Manuale d’uso / di manutenzione del prodotto SIGMA 900 del fabbricante Hach
Vai alla pagina of 52
Catalog Number 8544 Sigma 900 Composite Sampler INSTR UMENT MANU AL © Hach Company , 2003. All rights reserved. Printed in the U .S.A. eac 07/0 3 2ed.
Visit http: //www .hach.com.
Pag e 3 8544T OC.fm T able of Contents Ta b l e o f C o n t e n t s Section 1 Specifications ............ ................ ................ ............. ................. ................ ................ ................ ........... 5 Section 2 General Inf ormation .
Pag e 4 T able of Conten ts 8544T OC.fm T able of Contents Section 5 Maintenance ........ ................ ................ ................ ................ ................ ............. ................ ................ 3 3 5.1 Cleaning the Sampler .
Pag e 5 8544specs.fm Specifications Section 1 Specifications Specifications are subject t o change without notice . Contr oller Dimensions Height: 34 cm (13.
Visit http: //www .hac h.com.
Pag e 7 8544intro .fm General Inf ormation Section 2 General Inf ormation 2.1 Instrument Description The instrument is desig ned to automatical ly collect and preserve samples from a liquid so urce. The sampler is suit able f or collection of con ventional and to xic pollutants and suspen ded solids.
Pag e 8 Front P anel 8544intro .fm Section 2 Figure 1 Interface Con nector s 2.3 Fr ont P anel The front panel cons ists of the k eypad , liquid cr ystal displa y , and the internal case humidity indicator ( Figure 2 ).
Section 2 Pag e 9 8544intro .fm Safety Information Figure 2 900 Composite Sampler 2.4 Safety Inf ormation Please read th is entire man ual before unpa ck ing, setting up , or oper ating this instru ment. P ay particular attention to all danger a nd caution statemen ts.
Pag e 1 0 Safety Info rmation 8544intro .fm Section 2 2.4.1 Use of Hazard Information If multiple hazards e xist, this manu al will use the signal word (Danger , Caution, Note) correspo nding to the gr eatest hazard.
Section 2 Pag e 1 1 8544intro .fm Safety Information D ANGER Bien que certains produits Hac h soient conçu s et certifiés pour être in stallés dans des endroits dangere ux tels que définis par l e National Electric Code , de nombreux produits Hach ne con viennent pa s pour de tels endroits.
Pag e 1 2 Safety Info rmation 8544intro .fm Section 2 2.4.5 Definition of Confined Space A Confined Space is an y location or enclosure t hat presents or has the immediate p otential to present on e or more o f the follo wing conditions : • An atmosphere with less tha n 19.
Pag e 1 3 8544install.fm Installation Section 3 Installation D ANGER Only qu alified per sonnel should conduct the install ation tasks described in t his section of the manual 3.1 Unpac king the Instrument Remov e the samp ler from th e shipping car ton and inspect it f or any damage.
Pag e 1 4 Mounting the Sampler 8544install.f m Section 3 Figure 4 W all Mount Brack et 3.3 Mounting the Sampler Attach the w all mount brack et to a stable pane l or wall ( Figure 4 ). Mount the brac ket so the sample displa y and ke ypad are easily accessible .
Section 3 Pag e 1 5 8544install.fm Setting Up the Intake Line and Strainer 4. Slide the tubing clamp ov er the silicone tubing until it is ov er the stainless steel fitting on t he liquid sensor . Tighten to secu re. 5. Repeat the proced ure f or the intake st rainer .
Pag e 1 6 Pow e r C o n n e c t i o n s 8544install.f m Section 3 Figure 5 Full Bottle Shut-off 3.8 P ower Connections The sampler controlle r operated on 12 V dc which is supplied by an optional ac/dc pow er con v er ter and/or battery . Note: Make sure both rubber hold- downs are attached to the clips at each end of the power supply .
Section 3 Pag e 1 7 8544install.fm 12 V dc • Bef ore performing an y maintenance , disconnect the sampler fro m the pow er source. • Do not attempt to mak e any connection or othe rwise handle the electrical components of the sampler wh en connected to ac line po wer if the immediate a rea is wet, or if hands or c lothing are wet.
Pag e 1 8 A uxiliary Connection 8544install.f m Section 3 Contact Closures The sampler controlle r is capab le of accum ulating 1 to 99 99 contact clo sures.
