Manuale d’uso / di manutenzione del prodotto 3081 pH/ORP del fabbricante Emerson Process Management
Vai alla pagina of 150
Model 3081 pH/ORP Smart T wo-Wire Microprocessor T ransmitter Instruction Manual PN 51-3081pH/rev .D August 2002.
ESSENTIAL INSTRUCTIONS READ THIS P AGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures, and tests it s products to meet many national and international standards.
WHA T YOU NEED T O KNOW BEFORE INST ALLING AND WIRING A ROSEMOUNT ANAL YTICAL SENSOR T O THE MODEL 3081 pH/ORP TRANSMITTER 1. THE MODEL NUMBER OF THE SENSOR • Look on the label. • Also note the model option string. • If the label is missing or unreadable, see the flowcharts on p ages 28 through 30.
• If the sensor is wired directly to the transmitter , the preamplifier can be in either the sensor or the transmitter . • Look at the wires in the sensor cable. A GREEN wire means the preamplifier is in the sen- sor . An coaxial cable means the preamplifier is in the transmitter .
QUICK ST ART GUIDE FOR MODEL 3081pH/ORP Before using this Quick St art Guide, please read “WHA T YOU NEED T O KNOW BEFORE INST ALLING AND WIRING A ROSEMOUNT ANAL YTICAL SENSOR TO THE MODEL 3081 pH/ORP TRANSMITTER” on the preceding p age. Section 1.
Section 2 Wiring A. Wire sensor Model 381+-55, 385+-03, or 396P-01-55 directly to the transmitter as shown in Figure 3. B. Wire sensor Model 381+-52, 385+-04, 396P-02-55, 396P-02-55, or 396R-54 as shown in Figure 4. C. Wire the 12 - 42.4 Vdc power supply to TB-15 (- 4 - 20 mA) and TB-16 (+ 4 - 20 mA).
QUICK REFERENCE GUIDE MODEL 3081PH/ORP Automatic Buffer Calibration Note: A pH measurement is only as good as the calibration, and the calibration is only as good as the buffers used. For best results, calibrate with buf fers having the same temperature as the process.
HART Communicator Fast Key Sequences T echnical Support Hotline: For assistance with technical problems, please call the Customer Support Center (CSC). The CSC is staffed from 5:00am to 5:00pm PST . Phone (US only): 800-854-8257 Phone: 949-757-8500 Fax: 949-863-9159 World W ide Web: www .
i MODEL 3081 pH/ORP T ABLE OF CONTENTS MODEL 3081 PH/ORP MICROPROCESSOR TRANSMITTER T ABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICA TIONS ................................................................ 1 1.1 Features .............
MODEL 3081 pH/ORP T ABLE OF CONTENTS T ABLE OF CONTENTS CONT’D ii 8.0 PROGRAMMING FOR pH MEASUREMENTS ....................................................... 56 8.1 General ............................................................................
iii MODEL 3081 pH/ORP T ABLE OF CONTENTS T ABLE OF CONTENTS CONT’D 13.0 pH MEASUREMENTS ............................................................................................. 1 14 13.1 General .................................................
iv MODEL 3081 pH/ORP T ABLE OF CONTENTS T ABLE OF CONTENTS CONT’D LIST OF FIGURES Number Title Page 2-1 Model 3081 pH/ORP T ransmitter - Exploded Drawing of Circuit Board S tack ......... 8 2-2 Model 3081 pH/ORP T ransmitter - Analog Board .........
v MODEL 3081 pH/ORP T ABLE OF CONTENTS T ABLE OF CONTENTS - CONT’D LIST OF FIGURES - CONT’D Number Title Page 3-23 Procedure for Removing BNC Connector and Preparing Coaxial Cable for ............ Connection to the Model 3081 pH/ORP Transmitter ..
vi MODEL 3081 pH/ORP T ABLE OF CONTENTS T ABLE OF CONTENTS CONT’D LIST OF T ABLES Number Title Page 3-1 Wiring Diagrams for Model 399 Sensors .................................................................. 15 3-2 Wiring Diagrams for Model 397 Sensors .
1 MODEL 3081 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.1 Features 1.2 Accessories 1.3 Specifications - General for Model 3081 pH/ORP 1.4 Specifications - pH 1.5 Specifications - ORP 1.6 Ordering Information • CHANGING FROM pH T O ORP operation takes only seconds.
2 CALIBRA TION: T wo-point, temperature-corrected buffer calibration is standard. T o reduce errors caused by impatient operators, the Model 3081 pH/ORP transmitter does not accept calibration data until programmed st abil- ity limits have been met.
3 MODEL 3081 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS minimum voltage and load for HART communication 1.3 SPECIFICA TIONS - GENERAL FOR MODEL 3081 pH/ORP Case: Cast aluminum containing less than 8% magnesium. NEMA 4X (IP65), NEMA 7 (explosion-proof) Epoxy-polyester paint, Neoprene O-ring seals Dimensions: 6.
