DIGITAL PANEL METER. 4.1.2. Positive sign indicates charge current. 4.1.3. Minus sign indicates discharge current. 4.2 M1 - Voltmeter 4.
Operator Manual, Supercharg er 60 R3.0 Page 19 of 62 4.8 SW4 - CELL SELECTOR: 4.8.1. Two digit selector to program the battery terminal voltage as a function of the number of cells. 4.8.2. Rate: 1.75 volts per cell for charge a nd 1.0 volts per cell for discharge.
Operator Manual, Supercharg er 60 R3.0 Page 20 of 62 Figure 2 - Front Panel.
Operator Manual, Supercharg er 60 R3.0 Page 21 of 62 4.12 DS1A - RESET: 4.12.1. Indicates that the unit is in standby. 4.13 DS1B - CYC END: 4.13.1. Flashing (with pulsating beeper) cycle end. Indicates that the unit has completed its cycle. 4.14 DS2A - DUAL: 4.
Operator Manual, Supercharg er 60 R3.0 Page 22 of 62 4.25 DS7B - OVR HEAT: 4.25.1. Indicates that the discharge bank is overheated. 4.26 DS8A - VOLT FLT: Indicates voltage fault. 4.26.1. Continuous: battery connected with polarity reversed. 4.26.2. Flashing: (during charge) battery voltage in excess of 1.
Operator Manual, Supercharg er 60 R3.0 Page 23 of 62 5. MODES OF OPERATION 5.1 CONSTANT CURRENT CHARGE 5.1.1. DUAL RATE: 5.1.1.1. The battery is charged at the main rate for a preset time, at the end of which, the current is switched automa tically to the topping rate for the remainder of the total time selected.
Operator Manual, Supercharg er 60 R3.0 Page 24 of 62 5.2 CONSTANT VOLTAGE CHARGE. 5.2.1. The battery is charged at the topping rate for a preset time. 5.
Operator Manual, Supercharg er 60 R3.0 Page 25 of 62 5.3 PEAK CHARGE. 5.3.1. Transfer from Main to Topping on Peak voltage. 5.3.1.1. The battery is charged at the main rate for a preset time. 5.3.1.2. When the battery voltage reaches the programmed peak voltage, the charging current transfers from main to topping.
Operator Manual, Supercharg er 60 R3.0 Page 26 of 62 5.4 DISCHARGE. NOTE: For battery voltages greater than 32V, the discharge current is automatically limited to 30A. A current fault condition will result if more than 30A is programmed. 5.4.1. ANALYSIS (AUTO CUT-OFF): 5.
Operator Manual, Supercharg er 60 R3.0 Page 27 of 62 6. OPERATING INSTRUCTIONS 6.1 GENERAL 6.1.1. Verify RESET 6.1.1.1. Verify that the reset button is depressed before turning power on. 6.1.2. Connect the Battery 6.1.2.1. Connect the battery cable to the batteries and the Temp-Plate.
Operator Manual, Supercharg er 60 R3.0 Page 28 of 62 6.2 CONSTANT CURRENT CHARGE. 6.2.1. Main Charge 6.2.1.1. Program Main charging rat e and time duration. 6.2.1.2. Verify that the Timer Speed is consistent with the required elapsed time. 6.2.2. Topping Charge 6.
Operator Manual, Supercharg er 60 R3.0 Page 29 of 62 6.3.4. RUN 6.3.4.1. Start the Charger-Analyzer in the Single Rate mode. The charger will start to reduce the charging current when the battery voltage is within 0.5V of the float level. 6.3.4.2. The charge will terminate autom a tically (CYCLE END) when the TOTAL TIME selected is reached.
Operator Manual, Supercharg er 60 R3.0 Page 30 of 62 6.5 DISCHARGE NOTE: Do not discharge batteries on the Te mp-Plate. Normal battery heating of the Temp-Plate during discharge will generate a false battery Overtemp fault when attempting to recharge it.
Operator Manual, Supercharg er 60 R3.0 Page 31 of 62 6.6 Float/Peak Voltage Chart CHART RELATING THE COMPUTED FLOAT AND PEAK VOLTAGES TO THE SELECTED NOMINAL BATTERY VOLTAGE 6.6.1. The Cell Selector becomes a Nominal Battery Voltage selector in other than the Constant Current settings of the Voltage Mode Selector.
Operator Manual, Supercharg er 60 R3.0 Page 32 of 62 6.7 OPERATING NOTES AND PRECAUTIONS 6.7.1. Make sure that the instrument is in th e reset position before turning power on or off, or before connecting or disconnecting batteries.
