Manuale d’uso / di manutenzione del prodotto 4909 CPVC del fabbricante Honeywell
Vai alla pagina of 34
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 70-82-25-19 Rev. 1 7/99.
ii 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 1999 by Honeywell Inc.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual iii A bout This Document Abstract The purpose of this document is to support the installation operation and m aintenance of the 4909 CPVC Conductivity Cell Remov al/Insertion Assem bly.
iv 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Organization Telephone Address Honeywell TAC 1-800-243-9883 Voice 1100 Virginia Drive Fort Washington, PA 19034.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual v Contents 1. INTRODUCTION ....................................................................................................1 1.1 Overview ............................
vi 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4. MA INTENA NCE A ND REPLAC EMENT PA RTS .................................................17 4.1 To Clean The Cell ...................................................
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual vii Tables Table 4-1 Replacem ent Parts __________________________________________________________ 18 Table 4-2 Voltage.
.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 1 1. Introduction 1.1 Overview The 4909 Conduct ivit y C ell Ins ertion/R emoval A ssembly is design ed fo.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 2 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 1.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 3 1/ 4" NP T Plug Cabl e 1 - 1/ 2" NPT Nipple Conducti vity Cell P urg e Po rt C ompr es s ion H.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 4 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Cable C onductivity Cell Ad ap t er * * Au t oma t ic Tempera t ur e Com pensato r , w hen specifie d by cat a log nu mber, is m ounted insid e ad apter.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 5 A D C B Ter mina l Boar d As s e m b l y #6- 32 Bi ndi ng He ad Sc r e w & Wa s h e r Vi e w of B a .
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 6 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99.
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7 2. Specifications and Model Selection Guide 2.1 Specifications Cell Constants 0.01, 0.1,1.0, 5, 10, 20, 25 and 50 cm -1 . Electrode Material Nickel or Platinum , as specified.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 8 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Integral Automatic Temperature Compensation Refer to Section 2.2 – Model Selection Guid e Leadwire Tefzel covered, 18-g age cable (7 feet, 20 f eet, or Universal Head, as specified).
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 9 049 TABLE IV Selection 09 08 Leadwire Length ’ 7 ft.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 10 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 049 TABLE VII - OPTIONS Tagging None 0_ Linen L_ Stainless Steel S_ Certificate of Calibration No _ 0 Yes _ 1 Notes: 1.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 11 3. Installation 3.1 Requirements To insur e that a re presenta tive sample is being measured at all times, the solutio n must m o ve thr ough and completely purge out the cell channels or guard tube.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 12 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Allow at least 1/2 inch clearance bey ond the end of th e cell and 1/8 to 3/16 inch radius clearance surrounding the cell to p e rmit circulation of the solutio n.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 13 3. Using a strap wrench on the m ounting nipple, tighten the assembly an additional 1-1/2 to 2 turns. Do not use stillson or chain type wrenches as they m ay damage and weaken the CPVC plastic.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 14 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 3.5 Removal 1. REDUCE PROCESS PRESSURE TO 50 PSIG OR LESS. Sh ut off purge line, if used. Dis connect wi ring conn ections if X1 is selected for Table IV in the MSG.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 15 2.5" (64mm) App r ox 4.1" (105mm) App r ox 12.5" (317mm) X " See Bel ow App r ox 19.7" max . (502mm) Allow a minim um clearance o f 36" (914 m m) beyond this poin t for removal of electrode.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 16 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 2. 5" (64mm) Ap prox 4 .1" (105mm) Approx 12. 5" (317mm) X " See Below A pprox 19.
Maintenance and Replacement Parts 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 17 4. Maintenance and Replacement Parts If a series of below normal condu ctivity readings or abov e normal resistiv ity readings occu r, this m ay indicate that the cell is not filled with process solution.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 18 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4.3 Replacement Parts Table 4-1 Replacement Parts Description Part Number Complete Ass embly See Figs.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 19 5. Platinization and Platinum Black Only the electrodes havi ng cons tants from 5 t.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 20 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 A D C B View o f J unct ion Box Head w ith Cap Re moved WH I T E BLA CK RED GREEN Cell Temp. Com p.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 21 A A D C B View o f J unct i on Box He ad with Cap Removed WH I T E BLA CK RED GREEN Cell Temp.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 22 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 WH I T E BLACK RED GREE N Cell Temp. Comp. WK S H G R Five Point Termin al Board. Each Terminal W i ll Accept #16 G age Max .
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 23 WH I T E BLA CK RED GREEN Cell Temp.
.
.
Industrial Auto mation and Co ntrol Honey well, Inc. 1100 Virginia D rive Fort Wa shington, Pennsy lvania 19034.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Honeywell 4909 CPVC è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Honeywell 4909 CPVC - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Honeywell 4909 CPVC imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Honeywell 4909 CPVC ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Honeywell 4909 CPVC, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Honeywell 4909 CPVC.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Honeywell 4909 CPVC. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Honeywell 4909 CPVC insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.