4 MODEL 3081 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.6 ORDERING INFORMA TION The Model 3081 pH/ORP Smart two-wire microprocessor transmitter is housed in a NEMA 4X case. Communication with the transmitter is through a hand-held infrared remote controller , a Model 275 HART communicator , or any HART com- patible device.
5 MODEL 3081 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS MODEL 3081 pH/ORP TRANSMITTER-SENSOR COMP A TIBILITY CHART PREAMPLIFIER LOCA TION Sensor-mounted Remote MODEL pH ORP Sensor junction box.
6 MODEL 3081 pH/ORP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS MODEL/PN DESCRIPTION SHIPPING WEIGHT 515 DC loop power supply , see Section 1.2.1 for details 3 lb/1.
2.1 UNP ACKING AND INSPECTION Inspect the shipping container . If it is damaged, cont act the shipper immediately for instructions. Save the box. If there is no apparent damage, remove the transmitter . Be sure all items shown on the p acking list are present.
MODEL 3081 pH/ORP SECTION 2.0 INST ALLA TION 8 FIGURE 2-1. Model 3081 pH/ORP T ransmitter - Exploded Drawing of Circuit Board St ack 2.2.4 Preamplifier Location pH sensors produce a high impedance voltage signal that must be preamplified before use.
9 5. Set the jumpers and the slide switch on the analog board. Refer to Figure 2-2. a. T emperature element jumper . Jumper position T emperature element JP-1 Pt 1000 RTD JP-2 Pt 100 RTD JP-3 3K Balco RTD b. Reference impedance jumper . Jumper position Reference impedance JP-6 low JP-7 high c.
10 MODEL 3081 pH/ORP SECTION 2.0 INST ALLA TION 6. There are more jumpers on the CPU board. Refer to Figure 2-3. These jumpers are factory set and should NOT need to be moved. This step is for trou- bleshooting purposes only . V erify that jumpers JP-1, JP-3, and JP-4 on the CPU board are in the positions shown in Figure 2-3.
11 MODEL 3081 pH/ORP SECTION 2.0 INST ALLA TION FIGURE 2-4. Mounting the Model 3081 pH/ORP T ransmitter on a Flat Surface MILLIMETER INCH 4. Use weathertight cable glands to keep moisture out of the transmitter . 5. If conduit is used, plug and seal the connections at the transmitter housing to prevent moisture from getting inside the transmitter .
12 MODEL 3081 pH/ORP SECTION 2.0 INST ALLA TION FIGURE 2-5. Using the Pipe Mounting Kit to Attach the Model 3081 pH/ORP T ransmitter to a Pipe MILLIMETER INCH 2.4.3 Pipe Mounting. See Figure 2-5. The pipe mounting kit (PN 2002577) accommodates 1-1/2 to 2 in.
13 MODEL 3081 pH/ORP SECTION 2.0 INST ALLA TION 2.5 POWER SUPPL Y/CURRENT LOOP 2.5.1 Power Supply and Load Requirement s. Refer to Figure 2-6. The minimum power supply voltage is 12.5 Vdc and the maximum is 42.4 Vdc. The top line on the graph gives the voltage required to maint ain at least 12.
14 MODEL 3081 pH/ORP SECTION 3.0 WIRING SECTION 3.0 WIRING 3.1 GENERAL INFORMA TION pH and ORP sensors manufactured by Rosemount Analytical can be wired to the Model 3081 pH/ORP transmitter in three ways: 1. directly to the transmitter , 2. to a sensor-mounted junction box and then to the transmitter , 3.
MODEL 3081 pH/ORP SECTION 3.0 WIRING 3.2 WIRING DIAGRAMS FOR pH and ORP SENSORS Refer to T ables 3-1 through 3-12 to locate the appropriate wire function and wiring diagram. There is a separate t able for each model. The sensor models having the highest number appear first.
16 MODEL 3081 pH/ORP SECTION 3.0 WIRING T able 3-4 W iring Diagrams for Model 396P Sensors Sensor Junction Box Preamplifier RTD W ire Function Wiring Diagram 396P-01-55 none in sensor Pt 100 Figure 3-.
17 MODEL 3081 pH/ORP SECTION 3.0 WIRING T able 3-7 W iring Diagrams for Model 385+ Sensors Sensor Junction Box Preamplifier RTD W ire Functions Wiring Diagram 385+ -02 sensor-mounted in sensor-mounted.
18 MODEL 3081 pH/ORP SECTION 3.0 WIRING REMOVE BNC AND TERMINA TE COAXIAL CABLE BEFORE WIRING SENSOR TO TRANSMITTER. SEE FIGURE 3-23. AL TERNA TIVEL Y , USE A BNC ADAPTER (PN 9120531) OR ORDER MODEL OPTION -62 (SENSOR WITH BNC REMOVED AND TER- MINA TIONS COMP A TIBLE WITH 3081 pH/ORP).