Operator Manual, Supercharg er 60 R3.0 Page 33 of 62 7. INSTALLATION 7.1 BENCH SPACE. 7.1.1. The Charger-Analyzer system occupies 19" x 17" (48.3 cm x 43.2 cm) of table top space for the charger and 10" x 25" (25 cm X 63.5 cm) for the Temp-Plate.
Operator Manual, Supercharg er 60 R3.0 Page 34 of 62 Figure 4 - Line Voltage Wiring, 208/230V Figure 5 - Line Voltage Wiring, 230/245V.
Operator Manual, Supercharg er 60 R3.0 Page 35 of 62 8. VERIFICATION OF PERFORMANCE The Charger-Analyzer has been designed, manufactur ed and tested to give thousands of hours of trouble free operatio.
Operator Manual, Supercharg er 60 R3.0 Page 36 of 62 8.3 TIMER VERIFICATION. 8.3.1. Settings: 8.3.1.1. Set the Main and Topping Charge Current Selectors to zero. 8.3.1.2. Set both Time Selectors to zero. 8.3.1.3. Set the Timer Speed Switch to the test (center) position.
Operator Manual, Supercharg er 60 R3.0 Page 37 of 62 8.4 BATTERY OVERTEMP CUT-OFF. 8.4.1. Settings: 8.4.1.1. Set current selectors to zero. 8.4.1.2. Set the timer speed switch in the normal position (to t he right). 8.4.1.3. Set time selectors to 1/01 and cell selector to 20.
Operator Manual, Supercharg er 60 R3.0 Page 38 of 62 8.6 REVERSE POLARITY TEST. 8.6.1. Connect the battery cable to a voltage source (CAL-100 or low current power supply and reverse the polarity). Start at zero and advance towards -0.5V 8.6.2. The unit will show a fault through the alar m and the voltage fault light at about - 0.
Operator Manual, Supercharg er 60 R3.0 Page 39 of 62 8.9 FULL DISCHARGE. 8.9.1. Settings: 8.9.1.1. Set the Current selectors to zero. 8.9.1.2. Set the Time selectors to 1/01. 8.9.1.3. Set the Cell selector to 20. 8.9.2. Start the unit in the full discharge mode (red).
Operator Manual, Supercharg er 60 R3.0 Page 40 of 62 8.12 FLOAT VOLTAGE: 8.12.1. Settings: 8.12.1.1. Set the Voltage Control Selector to Float 8.12.1.2. Set the Current Selectors to zero. 8.12.2. Connect the Battery Cable to the battery and note the battery voltage.
Operator Manual, Supercharg er 60 R3.0 Page 41 of 62 8.14 METERS 8.14.1. Verify the battery voltage and charge/discharge current readings by comparing against a CAL-100 Calibrator or a refere nce voltmeter and a reference ammeter (or reference shunt).
Operator Manual, Supercharg er 60 R3.0 Page 42 of 62 9. CALIBRATION • NOTE 1: The Charger-Analyzer must be verified (and calibrated if required) at least once every 12 months, or earlier if deviation from performance is observed. (see first note on section 7).
Operator Manual, Supercharg er 60 R3.0 Page 43 of 62 • NOTE 3: Insert an external shunt/ammeter in series with any of the battery leads (Use the Single Cell adaptor to facilitate the connection or simply use the CAL-100 which is a calibrator designed specifically to test and certify Charger-Analyzer Battery Charger-Analyzers.
Operator Manual, Supercharg er 60 R3.0 Page 44 of 62 Figure 6 - Meters Board Adjustments.
Operator Manual, Supercharg er 60 R3.0 Page 45 of 62 Figure 7 - Control S witch Board Adjustments 9.2.3. CONTROL SWITCH (see fig 8). NOTE: For voltage measurements use pin 14 on the Current Control Board as reference common. 9.2.3.1. R10 - AMPLIFIER ZERO.
Operator Manual, Supercharg er 60 R3.0 Page 46 of 62 • Set the Cell Selector at 20, connect a voltage source set to approximately 21V and start the Charger-Analyzer in AUTO Discharge • Adjust for a Discharge CUT-OFF (Capacity Failure) at 20.0 volts.
Operator Manual, Supercharg er 60 R3.0 Page 47 of 62 • NOTE 1: Other currents can also be used for High Current calibration (20A minimum). • NOTE 2: Adjustments of R28 and R25 a re interactive. Therefore, It will be necessary to repeat the adjustments once or twice to achieve exact results.
