NI PXI/PCI-5124 Specifications 12-Bit 200 MS/s Digitizer This document lists the sp ecifications for the NI PX I/PCI-5124 (NI 5124) high-speed digitizer.
NI PXI/PCI-5124 Specifications 2 ni.com PFI 0 and PFI 1 (Programmabl e Function Interfac e, AUX Front Panel Connector) ................... .................... ................... .....20 TClk Specifications .............. .................... .......
© National Instruments Corporat ion 3 NI PXI/PCI-5 124 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable V ertical Offset 50 Ω 1M Ω — Range (V pk-pk ) V ertical Of fset Range (V) Range (V pk-pk ) V ertical Offs et Range (V) 0.
NI PXI/PCI-5124 Specifications 4 ni.com DC Drift Range (V pk-pk ) 50 Ω and 1 M Ω — 0.2, 0.4, 1, and 2 ±(0.057% of Input + 0.006% of FS +1 0 0 μ V) per °C 4, 10, and 20 (1 M Ω only) ±(0.057% of Input + 0.006% of FS + 900 µV) per °C A C Amplitude Accuracy 50 Ω 1M Ω W ithin ±5 °C of self-calibration temperature ±0.
© National Instruments Corporat ion 5 NI PXI/PCI-5 124 Specifications Figure 1. NI 5124 Frequency Response (T ypical) A C Coupling Cutof f (–3 dB) 12 Hz A C coupling av ailable on 1M Ω only Passband Flatness Filter Settings Range (V pk-pk ) 50 Ω and 1 M Ω Referenced to 50 kHz Filters Off All ranges except 0.
NI PXI/PCI-5124 Specifications 6 ni.com Specif ication Va l u e Comments Spectral Characteristics Spurious-Free Dynamic Range with Harmonics (SFDR) (Physical), T ypical Range (V pk-pk ) 50 Ω 1M Ω Fi.
© National Instruments Corporat ion 7 NI PXI/PCI-5 124 Specifications Intermodulation Distortion, T ypical 0.2 V pk-pk to 2.0 V pk-pk Ranges on 50 Ω Input Filters of f or antialias filter on T wo tones at 10.
NI PXI/PCI-5124 Specifications 8 ni.com Figure 2. NI 5124 Dynamic Performance, 50 Ω , 1 V pk-pk Range, 262,144 Point FF T (T ypical) –110 –100 –90 –80 –70 –60 –50 –40 –30 –20 –10 0 Amplitude (dBFS) 10 M 20 M 30 M 40 M 50 M 60 M 70 M 80 M 90 M 100 M 1 M F requency (Hz) 9.
© National Instruments Corporat ion 9 NI PXI/PCI-5 124 Specifications Specif ication Va l u e Comments RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Noise filter on 50 Ω terminator connected to input 0.2 NI PXI-5124: 94 μ V rms (0.047% FS) NI PCI-5124: 106 μ V rms (0.
NI PXI/PCI-5124 Specifications 10 ni.com RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Antiali as fil t e r on 50 Ω terminator connected to input 0.2 NI PXI-5124: 112 μ V rms (0.056% FS) NI PCI-5124: 126 μ V rms (0.063% FS) NI PXI-5124 : 130 μ V rms (0.065% FS) NI PCI-5124: 146 μ V rms (0.
© National Instruments Corporat ion 11 NI PXI/PCI-5 124 Specifications RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Filters off 50 Ω terminator connected to input 0.2 NI PXI-5124: 114 μ V rms (0.057% FS) NI PCI-5124: 128 μ V rms (0.064% FS) NI PXI-5124 : 164 μ V rms (0.
NI PXI/PCI-5124 Specifications 12 ni.com Figure 3. Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Noise Filter Enabled, 1 M Ω Input Impedance Figure 4.
© National Instruments Corporat ion 13 NI PXI/PCI-5 124 Specifications Horizontal Sample Clock Specif ication Va l u e Comments Sources NI PXI-5124 NI PCI-5124 * Internal Sample Clock is locked to th.
NI PXI/PCI-5124 Specifications 14 ni.com Sample Clock Jitter , T ypical ≤ 1 ps rms (10 0 Hz to 100 kHz) ≤ 2p s r m s ( 1 0 0 H z t o 1M H z ) Includes the ef fects of the con verter aperture uncer.
© National Instruments Corporat ion 15 NI PXI/PCI-5 124 Specifications External Sample Clock Sources NI PXI-5124 NI PCI-5124 — CLK IN (front panel SMB connector) PXI Star Trigger (backplane connect.
NI PXI/PCI-5124 Specifications 16 ni.com Phase-Locked Loo p (PLL) Reference Clock Sample Clock Exporting Exported Sample Clock Destinations Destination Maximum Frequency * Decimated Sample Clock only CLK OUT (front panel SMB connector) 210 MHz PXI_T rig<0.