19 MODEL 3081 pH/ORP SECTION 3.0 WIRING REMOVE BNC AND TERMINA TE COAXIAL CABLE BEFORE WIRING SENSOR TO TRANSMITTER. SEE FIGURE 3-23. AL TERNA TIVEL Y , USE A BNC ADAPTER (PN 9120531) OR ORDER MODEL OPTION -62 (SENSOR WITH BNC REMOVED AND TERMINA TIONS COMP A TIBLE WITH 3081 pH/ORP).
20 MODEL 3081 pH/ORP SECTION 3.0 WIRING FIGURE 3-10. Wire functions for Models 396R-50, 396R-54, 396R-54-61, 396P-02-50, 396P-02-54, 396P-02-55, 385+ -04, and 385+ -41-52. NOTES: 1. PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "TRANS- MITTER" POSITION (SEE SECTION 2.
21 MODEL 3081 pH/ORP SECTION 3.0 WIRING FIGURE 3-13. Wire functions for Models 396P-01-55, 385+ -03, 381+ -40-55, and 381+ -43-55. NOTES: 1. PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "SENSOR/JUNCTION BOX" POSITION (SEE SECTION 2.
22 MODEL 3081 pH/ORP SECTION 3.0 WIRING FIGURE 3-15. Wire functions for Model 385+ -02. NOTES: 1. PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "SENSOR/JUNCTION BOX" POSITION (SEE SECTION 2.2). 2. SEE FIGURE 3-24 FOR TERMINA TION OF RAW INTERCONNECTING CABLE (PN 9200273).
23 MODEL 3081 pH/ORP SECTION 3.0 WIRING NOTE: MODEL 328A-08 HAS UNDRESSED COAXIAL CABLE. SEE SENSOR INSTRUCTION SHEET FOR CABLE PREP ARA TION PROCEDURE. FIGURE 3-17. Wire functions for Model 328A-07. NOTES: 1. PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "TRANS- MITTER" POSITION (SEE SECTION 2.
24 MODEL 3081 pH/ORP SECTION 3.0 WIRING NOTES: 1. PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "TRANS- MITTER" POSITION (SEE SECTION 2.2). 2. TERMINALS IN JUNCTION BOX ARE NOT NUMBERED. COUNT POSITION FROM LEFT TO RIGHT AS SHOWN IN DRA WING . 3.
25 MODEL 3081 pH/ORP SECTION 3.0 WIRING FIGURE 3-21. Wire Functions for Model 399-33 FIGURE 3-22. Wiring Diagram for Model 399-33.
26 MODEL 3081 pH/ORP SECTION 3.0 WIRING FIGURE 3-23. Procedure for Removing BNC Connector and Preparing Coaxial Cable.
27 MODEL 3081 pH/ORP SECTION 3.0 WIRING FIGURE 3-24. Preparation of Raw Connecting Cable (PN 9200273)..
28 MODEL 3081 pH/ORP SECTION 3.0 WIRING SENSOR FLOW CHART (continued on page 29).
29 MODEL 3081 pH/ORP SECTION 3.0 WIRING SENSOR FLOW CHART (continued on page 30).
30 MODEL 3081 pH/ORP SECTION 3.0 WIRING SENSOR FLOW CHART.
31 4.1 INTRINSICALL Y SAFE INST ALLA TIONS The installation wiring, operating p arameters, or agency tags for intrinsically safe operation are given in Figures 4-1, 4-2, and 4-3. 4.2 EXPLOSION PROOF INST ALLA TIONS The installation wiring and operating p arameters for explosion proof operation are given in Figure 4-4.
32 FIGURE 4-2. FMRC Inst allation for Model 3081 pH/ORP T ransmitter.
33.
FIGURE 4-3. CSA Installation for Model 3081 pH/ORP T ransmitter 34.
35.
FIGURE 4-4. Explosion-Proof Inst allation for Model 3081 pH/ORP T ransmitter 36.
37 MODEL 3081 pH/ORP SECTION 5.0 OPERA TION WITH REMOTE CONTROLLER SECTION 5.0 OPERA TION WITH REMOTE CONTROLLER CALIBRA TE PROGRAM DIAGNOSE CALIbrAtE EXIT NEXT ENTER 7.00 pH MV F A U L T H O L D ♥ Appears when transmitter is in hold (see Section 8.
38 MODEL 3081 pH/ORP SECTION 5.0 OPERA TION WITH REMOTE CONTROLLER 5.2 INFRARED REMOTE CONTROLLER (IRC) - KEY FUNCTIONS The infrared remote controller is used to calibrate and program the transmitter and to read diagnostic messages. See Figure 5-3 for a description of the function of the keys.