Operator Manual, Supercharg er 60 R3.0 Page 48 of 62 Figure 9 - Monitor Board Adjustments 9.2.4. MONITOR (see fig 12) NOTE: Use pin 14 on the card connector (or the Test Point E0 on the Control Board) as ground reference for voltage measurements. 9.2.
Operator Manual, Supercharg er 60 R3.0 Page 49 of 62 9.2.4.3. R62 – OVERVOLTAGE • (no battery) • Set the Current Selectors to zero and set the Cell Selector to 20 (1.75V/cell) • Connect an external voltage source (Power Supply or CAL-100) to simulate the battery voltage, set at approximately 34V.
Operator Manual, Supercharg er 60 R3.0 Page 50 of 62 Figure 10 - Voltage Control Board Adjustmen ts 9.2.5.2. VOLTAGE CONTROL - See [ Figure 10 ] 9.2.5.2.1. R11, FLOAT VOLTAGE: • Connect a CAL-100 Calibrator or a power supply capable of producing at least 10V (very little current required) to the CAL (-) and CAL (COMMON) test points.
Operator Manual, Supercharg er 60 R3.0 Page 51 of 62 • Adjust R11 as required to reduce the charging current to zero at a simulated battery voltage of 28.0V • Note: This test can also be performed directly on a battery (no calibrator required), preferably one that is “soft” (rapid voltage change).
Operator Manual, Supercharg er 60 R3.0 Page 52 of 62 Figure 11 - Circuit Board Sequence.
Operator Manual, Supercharg er 60 R3.0 Page 53 of 62 10. TROUBLESHOOTING AND REPAIRS 10.1 TROUBLESHOOTING The following are hints and directions to help you locate the most probable causes of deviation of performance as establis hed in the procedures of section 8.
Operator Manual, Supercharg er 60 R3.0 Page 54 of 62 Do not connect a battery under this condition! 10.1.12. 9.1.12 CURRENT TRACKING IS OFF BY A SMALL CONSTANT AMOUNT. Amplifier zero or low current calibrate out of adjustment (CONTROL board). 10.1.13.
Operator Manual, Supercharg er 60 R3.0 Page 55 of 62 10.2 FINDING A SHORTED DISCHARGE TRANSISTOR Since all discharge transistors are wired in pa rallel, a single shorted transistor will short the entire bank.
Operator Manual, Supercharg er 60 R3.0 Page 56 of 62 Figure 12 - Measuring for a shorted transistor.
Operator Manual, Supercharg er 60 R3.0 Page 57 of 62 Figure 13 - Location of Current Limiters.
Operator Manual, Supercharg er 60 R3.0 Page 58 of 62 Figure 14 - Po wer Block Diagram.
Operator Manual, Supercharg er 60 R3.0 Page 59 of 62 11. REPLACEABLE MODULES AND PARTS 11.1 CIRCUIT BOARDS 9879101001 - METERS (LED) 9879903004 - VOLTAGE CONTROL 9879960001 – CONTROL SWITCH 98799030.
Operator Manual, Supercharg er 60 R3.0 Page 60 of 62 12. BATTERY TESTING NOTES Note, hazardous condition: Handle batteries with extreme care as they are capable of very high discharge rates if short circuited. Check the Battery Manual, IMM or Aircraft Manufacturer Instructions for information on charging, discharging and testing.
Operator Manual, Supercharg er 60 R3.0 Page 61 of 62 13. DISCLAIMER • The Supercharger is a precision system inte nded to be operated by personnel qualified in the servicing of aircraft batteries. • The end user must establish suitability in the intended application.
Operator Manual, Supercharg er 60 R3.0 Page 62 of 62 14. REVISION INDEX Table 2 - Index of Revisions REVISION DATE NOTES 1.0 21 January 2001 Release 2.0 15 January 2003 New Clock-Timer Board and add itional Peak function 2.1 23 July 2003 Overvoltage is now set at 1.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Precision Trading SUPERCHARGER 60 R3.0 60 R3.0 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Precision Trading SUPERCHARGER 60 R3.0 60 R3.0 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Precision Trading SUPERCHARGER 60 R3.0 60 R3.0 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Precision Trading SUPERCHARGER 60 R3.0 60 R3.0 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Precision Trading SUPERCHARGER 60 R3.0 60 R3.0, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Precision Trading SUPERCHARGER 60 R3.0 60 R3.0.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Precision Trading SUPERCHARGER 60 R3.0 60 R3.0. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Precision Trading SUPERCHARGER 60 R3.0 60 R3.0 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.