15 Discovery Way, Acton, MA 01720 Phone: (978)263-3 584, Fax: (978)263-50 86 Web Site: www.princetoninstruments.com Operating Instructions A cton Se ries SP - 2300i 0.300 Meter Focal Length Triple Grating Imaging Monochromator / Spectrograph Rev 4 .0 9.
1 A cton SP-2300i Operating Inst ructions CONTENTS: Page I. Brief Introduction and SP- 2300i Description 2 II. Specifications 3 III. SP - 2300i Setup 4 A. Unpacking Note 4 B. Connecting the SP -2300i to a Computer 4 C. Cables and Connecti ons for SP-2300i 5 D.
2 I : Brief Introductio n and SP-2300i Descr iption Brief Introduction: This instruction manual is i ntended to assist you in s et - up and operation of your new SP -2300i monochromator/spectrograp h.
3 II : SP -2300i Specifications Every SP-2 300i monochromator or spec trograph includes a triple grating turret for 1, 2, or 3 gratings, 32 -bit microprocessor controll ed scanning, buil t-in RS - 232 and USB interfaces, and micrometer controlled e ntrance slit.
4 III : SP -2300i Setup Section III-A: Unpacking and Inspection Carefully unpack and examine the SP-2300i and any ac cessories purc hased. Note: Report any damage i mmediately t o the carrier and t o Princeton Instrum ents, and save all pac ki ng materia l.
5 Section III-C: Cables and Connections for SP-2300i The following computer cables are supplied wi th the SP-2300i: CC - 499 - 2 IBM AT or compatible 9 pin female ( DB9S) connector to 9 pin male connector (DB9P). 3650 -USB- 06 USB Cable Type A to Type B.
6 Section III-D: Mounting Accessories to the SP -2300i Slit A sse mblies All Acton accessories come with thei r own set of instructions for proper mounting and operation. The instructions below are only general i nformation. Pleas e refer to the individual instructions for detailed information.
7 Section III-E: Detector Mounting Instructions: Mounting Focal Plane Detectors to the SP-2300i Spectrograph The standard mounting f lange for CCDs and diode arrays acc ommodates detectors with two diff erent bolt circles. There are three (3) equal ly spaced #10 - 32 tapped holes on a 3.
8 Section III-F: Slit Width A djustments in the Model 716 Bilateral Slits The slit width of eac h bilateral slit assembly (716 type) is adjustable f rom 0.010 millimeters to 3 millimeters (10 to 3,000 µm) by a micrometer knob loc ated on the top of the sl it housing.
9 IV : SP-2300 i Operation Section IV-A: Initialization W hen power is turned ON to the SP -2300i, it ini tializes to a wavelength of 0.0 nm for grating number 1. If the power is switched OFF and then ON again to the SP-230 0i, it will re-ini tialize.
10 NM/MIN Sets the scan rate i n nm/min to 0.01 nm/min resolution with units nm/min ?NM/MIN Returns present scan rate i n nm/min to 0.01 nm/min resoluti on with units nm/min Grating Contro l Commands : GRAT ING Places specified grating in position to the wavelength of the wavelength on the present grating.
11 Slit Width Cont rol Command s (for Motorized Slits only ): FRONT-EX IT-SLIT Designates front ex it slit to receive slit control commands. Note : The designation remains in effec t until changed by another slit des ignator. This command does not have to be repeated unti l the designate d sl it is changed.
12 Section IV-C : Focusing and A lignment of A rra y Detectors W ith the array detector properly mounted to the SP-2300 i, use the followi ng procedure to align and foc us the array detector to the SP-230 0i optical system. It is as sumed that the array detection s ystem is running with SpectraSense or W inSpec.
13 Section IV-D : Conv erting the SP-2300i from Monochromator to Spectrograph, or Spectrograph to Monochromator Monochromator to Sp ectrograph: This procedure requires that you ha ve the monochromator version, Model SP- 2355 or SP- 2357 , plus the array detector mounting fl an ge.
14 V : Ap pendices and Schemat ic Dra wings A ppendix V -A: Selecting Alternate Start-Up Para meters The following are the Start-U p parameters and their def ault values: GRATING #1 WAVELENGTH 0.0 nm SCAN SPEED 100.0 nm/min Each of the above may be changed through the RS - 232 port or USB port using the f ollowing commands.
15 A ppendix V-B: Cable Connection Diagrams - Monochromators & Spectrographs to Computer Figure 9 . RS -232 Cable Wir ing Figure 10 . Cable Connections (Back Panel ) POWER Power connection +5 and +24 Vol ts USB HUB USB Hub 1.1 compatible. USB USB connection to computer.
16 A ppendix V -C: Standard Slit Assembly Drawing.
17 A ppendix V -D: Accessories Listing SP -2300i M onochromators & Spectrographs SP - 2355 SP - 2300i, 300mm Monoch romator: Micrometer adjustabl e entrance/exit slits, 90° optical path, internal.
18 TDS- 429 Dual Light Source with 30 watt deuterium & 30 watt tungsten -halogen lamps. Inc ludes housing with light c ollection mirror, mounting flange, and power supply. Provides c ombined output from 190 to >2.0 µm. XS - 43 3 75 Watt Xenon Light Source includes power s upply, housing, mounting flange and light collection mirror.
19 Filter Wheels and Order Sorting Filter Ass emblies FA - 2448 Six Position Filter Wheel Assembl y. Manual sample index ing, holds up to six samples, each 1.
20 A ppendix V -E: Gratings for SP-2300i Grooves/mm Blaze Wavelength Standard Part Number Mechanical Limit Notes: † Bl az ed H ol og rap hi c, 19 0- 450 nm , op tim um ra ng e †† H ol ogr ap hi .
21 A ppendix V -F: Certification and Warranty Certification Princeton Instruments certifies that thi s instrument was thoroughly tested and f ound to meet the specif ications furnished by Princ eton Instruments when it was shipped f rom the factory.
22 A ppendix V -G: Drawings.
23.
24.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Princeton Digital (USA) SP-2300i è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Princeton Digital (USA) SP-2300i - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Princeton Digital (USA) SP-2300i imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Princeton Digital (USA) SP-2300i ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Princeton Digital (USA) SP-2300i, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Princeton Digital (USA) SP-2300i.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Princeton Digital (USA) SP-2300i. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Princeton Digital (USA) SP-2300i insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.