39 MODEL 3081 pH/ORP SECTION 5.0 OPERA TION WITH REMOTE CONTROLLER PROGRAM CALIBRA TE OutPut GIMP 1000 V Er 81PH.21 tEMP 25 C InPut 58.9 ShoW FLt nonE rIMP 10 dIAGnOStIC tEMP bUFFEr dISPLA Y ISOPOtntAL SIM OUtPUt CALIbrAtE Std tEMP AdJ tEMP 25.0 4 MA 00.
40 5.5 MENU TREE - ORP The Model 3081 ORP transmitter has three menus: CALIBRA TE, PROGRAM, and DIAGNOSE. Under the Calibrate and Program menus are several sub-menus. For example, under CALIBRA TE, the sub-menus are St d (standard) and tEMP AdJ (temperature adjust).
41 5.7 SECURITY 5.7.1 General. Use the programmable security code to protect program and calibration settings from accidentally being changed. The transmitter is shipped with the security fea- ture disabled. T o program a security code, refer to Section 8.
42 6.1 Note on Model 275 HART Communicator The Model 275 HART Communicator is a product of Rosemount Measurement. This section contains selected information on using the Model 275 with the Rosemount Analytical Model 3081 pH/ORP T ransmitter . For complete information on the Model 275 Communicator , see the Model 275 instruction manual.
43 MODEL 3081 pH/ORP SECTION 6.0 OPERA TION WITH MODEL 275 6.3 Operation 6.3.1 Off-line and On-line Operation The Model 275 Communicator features off-line and on-line communications. On-line means the communicator is connected to the transmitter in the usual fashion.
44 MODEL 3081 pH/ORP SECTION 6.0 OPERA TION WITH MODEL 275 FIGURE 6-2a. pH Menu T ree (HART) (Figure 6-2 is continued on page 41).
45 MODEL 3081 pH/ORP SECTION 6.0 OPERA TION WITH MODEL 275 FIGURE 6-2b. pH Menu T ree (HART).
46 MODEL 3081 pH/ORP SECTION 6.0 OPERA TION WITH MODEL 275 FIGURE 6-3a. ORP Menu T ree (HART) (Figure 6-3 is continued on page 43).
47 MODEL 3081 pH/ORP SECTION 6.0 OPERA TION WITH MODEL 275 FIGURE 6-3b. ORP Menu T ree (HART).
48 MODEL 3081 pH/ORP SECTION 7.0 CALIBRA TION OF pH MEASUREMENTS SECTION 7.0 CALIBRA TION OF pH MEASUREMENTS 7.1 General 7.2 Entering and Leaving the Calibrate Menu 7.3 Using the Hold Function 7.4 T emperature Calibration 7.5 Auto Calibration 7.6 Manual Calibration 7.
49 CALIBRA TE tEMP A A d J EXIT NEXT ENTER CALIBRA TE tEMP EXIT ENTER 0 25.0 MODEL 3081 pH/ORP SECTION 7.0 CALIBRA TION OF pH MEASUREMENTS 7.4 TEMPERA TURE CALIBRA TION 7.4.1 Purpose 1. As discussed in Section 13.6, Glass Electrode Slope, measuring tempera- ture is an important p art of measuring pH.
50 7.5 AUTO CALIBRA TION 7.5.1 Purpose 1. New sensors must be calibrated before use. Regular recalibration is also necessary . 2. The use of auto calibration instead of manual calibration is strongly recommended. Auto calibra- tion avoids common pitfalls and reduces errors.
7.5.4 Procedure 1. Refer to Section 8.7, Buffer Calibration Parameters. V erify that auto calibration is activated. Identify the buffers being used and set the st ability limits. 2. Enter the CALIBRA TE menu by pressing CAL on the IRC. The CALIbrAtE sub-menu appears (pic- tured above left).
52 MODEL 3081 pH/ORP SECTION 7.0 CALIBRA TION OF pH MEASUREMENTS 7.6.1 Purpose 1. New sensors must be calibrated before use. Regular recalibration is also necessary . 2. Manual calibration is an alternative to auto calibration. Because auto calibration eliminates many common calibration errors, it is strongly recommended.
53 7.6.3 Procedure 1. Before starting, refer to Section 8.7, Buffer Calibration Parameters, to deactivate auto cali- bration. 2. Enter the CALIBRA TE menu by pressing CAL on the IRC. The CALIbrAtE sub-menu appears (pictured above left). 3. At the CALIbrAtE sub-menu, press ENTER .
54 MODEL 3081 pH/ORP SECTION 7.0 CALIBRA TION OF pH MEASUREMENTS 7.7 MAKING THE TRANSMITTER READING MA TCH A SECOND pH METER (ST ANDARDIZA TION). 7.7.1 Purpose 1. This section describes how to make the transmitter reading match the reading from a second pH meter .