Section 3 Pag e 1 9 8544install.fm A uxiliary Conne ction Figure 6 4–20 mA Interface and Pu lse Duration Input F or details o n progr amming the sample r f or flow propor tional oper ation, ref er to SECTION. Splitter Interface This splitter interf ace (Cat.
Visit http: //www .hac h.com.
Pag e 2 1 8544operation s.fm Operation Section 4 Operation 4.1 Ke ypad and Key Functions Figure 8 shows the 18-p osition k eyp ad and 24-char acter alphan umeric displa y .
Pag e 2 2 Tips and T echniques for Programming the Instrument 8544operations .fm Section 4 Figure 8 Front P anel 4.2 Tips and T echniques for Pr ogramming the Instrument • The instrument beeps each time a k ey is pressed.
Section 4 Pag e 2 3 8544operation s.fm Setting the Ti me and Da te 4.3 Setting the Time and Date Bef ore programming t he sampler f or the first time, check the real time cloc k and date setting s. Note: When setting the time and date, the * ke y causes the display to back up to the pre viously fl ashing character .
Pag e 2 4 Explanations of Pr ogram Messages 8544operations .fm Section 4 4.5 Explanations of Pr ogram Messages Item Number Message Descr iption of Message 1 READ Y T O ST ART Indicates that a program is ready to star t. T o star t, press ST ART PROGRAM .
Section 4 Pag e 2 5 8544operation s.fm Data Logg ing 4.6 Data Logging The sampler records the tim e and date that each sa mple was collecte d, up to 400 samples.
Pag e 2 6 Measurements T riggered by Liquid Le vels or Flow 8544operations .fm Section 4 5. T o e x it the man ual mode, press CLEAR ENTR Y . This retur ns the sa mpler to the standb y state. While the sampler is in standb y , a samp le cyc le can be initiated by pressing the T AKE SAMPLE key .
Section 4 Pag e 2 7 8544operation s.fm Pr ogramming the Sampler for Le vel Contro l 4.9 Pr o gramming the Sampler f or Level Contr ol Put the sampler in a standb y state (i.e. "Prog ram Halted", "Ready T o Star t" or "Program Complete") hold CLEAR ENTR Y fo r two seconds.
Pag e 2 8 Flow Proportional Operation 8544operations .fm Section 4 4.11 Flo w Pr oportional Operation 4.11.1 Ho w to Calculate Pulses/Counts 4.11.1.1 Flo w Propo rtional Sampling Inter vals—Using Ex.
Section 4 Pag e 2 9 8544operation s.fm Flow Pr oporti onal Operatio n 1. Determine the flo w increment between samples: 2. Multiply the flow increment, f , by the pu lse frequency output o f the flo w meter . Theref ore, the v alue fo r INTV = _ _ _ _ CNTS is 71.
Pag e 3 0 Flow Proportional Operation 8544operations .fm Section 4 Use the f ollowing e xamples to help you dete r mine the v alue that you need to enter when prog ramming f or sampling based on counts . Example 1 Y ou want to collect 24 samples o ver a 24-hour period on a 4–20 mA flo w propor tional basis .
Section 4 Pag e 3 1 8544operation s.fm Flow Pr oporti onal Operatio n 2. Calculate t . 3. Multiply Q x t . Theref ore, the v alue entered f or INTV = _ _ _ _ CNTS is 4.
Visit http: //www .hac h.com.
Pag e 3 3 8544Maintenance .fm Maintenance Section 5 Maintenance D ANGER Only qu alified per sonnel should conduct the install ation tasks described in t his section of the manual 5.1 Cleaning the Sampler 5.1.1 Cleaning the Sampler Housing The interior and e xterior sampler housing ma y be cleaned with a damp sponge and mild deter gent.
Pag e 3 4 Re-installing the Bottom Panel 8544Maintenance .fm Section 5 4. Place the controller , face do wn, on a cloth-cov ered workbench to pre vent scratching the front panel. 5. Inside the Center Sec tion, remove the f our screws that fasten the Controller .
Section 5 Pag e 3 5 8544Maintenance .fm Internal Desiccant Module 5.8 Internal Desiccant Module The Internal Desiccant Module (Cat. No . 787) consists of a moisture-a bsorbent m ateri al inside a poly bag. I f the hum idity indicator on the front panel turns pink, replace the module and the bac k panel perimeter gask et.
