OPERATING INSTRUCTIONS NI 9219 4-Channel, 24-Bit, Universal Analo g Input Module.
NI 9219 Operating Instructions 2 ni.com These operating instructions desc ribe ho w to use the National Instruments 9219. For information about installing, conf iguring, and programming the syst em, refer to the system documentation.
© National Instruments Corp. 3 NI 9219 Operating Instructions Safety Guidelines for Hazardous V oltages If hazardous voltages are connect ed to the module, take the follow ing precautions. A hazardous voltage is a voltage greater than 42.4 V pk or 60 VDC to earth ground.
NI 9219 Operating Instructions 4 ni.com Figure 1 sho ws the NI 9972 connector backshell. Figure 1. NI 9972 Connector Backshell Safety Guidelines fo r Hazardous Loca tions The NI 9219 is suitable for u.
© National Instruments Corp. 5 NI 9219 Operating Instructions Caution Do not remove modules unle ss power has been switched off or the area is known to be nonhazardous. Caution Substitution of component s may impair suitability for Class I, Division 2.
NI 9219 Operating Instructions 6 ni.com Wiring the NI 9219 The NI 9219 has four 6-terminal connectors that pro vide connections for four analog input channels. Connect the positi ve signal of the signal source to the positive input signal terminal (HI) and the negati ve signal of the signa l source to the negati ve input signal terminal (LO).
© National Instruments Corp. 7 NI 9219 Operating Instructions Table 1. Signal N ames Module T erminal Signal Name Signal Description 1 T+ TEDS Data 2 T– TEDS COM 3 EX+/HI * Pos itive e xcitation or.
NI 9219 Operating Instructions 8 ni.com Table 2. Terminal Assignment s Mode T erminal 123456 V oltage T+ T – — HI LO — Current T+ T – HI — LO — 4-W ire Resistance T+ T – EX+ HI EX – LO.
© National Instruments Corp. 9 NI 9219 Operating Instructions Connecting Wires to the NI 9219 Connectors Use a flathead scre wdriv er with a blade smaller than 2.3 × 1.0 mm (0.09 × 0.04 in.) to connect wires to the detachable spring-terminal connectors.
NI 9219 Operating Instructions 10 ni.com Wiring TEDS Cha nnels The NI 9219 supports only Class II TEDS sensors. Connect th e two TEDS lin es to TEDS Data (T+) and TEDS COM (T–) and ensure that neither T+ nor T– is tied in common to an y of the signal inputs (terminals 3 through 6) on the NI 9219.
© National Instruments Corp. 11 NI 9219 Operating Instructions Figure 3. Connecting a G rounded Signal Sour ce to the NI 9219 Figure 4. Connecting a Floating Signal Source to the NI 9219 + – T wist.
NI 9219 Operating Instructions 12 ni.com NI 9219 Timing Options The NI 9219 supports four dif f erent timing options that are optimized for dif ferent types of applications by using dif ferent ADC con version times.
© National Instruments Corp. 13 NI 9219 Operating Instructions Figure 5. Input Circuitry for One Channel Volta ge and Current Modes In V oltage and Current modes, c onnect t he signal source to the NI 9219 across the HI and LO terminals. The current is computed from the v oltage that the ADC measures across an interna l shunt resistor .
NI 9219 Operating Instructions 14 ni.com 4-Wire Resistance and 4-Wire RTD Modes 4-W ire Resistance and 4-W ire R T D modes source a current, which va ries based on the resistance of the load, between the EX+ and EX– terminals. The me asured resistance is computed from the resulting v oltage readi ng.
© National Instruments Corp. 15 NI 9219 Operating Instructions lead wires have the same resistance . A gain of 2x is applied to the volta ge across the neg ativ e lead wire and the ADC uses this v oltage as the negati ve reference to canc el the r esistance error across the positiv e lead wire.
NI 9219 Operating Instructions 16 ni.com for lead wire resistance. Refer to Figure 9 for an illustration of the connections. Figure 9. Connections in 2-Wire Resistan ce and Quarter-Bridge Modes Half-B.
© National Instruments Corp. 17 NI 9219 Operating Instructions Figure 10. Connections in Half-Br idg e and Full-Bridge Modes Thermocouple Mode In Thermocouple mode, connect the positi ve end of the thermocouple to HI and the ne gati ve end of the thermocouple to LO.
NI 9219 Operating Instructions 18 ni.com en vironment and av oid placing heat sources near the module or its connectors. Refer to the Specifications section for more information about accuracy . Th e NI 9219 does not support open thermocouple detection.
© National Instruments Corp. 19 NI 9219 Operating Instructions circuit is open, make sure no more than ±60 V is sourced across the switch. Open Contact mode is supported only in CompactRIO systems. Refer to Figure 12 for an illustration of the connections.