55 MODEL 3081 pH/ORP SECTION 7.0 CALIBRA TION OF pH MEASUREMENTS CALIBRA TE Std EXIT NEXT ENTER CALIBRA TE SLOPE EXIT NEXT ENTER CALIBRA TE Std EXIT NEXT ENTER 08 .00 5 9.00 7.7.3 Procedure 1. Enter the CALIBRA TE menu by pressing CAL on the IRC. The CALibrAtE sub-menu appears (pictured above left).
56 MODEL 3081 pH/ORP SECTION 8.0 PROGRAMMING FOR pH MEASUREMENTS SECTION 8.0 PROGRAMMING FOR pH MEASUREMENTS 8.1 General 8.2 Entering and Leaving the Program Menu 8.3 Output Ranging 8.4 Diagnostic Parameters 8.5 T emperature Related Settings 8.6 Display Unit s 8.
57 ITEM MNEMONIC DISPLA Y LIMITS F ACTOR Y SETTINGS USER SETTINGS PROGRAM LEVEL (Sections 8.0 - 8.9). A. Output Range (Section 8.3) OutPut 1. 4 mA Output 0 - 14 pH 0.00 pH _______ 2. 20 mA Output: 0 - 14 pH 14.00 pH _______ 3. Hold HOLd 3.80 to 22.00 mA 21.
58 MODEL 3081 pH/ORP SECTION 8.0 PROGRAMMING FOR pH MEASUREMENTS 8.3 OUTPUT RANGING 8.3.1 Purpose This section describes how to do the following: 1. assign pH values to the 4 and 20 mA outputs, 2. set the output current generated by the transmitter during hold, 3.
59 8.3.3 Procedure 1. Enter the Program menu by pressing PROG on the IRC. The OutPut sub-menu appears. 2. Press ENTER . The screen displays the 4 MA prompt. Use the editing keys to change the displayed number to the desired pH. The allowed range is 0.
60 8.4 DIAGNOSTIC P ARAMETERS 8.4.1 Purpose This section describes how to do the following: 1. change the standardization or reference of fset, 2. enable and disable sensor diagnostics, 3. enable and disable glass impedance temperature compensation, 4.
61 4. REFERENCE IMPEDANCE. The majori- ty of reference electrodes used in indus- try are low impedance silver-silver chlo- ride electrodes. Every pH and ORP sen- sor manufactured by Rosemount Analytical has a low impedance refer- ence. However , there are applications that call for either a high impedance sodi- um or pH glass reference electrode.
62 8.4.3 Procedure 1. Press PROG on the infrared remote controller (IRC). 2. Press NEXT until the diAGnOStIC sub-menu appear . Press ENTER . 3. The screen displays the rOFFSt prompt. Use the editing keys to change the flashing display to the desired standardization (reference) of fset (in millivolts).
63 12. The rFH prompt appears. Use the editing keys to change the display to the desired ref- erence electrode high impedance fault value. The allowed ranges are T ype of reference electrode Allowed range Low impedance ( LO in step 1 1) 0 - 2000 kilohms High impedance ( HI in step 1 1) 0 - 2000 megohms Entering 0000 disables the feature.
64 MODEL 3081 pH/ORP SECTION 8.0 PROGRAMMING FOR pH MEASUREMENTS 8.5 TEMPERA TURE RELA TED SETTINGS 8.5.1 Purpose This section describes how to do the following: 1. activate and deactivate automatic temperature compensation, 2. set a manual temperature compensation value, 3.
65 8.5.3 Procedure 1. Press PROG on the infrared remote controller (IRC). 2. Press NEXT until the tEMP sub-menu appears in the display . Press ENTER . 3. The screen displays the tAUTO prompt. Press Ï or Ð to enable ( On ) or disable ( OFF ) automatic temperature compensation.
66 8.6 DISPLA Y UNITS 8.6.1 Purpose This section describes how to do the following: 1. switch the process display units between pH and ORP (millivolts), 2. select °C or °F for the temperature display , 3. select percent of full scale or milliamps for the output display , 4.
67 8.7 BUFFER CALIBRA TION P ARAMETERS 8.7.1 Purpose This section describes how to do the following: 1. activate or deactivate auto calibration, 2. identify which buffers will be used during auto calibration, 3. set the stability criteria for auto calibration.
68 3. ST ABILITY CRITERIA. For the Model 3081 pH/ORP transmitter to accept calibration data, the pH must remain within a specified range for a specified period of time. The default values are 0.02 pH units and 10 seconds. In other words, at the default setting, calibration data will be accepted as soon as the pH reading is const ant to within 0.
