SCXI-1125 Calibration Procedure 2 ni.com Equipment and Other T est Requirements This section describes the equipment, software, documentation, and test conditions required for calibrat ing the SCXI-1125.
© National Instruments Corpor ation 3 SCXI-1125 Calibration Procedur e Documentation The NI-DAQ Function Refe rence Online Help fi le and NI-DAQ User Manual for PC Compa tibles c o n t a i nd e t a i l e di n f o r m a t i o no nu s i n gt h e NI-DAQ driver.
SCXI-1125 Calibration Procedure 4 ni.com Setting Up Y our Module Complete the following steps to set up yo ur module for testing : 1. Install the SCXI-1125 in sl ot 1 of the SCXI chassis. 2. Connect a 68-to-68-pin cable between the S CXI module and the E Series D A Q de vice installed in the host computer .
© National Instruments Corpor ation 5 SCXI-1125 Calibration Procedur e 7. Connect the calibrator to the appro priate analog input, starting with channel 0. Refer to Figure 1 to determine which pins on the 96-pin front conn ector correspond to t he positiv e and negati ve inputs of the specified channel.
SCXI-1125 Calibration Procedure 6 ni.com 13. Call SCXI_Scale . Set the following parameters: • SCXIChassisID — The v alue assigned b y MAX • moduleSlot — 1 • channel — The same channel num.
© National Instruments Corpor ation 7 SCXI-1125 Calibration Procedur e • calOp — ND_SELF_CALIBRATE • s etOfCalConstants — ND_USER_EEPROM_AREA • calRefVolts — 0.
SCXI-1125 Calibration Procedure 8 ni.com • filterMode — 1 • cutoffFrequency — 10,000 9. Connect the calibrator to the analog input you w ant to adjust. Refer to Figure 1 to determine the pins on the 96-pin f ront connector that correspond to the positiv e and neg ati ve inputs of the specif ied channel.
© National Instruments Corpor ation 9 SCXI-1125 Calibration Procedur e 16. Record binary1 with volt1 and binary2 with volt2 for later use. 17. Repeat steps 7 through 16 for the remaining gain v alues you want to adjust on this channel. 18. Repeat steps 6 through 17 for the remaining chan nels you want to adjust.
SCXI-1125 Calibration Procedure 10 ni.com V erifying Adjusted V alues After completing the adjustment s to your SCXI-1125 module, it is important that you verify the analog input operation by repeating t he steps listed in the Veri fying Gain and Offset Errors section of this docu ment.
© National Instruments Corpor ation 11 SCXI-1125 Calibration Procedur e Front and Rear Signal Connector Diagrams Figure 1. Front Signal Connector Pin Assignments CHSGND +5 V RSVD CH7+ CH6+ CH5+ CH4+ .
SCXI-1125 Calibration Procedure 12 ni.com Figure 2. Rear Signal Connector Pin Assignments SERCLK DIGGND SLOT0SEL* DAQD*/A SERDATIN OUTREF MCH7+ MCH6+ MCH5+ MCH4+ MCH3+ MCH2+ MCH1+ MCH0+ AOGND SCANCLK .
© National Instruments Corpor ation 13 SCXI-1125 Calibration Procedur e Specifications The following table contains the sp ecifications you should use w hen verifying and calibrating y our SCXI-1125 module. If you r device has been calibrated within the last year, the test point value should fall between the Upper Limit and Lower Limit valu es.
SCXI-1125 Calibration Procedure 14 ni.com 200 – 0.025 – 0.024750 – 0.02468642 – 0 .02481358 250 0.02 0.019800 0.01985126 0.019 74874 250 0 0.000000 0.000022 – 0.000022 250 – 0.02 – 0.019800 – 0.01974874 – 0.0 1985126 500 0.01 0.009900 0.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il National Instruments SCXI -1125 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del National Instruments SCXI -1125 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso National Instruments SCXI -1125 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul National Instruments SCXI -1125 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il National Instruments SCXI -1125, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del National Instruments SCXI -1125.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il National Instruments SCXI -1125. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo National Instruments SCXI -1125 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.