Sierra Model 951 Digital Flo-Box Instruction Manual Part Number IM-951 Revision A 09-05 5 Harris Court, Building L Monterey , CA 93940 (831) 373-0200 (800) 866-0200 Fax (831) 373-4402 www .
Page 1 DESCRIPTION The Model 951 Digital Flow Box is a microprocessor-based digital indicator and power supply , capable of interfacing directly to a mass flow meter or a mass flow controller . The integrated +/-15 vdc @250ma power supply provides a well regulated, short circuit and thermal overload protected output.
INST ALLA TION AND WIRING INST ALLA TION The Model 951 enclosure is designated for panel mounting in a 1/8 DIN cutout. The cutout dimensions are shown below . T o panel mount the 951, perform the following steps: 1. Rotate the four pawl screws ( outside screws in each corner) several turns counter-clockwise to retract the pawls.
Page 3 CAL T ARE H1 H2 L1 L2 Pawl Screw (1 of 4 outside screws) These 4 screws are used to mount the instrument in a panel. Menu Switch S1 (Single Function Switch) Increment Switch S2 (Single Function.
Page 4 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | J1 P1 6 +15 vdc 8 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |.
Page 5 SWITCH DEFINITIONS Menu Switch (S1) The Menu switch (S1) is used to cycle through the flow controller setpoints, alarm setpoints, alarm hysteresis, calibration values and decimal point locations. The sequence is as follows. The designators, in parenthesis, identify text that will be displayed, momentarily , followed by the value.
Page 6 SWITCH DEFINITIONS (cont) Step and Min-Max Switch (S3) This switch has a dual function. When in the MENU sequence, it steps the digit to be incremented, one position to the right. Using S3 in conjunction with S2 allows rapid updating of all parameter values.
Page 7 FUNCTION DEFINITIONS The Mode switch (S1) cycles through the following functions. Flow Set Point (StPt) The Setpoint (0-5 vdc) command signal used to control the flow through a flow controller . The output voltage is 0-5 vdc. The setpoint output is defined as: Setpoint Output = StPt / CAL x 5.
Page 8 FUNCTION DEFINITIONS (cont) Alarm Set Point High 2 (SPH2) SPH2 is independent and functions identically to SPH1. SPH2 has its own open collector outputs and front panel LED (H2). Alarm Set Point Low 2 (SPL2) SPL2 is independent and functions identically to SPL1.
Page 9 FUNCTION DEFINITIONS (CONT) Decimal Point (dP) dP allows selection of decimal point locations for display . The locations selectable are x.xxxx, xx.
Page 10 OPERA TING INSTRUCTIONS TURN-ON INSTRUCTIONS Reference Figures 3 and 4 for proper input and output signal wiring. 1. Apply power to the instrument. 2. V erify the instrument readout blanks momentarily , then displays the value of the input signal.
Page 1 1 PROGRAM INSTRUCTIONS (cont) 2. Depress MODE (S1) switch once. The text followed by the current value of Alarm Set Point High 1 (SPH1) will be displayed. Use S2 and S3 to set Alarm Set Point High 1 to the desired value. Proceed to Step 3 to set Alarm Setpoint Low 1 (SPL1) or depress S4 to exit MODE sequence.
Page 12 PROGRAM INSTRUCTIONS (cont) 7. Depress MODE (S1) switch once. The text followed by the current value of Hysteresis Low will be displayed. Repeat Step 2 to set Alarm Hysteresis Low to the desired value. The maximum value for HL is 99 counts.
Page 13 Prior to connecting the 951 to the Flowmeter or Controller , verify that the Controller's +/-15 vdc requirement is no more than 40 0 ma. The 951 is capable of supplying up to 500ma for short periods of time ( in 1/2 or 1 hour increments).
Page 14 CALIBRA TION INSTRUCTIONS (cont) 6. Depress and hold CAL (S5) for approximately 3 seconds. The FS CAL LED should illuminate and the readout should display the FS Cal value within +/- 1 digit.
Page 15 RS232 SERIAL COMMUNICA TIONS (Special Order Only) The Parity (Even or None) and the Baud Rate (2400 or 9600) are selectable by setting DIP switches SW a and SWb, accessible from the right side of the instrument case.
Page 16 RS232 SERIAL COMMUNICA TIONS (Special Order Only) [Z] [CR] [LF] Zero the displayed reading [F] [CR] [LF] Calibrate Full Scale (Input signal is scaled to display Full Scale) [S] [counts] [CR] [LF] Set Full Scale display value.
MODEL SR5 PROGRAM MENU DIAGRAM Depress MODE ST ART Set Depress Depress Increments Flashing Digit Steps Flashing Digit to Right Set Depress MODE Depr.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Sierra 951 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Sierra 951 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Sierra 951 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Sierra 951 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Sierra 951, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Sierra 951.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Sierra 951. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Sierra 951 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.