69 MODEL 3081 pH/ORP SECTION 8.0 PROGRAMMING FOR pH MEASUREMENTS 8.8 ISOPOTENTIAL P ARAMETERS 8.8.1 Purpose This section describes how to do the following: 1. convert the pH at the measurement temperature to the pH at a reference temperature by entering a solution temperature coefficient, 2.
70 8.8.3 Procedure 1. Press PROG on the infrared remote controller (IRC). 2. Press NEXT until the ISOPOtntAL sub-menu appears. Press ENTER . 3. The screen displays the tCOEFF prompt. Use the editing keys to change the display to the desired solution temperature coefficient.
71 PROGRAM tESt EXIT ENTER 1 2.00 PROGRAM SIM O O UtPUt EXIT NEXT ENTER MODEL 3081 pH/ORP SECTION 8.0 PROGRAMMING FOR pH MEASUREMENTS 8.9 GENERA TING A TEST CURRENT 8.9.1 Purpose This section describes how to generate output currents for testing recorders and dat a han- dling systems.
72 MODEL 3081 pH/ORP SECTION 9.0 CALIBRA TION OF ORP MEASUREMENTS SECTION 9.0 CALIBRA TION OF ORP MEASUREMENTS 9.1 General 9.2 Entering and Leaving the Calibrate Menu 9.
73 CALIBRA TE tEMP A A d J EXIT NEXT ENTER CALIBRA TE tEMP EXIT ENTER 0 25.0 MODEL 3081 pH/ORP SECTION 9.0 CALIBRA TION OF ORP MEASUREMENTS 9.4 TEMPERA TURE CALIBRA TION 9.4.1 Purpose 1. As discussed in Section 14.6 (ORP , Concentration, and pH), ORP is a function of temperature.
74 9.5.3 Procedure 1. Place the transmitter in ORP mode. See Section 10.6.3, steps 1 - 3. Af ter selecting and saving OrP , press EXIT twice to return to the main display . 2. Enter the CALIBRA TE menu by pressing CAL on the IRC. The St d sub- menu appears.
75 10.1 GENERAL This section describes how to do the following: 1. assign ORP values to the 4 and 20 mA outputs, 2. set the current generated by the transmitter during hold, 3. set the current generated by the transmitter when a fault is detected, 4. change sensor diagnostic limits, 5.
76 MODEL 3081 pH/ORP SECTION 10.0 PROGRAMMING FOR ORP MEASUREMENTS ITEM MNEMONIC DISPLA Y F ACTOR Y USER LIMITS SETTINGS SETTINGS PROGRAM LEVEL A. Output Range (Section 10.3) OutPut 1. 4 mA Output -1400 to 1400 mV -1400 mV _______ 2. 20 mA Output: -1400 to 1400 mV 1400 mV _______ 3.
77 MODEL 3081 pH/ORP SECTION 10.0 PROGRAMMING FOR ORP MEASUREMENTS 10.3 OUTPUT RANGING 10.3.1 Purpose This section describes how to do the following: 1. assign ORP values to the 4 and 20 mA outputs, 2. set the output current generated by the transmitter during hold, 3.
78 MODEL 3081 pH/ORP SECTION 10.0 PROGRAMMING FOR ORP MEASUREMENTS 10.3.3 Procedure 1. Enter the Program menu by pressing PROG on the IRC. The OutPut sub-menu appears. 2 . Press ENTER. The screen displays the 4 MA prompt. Use the editing keys to change the displayed number to the desired ORP .
79 MODEL 3081 pH/ORP SECTION 10.0 PROGRAMMING FOR ORP MEASUREMENTS 10.4 DIAGNOSTIC P ARAMETERS 10.4.1 Purpose This section describes how to do the following: 1. change the standardization (reference) of fset, 2. enable and disable sensor diagnostics, 3.
80 MODEL 3081 pH/ORP SECTION 10.0 PROGRAMMING FOR ORP MEASUREMENTS PROGRAM rEF EXIT ENTER L O PROGRAM IMPtC EXIT NEXT ENTER O N PROGRAM rOFFSt EXIT ENTER PROGRAM dIAGnOSTIC EXIT NEXT ENTER 0 6 0 PROGRAM dIAG EXIT ENTER OFF 10.4.3 Procedure 1. Press PROG on the infrared remote controller (IRC).
81 8. The rWH prompt appears. In the display , W appears as wj . Use the editing keys to change the display to the desired reference electrode high impedance warn- ing value.
82 MODEL 3081 pH/ORP SECTION 10.0 PROGRAMMING FOR ORP MEASUREMENTS 10.5 TEMPERA TURE ELEMENT 10.5.1 Purpose This section describes how to match the transmitter to the type of temperature element in the ORP sensor . 10.5.2 Definition TEMPERA TURE ELEMENT : ORP sensors use a variety of temperature ele- ments.