© National Instruments Corporat ion 17 NI PXI/PCI-5 124 Specifications CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) CLK OUT (Sample Clock and Reference Clock Output, Front Panel Connector) T rigger Reference (Stop) T rigger Specification Va l u e Comments Input V oltage Range Sine wav e: 0.
NI PXI/PCI-5124 Specifications 18 ni.com Ti m e Resolution TDC Onboard Clock External Clock TDC = T ime to Digita l Con version Circuit On 50 ps N/A Of f 5n s External Clock Period Rearm T ime TDC Rearm T ime Ho ldoff s et to 0 On 10 μ s Of f 2 μ s Holdoff TDC Onboard Clock External Clock — On 10 μ s to 85.
© National Instruments Corporat ion 19 NI PXI/PCI-5 124 Specifications TRIG (External T rigger , Front Panel Connector) T rigger Filters Lo w-Frequenc y (LF) Reject High-Frequency (HF) Reject — 50 kHz 50 kHz Digital Trigger (Digital Trig ger Type) Sources NI PXI-5124 NI PCI-5124 — PXI_T rig<0.
NI PXI/PCI-5124 Specifications 20 ni.com PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connector) Specif ication Va l u e Co mments Connector 9-pin mini-circu lar DIN — Direction.
© National Instruments Corporat ion 21 NI PXI/PCI-5 124 Specifications TClk Specifications National Instruments TClk synchronizati on method and the NI-TClk driver are used to align the sample clocks on any numb er of SMC-based modules in a chassis.
NI PXI/PCI-5124 Specifications 22 ni.com W aveform Specifications Specification Va l u e Comments Onboard Memory Size 8 MB per channel stan dard (4 megasamples per channel) * NI PXI-5124 only 32 MB pe.
© National Instruments Corporat ion 23 NI PXI/PCI-5 124 Specifications Calibration Power Specification Va l u e Comments Self-Calibrati on Self-calibration i s done on soft ware command. The calibration corrects for gain, offset, frequency response, triggering, and timing adjust ment errors for all input ranges.
NI PXI/PCI-5124 Specifications 24 ni.com Software Specif ication Va l u e Comments Dri ver Software NI-SCOPE 2.7 or later NI-SCOPE is an IVI-comp liant driv er that allows you to configure, control, and calib rate the NI 5124. NI-SCOPE provides application program ming interf aces for many dev elopment environments.
© National Instruments Corporat ion 25 NI PXI/PCI-5 124 Specifications Environment NI PXI-5124 Note T o ensure that the NI PXI-5124 cools ef fectiv ely , follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5124 kit.
NI PXI/PCI-5124 Specifications 26 ni.com NI PCI-5124 Note T o ensure that the NI PCI-5124 cools effecti v ely , make sure that the chassis in which it is used has activ e cooling that provides at least some airflow acros s the PCI card cage. T o maximize airflo w and extend the life of the de vice, leav e any adjacent PCI slots empty .
© National Instruments Corporat ion 27 NI PXI/PCI-5 124 Specifications Safety , Electromagnetic Compatibility , and CE Compliance Safety This product is designed to meet the requirements of the follo.
NI PXI/PCI-5124 Specifications 28 ni.com Environmental Management National Instruments i s committed to designing and manu facturing products in an environmental ly responsible mann er. NI recognizes that eliminating certa in hazardous substances from our products is beneficial not only to t he environment but also to NI customers.
© National Instruments Corporat ion 29 NI PXI/PCI-5 124 Specifications Dimensions and W eight NI PXI-5124 Front Panel I ndicators Label Function Fo r m or e information, refer to the NI High-Speed Digitizers Help . A CCESS The A CCESS LED indicates the status of the PCI b us and the interface from the NI PXI-5124 to the co ntroller .
NI PXI/PCI-5124 Specifications 30 ni.com NI PCI-5124 Dimensions 35.5 × 2.0 × 11.3 cm (14.0 × 0.8 × 4.4 in.) We i g h t 455 g (16 oz) 2.0 cm (0.8 in.
© National Instruments Corporat ion 31 NI PXI/PCI-5 124 Specifications Where to Go for Support The National Instrum ents W eb site is your complete resource for technical support.
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more inform ation about Natio nal Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il National Instruments 12-Bit 200 MS/s Digitizer è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del National Instruments 12-Bit 200 MS/s Digitizer - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso National Instruments 12-Bit 200 MS/s Digitizer imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul National Instruments 12-Bit 200 MS/s Digitizer ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il National Instruments 12-Bit 200 MS/s Digitizer, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del National Instruments 12-Bit 200 MS/s Digitizer.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il National Instruments 12-Bit 200 MS/s Digitizer. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo National Instruments 12-Bit 200 MS/s Digitizer insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.