NI 9219 Operating Instructions 20 ni.com Binary Data NI 9219 modules in the system retu rn calibrated binary data when used in a CompactRIO chassis. For these modules, you can con vert the data to engineering units in software. Refer to the software documentation for inform ation about con verti ng data.
© National Instruments Corp. 21 NI 9219 Operating Instructions Input Characteristics Number of channels ............ ............. . 4 analog i nput channels ADC resolution .............. .............. ..... 24 bits T ype of ADC ........... ........
NI 9219 Operating Instructions 22 ni.com Con version time, no ch annels in TC mode High speed .. ............. ........... ........ 10 m s for all channe ls Best 60 Hz rejection ......... .......... 110 ms for all channels Best 50 Hz rejection .......
© National Instruments Corp. 23 NI 9219 Operating Instructions Input impedance V oltage and Digital In modes ( ±60 V , ±15 V , ±4 V ) ........... .......... 1 M Ω Current mode .... .............. ........... . < 40 Ω All other modes ..... ...
NI 9219 Operating Instructions 24 ni.com Cold-junction compensa tion sensor accuracy ....... .............. ........... . ±1°C typ 4-W ire and 3-W ire R TD, Pt 1000 ±0.1, ±0.5 ±240, ±640 4-W ire and 3-W ire R TD, Pt 100 ±0.1, ±0.5 ±2400, ±6400 Quarter-Bridge, 350 Ω ±0.
© National Instruments Corp. 25 NI 9219 Operating Instructions Stability Mode, Range Gain Drift (ppm of Reading/°C) Offset Drift (ppm of Range/°C) V oltage, ± 60 V ±20 ±0.2 V oltage, ± 15 V ±20 ±0.8 V oltage, ± 4 V ±20 ±3.2 V oltage, ± 1 V ±10 ±0.
NI 9219 Operating Instructions 26 ni.com Input noise in ppm of Range rms Full-Bridge, ± 62.5 mV/V ±3 ±20 Full-Bridge, ± 7.8 mV/V ±3 ±20 Mode, Range Con version Time High speed Best 60 Hz rejection Best 50 Hz rejection High res o - lution V oltage, ±60 V 7.
© National Instruments Corp. 27 NI 9219 Operating Instructions 4-W ire and 2-W ire Resistance, 10 k Ω 4.1 1.3 0.8 0.3 4-W ire and 2-W ire Resistance, 1 k Ω 7.1 1.8 1.2 0.7 4-W ire and 3-W ire R TD, Pt 1000 7.6 1.7 1.1 0.4 4-W ire and 3-W ire R TD, Pt 100 10.
NI 9219 Operating Instructions 28 ni.com Input bias curre nt ................ ........... ... <1 nA INL ........... ........... ........... .............. ..... ±15 ppm CMRR ( f in = 60 Hz) ............ ........... ... >100 dB NMRR Best 60 Hz rej ection .
© National Instruments Corp. 29 NI 9219 Operating Instructions Excitation le vel for Resistance , R TD, and Quarter-Bridge modes MTBF .......... .....
NI 9219 Operating Instructions 30 ni.com Thermal dissipation (at 70 °C) Activ e mode .............. ............ ..... 625 mW max Sleep mode . ........... ........... .......... 25 μ W max Physical Characteristics If you need to clean the module, wipe it with a dry towel.
© National Instruments Corp. 31 NI 9219 Operating Instructions Channel-to-earth ground Continuous .... ........... ............ 250 V AC, Measurement Category II W ithstand .
NI 9219 Operating Instructions 32 ni.com by module number or product line, and clic k the appropriate link in the Certification column. Hazardous Locations U.
© National Instruments Corp. 33 NI 9219 Operating Instructions Maximum altitude.......... ......... .......... 2,000 m Pollution Degree (IEC 60664) .......... 2 Shock and Vibration T o meet these specif ications, you must panel mount the system. Operating vibration Random (IEC 60068-2-34).
NI 9219 Operating Instructions 34 ni.com Note For EMC compliance, operate this de vice with shielded cabling. CE Compliance This product meets the essentia l requirements of applicable European direct.
© National Instruments Corp. 35 NI 9219 Operating Instructions Calibration Y ou can obtain the calibration cert ificate and information about calibration services for th e NI 9219 at ni.com/calibration . Calibration interval .... ............. ......
© 2007 National I nstruments Corp. All rights reserved. 374473D-01 Jun07 National Instruments, NI, n i.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about Nat ional Instruments trademarks.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il National Instruments NI 9219 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del National Instruments NI 9219 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso National Instruments NI 9219 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul National Instruments NI 9219 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il National Instruments NI 9219, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del National Instruments NI 9219.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il National Instruments NI 9219. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo National Instruments NI 9219 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.