PN 644226 July, 1997 Rev.1, 3/99 © 1997,1999 Fluke Corporation, Al l rights reserved. Printed in U .S.A. All p r oduct names are trademarks of their respective companies. 1 ® 16 Multimeter with Temperature Calibration Information W Warning To avoid electrical shock, remove test leads and any input signals from the Meter before opening the case.
16 Multimeter with Temperature 2 Accuracy specifications are given as follows: ±([% of reading] + [number of least significant digits]). Table 1. Specifications (cont) Function Range Resolution Accuracy f (50 to 400 Hz) 4000 mV 1 4.000 V 40.00 V 300.
Specifications 3 Table 1. Specifications (cont) Function Overload Protection 1 Input Impedance (Nominal) Common Mode Rejection Ratio (1 k Ω Unbalanced) 2 Normal Mode Rejection 2 f 600 V dc >5 M .
16 Multimeter with Temperature 4 Disassembly and Reassembly Referring to Figure 1, disassemble the Meter as follows. A Phillips-head screwdriver and small flat-blade screwdriver are required. 1. Remove the test leads and set the rotary knob to OFF. 2.
Recommended Test Equipment 5 Case, Top Decal Switch Assembly Actuator, Switch Main PCA Shield, Bottom Foot LCD Conn. Elastomeric Conn. Elastomeric LCD to PCA Support, Switch Assembly Battery Shock Absorber Case, Bottom Screw, THD Form, PH.P.STL, 5-14 x .
16 Multimeter with Temperature 6 1 2 3 Snap Snap Snaps ig2f.eps Figure 2. Removing and Reinserting the Printed Circuit Assembly.
Performance Test 7 Performance Test W Warning To avoid electric shock, do not execute the performance tests procedures unless the Meter is fully assembled. To ensure that the Meter is working properly and performs to specifications, use the following procedures.
16 Multimeter with Temperature 8 Table 3. Performance Tests Test No. Meter Range Input To Meter Switch Position Display Minimum Display Maximum 1 2 3 4 5 6 7 8 9 1 10 4.000 V ac 4.000 V ac, MIN MAX 4.000 V ac 4.000 V ac 4.000 V ac 4.000 V ac 40.00 V ac 40.
Calibration Adjustments 9 + + 5700A Ice Bath AK-80 Banana Plug Temperature Yellow Red Copper 0 ig6f.eps Figure 3. Performance Test for Temperature + _ 9V BATTERY + Clip to Base Contacts Adjust R4 for 3.997-4.003V DC Input Receptacles + ig1f.eps Figure 4.
16 Multimeter with Temperature 10 Temperature Note Adjust DC voltage (R4) before adjusting temperature (R38). The following describes three temperature calibration procedures: • Calibrating at room temperature. Use this procedure if it is not necessary to optimize the Meter’s performance.
Replacement Parts 11 ! CAL ENABLE 10A 100 mA 1000V 750V MAX ALL INPUTS 1kV MAX 500mA F 250V POWER AUTO 2ND MN MX RATE dB REL LOCAL THRESH ADDR BAUD HOLD LO HI COMP A A V FREQ DUAL DISPLAY MULTIMETER 45 V REF# REF V COM FUSED 80T-150U + _ 9V BATTERY + + R38 Clip to Base Contacts V m ig4f.
16 Multimeter with Temperature 12.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Fluke Fluke 16 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Fluke Fluke 16 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Fluke Fluke 16 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Fluke Fluke 16 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Fluke Fluke 16, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Fluke Fluke 16.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Fluke Fluke 16. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Fluke Fluke 16 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.