Manuale d’uso / di manutenzione del prodotto 750000 del fabbricante Graham Field
Vai alla pagina of 20
750000 AL TERNA TING PRESSURE / L OW AIR L OSS MA TTRESS SYSTEM USER MANUAL Impor tant: Do not operate the Mattress System without rst reading and understanding this manual! Save this manual for future use. Info: The most current version of this manual can be found online at www.
2 750000-INS-LAB-RevC11 CONTENTS INTRODUCTION..................................................................................... 3 INTENDEDUSEOFTHISDEVICE ........................................................ 3 CONTRAINDICA TION .
750000-INS-LAB-RevC11 3 INTR ODUCTION Use this manual for set-up and operation of the Lumex AltaDyne 750000 Alternating Pr essure / Low Air Loss Mattr ess System. Read all instructions before using the mattr ess system. Save this manual for future r eference.
4 750000-INS-LAB-RevC11 IMPORT ANT SAFETY PRECA UTIONS The safety statements pr esented in this chapter refer to the basic safety information that should be obser ved by those using the Alternating Pr essure / Low Air Loss Mattr ess System.
750000-INS-LAB-RevC11 5 WARNING To reducetheriskofburns,electrocution,r e,or personalinjury: W ARNING:Donotleavethisproductunattended whenpluggedin.
6 750000-INS-LAB-RevC11 W ARNING:Patiententrapmentwithbedsiderails maycauseinjuryordeath.Thebedframeandits components,includingthemat.
750000-INS-LAB-RevC11 7 EQUIPMENT SYMBOLS Attention! Consult accompanying documents (this manual) “BF”symbol: Indicates that this product is in accordance with the degree of protection against .
8 750000-INS-LAB-RevC11 PR ODUCT DESCRIPTION PUMP AND MA TTRESS SY STEM pumpandmattresssystem CPRvalves mattress-pump connectors mattress footend symbol pump Pump: front quick connec.
750000-INS-LAB-RevC11 9 Pump: rear Pump: top panel pressur eadjustmentknob normalpressur eindicator lowpressur eindicator mainpowerswitch staticfunctionswitch Pr essur e adjust knob : The pressur e adjust knob controls the air pr essure output.
10 750000-INS-LAB-RevC11 Static function switch : Pr ess the static function switch to suspend alternating function, if needed; the pr essure inside the air tubes will then be adjusted to identical pr essure. Pr ess the static function switch again to change back to alternating mode.
750000-INS-LAB-RevC11 11 INST ALLA TION UNP ACKING 1. Befor e unpacking, check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment pr ovider . 2. Remove all loose packing fr om the car ton.
12 750000-INS-LAB-RevC11 PUMP AND MA TTRESS INST ALLA TION 1. Place the mattr ess on top of the bed frame; ensur e that the foot end symbol faces up at foot end as shown at right. 2. Fold open the wir e hangers on the back of the pump as shown at right.
750000-INS-LAB-RevC11 13 W ARNING:Locateallcordssothattheywillnotbe steppedon,trippedover ,orotherwisesubjectedto damageorstress. W ARNING:Donotuseahouseholdextensioncordif theelectricalcorddoesnotreachthepoweroutlet.
14 750000-INS-LAB-RevC11 OPERA TION 1. T urn the power switch to the ON position. 2. Adjust the pr essure adjust knob to the maximum position. Info: Set the pressure adjust knob to the maximum position for the quickest inflation at set-up; the mattress can then be adjusted to the desired pressure.
750000-INS-LAB-RevC11 15 under the patient's buttock ar ea; your ngers should just t. If the t is too loose, decr ease the setting. If your ngers do not t well, incr ease the setting two notches rmer . W ait ten minutes and then check again.
16 750000-INS-LAB-RevC11 MAINTENANCE CLEANING Perform the following cleaning pr ocedures befor e use. Pump : W ipe the pump with a clean, damp cloth and mild deter gent, and keep it away fr om dust. Air dr y . s NOTICE:Donotusephenolicproductsorcorrosive orpowderedcleanserstocleanthepump.
750000-INS-LAB-RevC11 17 Fuse replacement tool needed: small screwdriver 1. If you suspect a blown fuse, disconnect the plug fr om the wall outlet immediately . 2. Use a small scr ewdriver to r emove the cover of the fuse holder (fuse locations ar e shown in pictur e at right).
18 750000-INS-LAB-RevC11 SPECIFICA TIONS Pump Specication Power Supply 120V System: AC 120V 60 Hz Info: see rating label on product Fuse Rating T1A, 250V Cycle time 8 minutes Air input >4 LPM Dimensions (L x W x H) 9.8" x 5.3" x 3.7" (25 x 13.
750000-INS-LAB-RevC11 19 WARRANTY GF Health Pr oducts, Inc. warrants the Lumex AltaDyne 750000 Alternating Pr essure / Low Air Loss Mattr ess System for a period of twelve months for defects in workmanship and materials. If a pr oduct is deemed to be under warranty , GF Health Products, Inc.
USA Corporate Headquar ters: GF Health Products, Inc. 2935 Nor theast Parkway Atlanta, Georgia 30360 telephone: 770-368-4700 fax: 770-368-2386 www .grahamfield.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Graham Field 750000 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Graham Field 750000 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Graham Field 750000 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Graham Field 750000 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Graham Field 750000, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Graham Field 750000.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Graham Field 750000. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Graham Field 750000 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.