Pag e 3 6 Motor/Gear Bo x 8544Maintenance .fm Section 5 5.10 Motor/Gear Bo x The motor/gear bo x requires no regular ma intenance . The special f or mulation gears are self lubricating and requir e no oil or grease .
Contact Inf ormation for U .S.A. & Outside Eur ope Pag e 3 7 Section 6 Contact Inf ormation for U .S.A. & Outside Europe Or dering Inf ormation f or the U .S.A. Informa tion Required Or dering Inf ormation f or Outside the U .S.A. and Europe Hach Company mainta ins a worldwide network of dealers and distributo rs.
Contact Inf ormation for Eur ope Pag e 3 8 Section 7 Contact Inf ormation for Eur ope F or technical suppo r t, repair ser vice, and or dering inf ormation please ref er to the contact inf or mation belo w f or y our specific country . F or additional inf ormation, visit www .
Limited W arranty Pag e 3 9 Section 8 Limited W arranty Hach Company w arrants its products to the original purchaser against an y def ects that are due to fau lty material or workmanship for a period of *one year* from date of shipment unless ot herwise noted in the product man ual.
Visit us at www .hach.com.
Pag e 4 1 8544parts.fm P ar ts and Accessories Section 9 P ar ts and Accessories Description Pa r t N u m b e r 3-wa y Splitter Assembly ............. ................ ................ ................ ................ ................ ........... .. .
Visit http: //www .hac h.com.
Pag e 4 3 8544apdx_pulsecounts .fm Programming Flo w Chart Appendix A Pr ogramming Flow Char t Pr ogram Flow Chart Ready to St art Program Halted Program Complete Bottle Setup - Press Enter units for.
Visit http: //www .hac h.com.
Pag e 4 5 8544apdx_batteries. fm Batteries and Chargers Appendix B Batteries and Char gers Lead-Acid (Gel Cell) Batteries Lead-acid batte ries are designed to pre vent electro lyte leakage from t he ter minals or c ase. The electrolyte is suspended in a gel, wh ich ensure s safe , efficient operation in any position.
Pag e 4 6 Batteries and Chargers 8544apdx_batteries .fm Appendix B Nic kel-Cadmium Batteries Nick el-cadmium batter ies provide superi or powe r capabilities when used at low temper atures. The y also perfo rm a higher number of charge/dischar ge cycles than lead-a cid batteries.
Pag e 4 7 8544apdx_engdra w .fm Exploded View Drawings Appendix C Exploded Vie w Drawings Exploded Vie w (1 of 6) Item Number Description QTY Pa r t N u m b e r 1 Pump Assemb ly 1 1075 2 Ov erlay 1 90.
Pag e 4 8 Exploded Vie w Draw ings 8544apdx_engdra w .fm Appendix C Exploded Vie w (2 of 6) Item Number Description QTY Pa r t N u m b e r 1 Latch 1 2279 2 K eypad Cable Assemb ly 1 8586 3 Screw 6 SE .
Appendix C Pag e 4 9 8544apdx_engdra w .fm Exploded View Drawings Exploded Vie w (3 of 6) Item Number Description QTY Pa r t Number Item Number Description QTY Pa r t Number 1 900 Controller Case 1 86.
Pag e 5 0 Exploded Vie w Draw ings 8544apdx_engdra w .fm Appendix C Exploded Vie w (4 of 6) Item Number Description QTY Pa r t Number Item Number Description QTY Pa r t Number 1 5 Amp Fuse Label 1 8537 6 He x Nut (tighten to 16 in.
Appendix C Pag e 5 1 8544apdx_engdra w .fm Exploded View Drawings Exploded Vie w (5 of 6) Item Number Description QTY Pa r t Number Item Number Description QTY Pa r t Number 1 Gearbo x Assembly 1 8910.
Pag e 5 2 Exploded Vie w Draw ings 8544apdx_engdra w .fm Appendix C Exploded Vie w (6 of 6) Item Number Description Pa r t N u m b e r 1 CPU PCB Assembly 8921 2 Distributor (J5) — 3 Auxiliary (J6) .
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Hach SIGMA 900 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Hach SIGMA 900 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Hach SIGMA 900 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Hach SIGMA 900 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Hach SIGMA 900, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Hach SIGMA 900.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Hach SIGMA 900. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Hach SIGMA 900 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.