83 MODEL 3081 pH/ORP SECTION 10.0 PROGRAMMING FOR ORP MEASUREMENTS PROGRAM OUtPUt EXIT ENTER CUR PROGRAM CODE EXIT ENTER 0 0 0 PROGRAM tYPE EXIT ENTER P H PROGRAM tEMP EXIT ENTER C PROGRAM dISPLAY EXIT NEXT ENTER 10.6 DISPLA Y UNITS 10.6.1 Purpose This section describes how to do the following: 1.
84 MODEL 3081 pH/ORP SECTION 10.0 PROGRAMMING FOR ORP MEASUREMENTS PROGRAM tESt EXIT ENTER 1 2 . 0 0 PROGRAM SIM O O UtPUt EXIT NEXT ENTER 10.7 GENERA TING A TEST CURRENT 10.7.1 Purpose This section describes how to generate output currents for testing recorders and data handling systems.
85 MODEL 3081 pH/ORP SECTION 1 1.0 MAINTENANCE SECTION 1 1.0 MAINTENANCE 1 1.1 OVERVIEW This section gives general procedures for routine maintenance of the 3081 pH/ORP transmitter and pH and ORP sensors. The transmitter needs almost no routine maintenance.
86 MODEL 3081 pH/ORP SECTION 1 1.0 MAINTENANCE T ABLE 1 1-1. Replacement Parts for Model 3081 pH/ORP T ransmitter PROBLEM CLEANING SUGGESTIONS Loose scale or debris Use a stream of water from a wash bottle to rinse away solids from the tip of the sensor .
87 When using acid or alkaline solvents, be careful to keep the solvent away from the liquid junction. If the cleaning sol- vent contact s the junction, hydrogen ions (acid solvent) or hydroxide ions (alkaline solvent) will diffuse into the junc- tion.
88 1 1.4 ORP SENSOR MAINTENANCE 1 1.4.1 Frequency of Cleaning The frequency at which an ORP sensor should be inspected and cleaned can be determined only by experience. If the process liquid coats or fouls the sensor , frequent cleaning may be necessary .
89 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING SECTION 12.0 TROUBLESHOOTING 12.1 W ARNING AND F AUL T MESSAGES 12.2 CALIBRA TION ERRORS 12.3 TROUBLESHOOTING - GENERAL 12.4 TROUBLESHOOTING WHEN A DIAGNOSTIC MESSAGE IS SHOWING 12.5 TROUBLESHOOTING WHEN NO DIAGNOSTIC MESSAGE IS SHOWING 12.
90 12.2 CALIBRA TION ERRORS If an error occurs during calibration, an error message appears in the main display , and the transmitter does not update the calibration. The calibration errors are S td Err , SLOPE Err LO , and SLOPE Err HI . See Section 12.
91 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.4.1 GLASSF AIL GLASSF AIL is an electrode fault message. It means the glass impedance is outside the programmed Glass Fault High ( GFH ) or Glass Fault Low ( GFL ) limit. Glass Fault High suggests the electrode is aging or the electrode is not immersed in the process liquid.
92 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING b. If the sensor was wired with the blue solution ground wire unattached and a jumper between terminals TB-8 and TB-7, remove the jumper and reatt ach the blue solution ground wire to TB-8. Keep the gray reference in wire attached to TB-7.
93 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING SLOPE ST A TUS OF SENSOR 54-60 mV/unit pH Sensor is in good condition. 48-50 mV/unit pH Sensor is nearing the end of its life. Once the slope drops below 48 mV/unit pH, the sensor can no longer be calibrated.
94 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING T roubleshooting Flowchart - rEF F AIL A. Be sure the sensor is completely immersed in the process liquid. If the diagnostic message disappears, the sensor is in good condition. If the diagnostic message remains, go to step B.
95 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.4.4 rEFW Arn rEF W Arn is an electrode fault message. It means the reference electrode impedance exceeds the programmed Reference W arning High ( RW H ) limit.
96 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING B. Check the sensor . Refer to the wiring diagrams in Section 3.0 to identify the RTD leads. Disconnect the R TD leads and measure the resistances shown in Figure 12-3. The measured resistance should agree with the value in T able 12-1 to within about 1%.
97 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING C. Check the transmitter . 1. Disconnect the RTD sensor leads and wire the circuit shown in Figure 12-4. Set the resist ance to the value for 25°C shown in T able 12-1. The measured temperature should equal 25°C to within ±1°C.
98 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING B. The RTD sense or the R TD return wire inside the sensor cable may be broken. Keep the sensor wires attached and jumper TB-3 and TB-4. If the diagnostic message disappears, either the RTD return or R TD sense wire is broken.
99 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.4.9 SLOPE Err LO SLOPE Err LO means that a two-point buffer calibration attempt has failed. The slope is too low (<40 mV/pH) for a good measurement. T roubleshooting Flowchart-SLOPE Err LO A. Repeat the calibration.
100 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.4.10 SLOPE Err HI SLOPE Err HI means that a two-point buffer calibration attempt has failed. The slope is too high (>62 mV/pH) for a good measurement. T roubleshooting Flowchart-SLOPE Err HI A. Repeat the calibration.
101 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING If the sensor is not rebuildable... T ry the reference electrode rejuvenation procedure described in Section 1 1.3.4. If the rejuvenated sensor can be calibrated, the problem has been corrected. If the sensor cannot be calibrated, replace the sensor .
102 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING SYMPTOM SECTION Id 000 appears in display when trying to program or calibrate transmitter 12.5.1 Error message flashing in display 12.4 T ransmitter does not respond to remote controller 12.5.2 Calibration Problems: SLOPE Err HI or SLOPE Err LO appears after calibration attempt 12.
103 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.5.1 Id 000 in Display A security code has been programmed into the transmitter . The correct code must be entered before the transmitter can be programmed or calibrated. T o retrieve a lost security code see Section 5.
104 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.5.5 pH Reading in Buffer Drif ts During Manual Calibration A. Allow adequate time for the temperature of the sensor to reach the temperature of the buffer .
105 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.5.8 Buffer Calibration Is Acceptable; Process pH is Grossly Different from Expected V alue. The symptoms suggest a ground loop (measurement system.
106 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 4. Occasionally , noise can travel into the transmitter housing from the met al it is mounted on. The noise is then radiated into the transmitter electronics.
107 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12. 5.13 T ransmitter Locks Up A. T urn the dc power off, then turn it back on. B. If the problem persists, replace the electronic board stack (PN 23574-02). 12. 5.14 T ransmitter Periodically Rest arts It self A.
108 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING Process pH measurement problem Perform buffer calibration (outside of the process) Refer to Section 7.5 Buffer calibration OK? Preamplifier and trans- mitter test OK? Put transmitter preamplifier switch (S-1) into junction box position.
109 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING Perform buffer calibration (outside of the process) Refer to Section 7.5 Buffer calibration OK? Analyzer test OK? Place transmitter internal pream- plifier switch in proper position. Likely Process Problem or Ground Loop: 1.
MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.7 DISPLA YING DIAGNOSTIC V ARIABLES 12.7.1 Purpose This section describes how to display the diagnostic variables listed below: DIAGNOSTIC MEASUREMENTS DIAGNOSTIC MESSAGES 1. Sensor voltage in mV ( InPut ) 1.
111 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 12.8.2 pH Simulation When the Preamplifier Is Located in the T ransmitter . 1. V erify that switch S-1 is set to "transmitter". See Section 2.2. 2. T urn off sensor diagnostics. See Section 8.
112 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING FIGURE 12-9. Simulate pH through Model 381+ Sensor Preamplifier DWG . NO. REV . 40381+05 A 12.8.4 pH Simulation with the Model 381+ Sensor 1. V erify that switch S-1 is set to "sensor or junction box".
113 MODEL 3081 pH/ORP SECTION 12.0 TROUBLESHOOTING 10. T ouch the other end of the second jumper to the center pin of the BNC connector on the preamplifier . DO NOT LET THE WIRE T OUCH THE OUTSIDE OF THE BNC CONNECT OR. 1 1. Measure the voltage. The measured voltage should be 0 mV , and the pH should be approximately 7.
114 SECTION 13.0 pH MEASUREMENTS 13.1 General 13.2 Measuring Electrode 13.3 Reference Electrode 13.4 Liquid Junction Potential 13.5 Converting V oltage to pH 13.6 Glass Electrode Slope 13.7 Buffers and Calibration 13.8 Isopotential pH 13.9 Junction Potential Mismatch 13.
115 MODEL 3081 pH/ORP SECTION 13.0 pH MEASUREMENTS designed, the liquid junction potential is usually small and rel- atively constant. All three potentials depend on temperature. As discussed in Sections 13.5 and 13.6, the factor relating the cell voltage to pH is also a function of temperature.
116 MODEL 3081 pH/ORP SECTION 13.0 pH MEASUREMENTS FIGURE 13-5. The Origin of Liquid Junction Potentials. The figure shows a thin section through a pore in the junction plug. The junction separates a solution of pot assium chloride on the left from a solution of hydrochloric acid on the right.
117 MODEL 3081 pH/ORP SECTION 13.0 pH MEASUREMENTS 4. the liquid junction potential. The second term, 0.1984 T pH, is the potential (in mV) at the outside surface of the pH glass.
118 MODEL 3081 pH/ORP SECTION 13.0 pH MEASUREMENTS pH 7.00 and pH 10.00. When the electrodes are placed in pH 7 buffer the cell volt age is V 7 , and when the electrodes are placed in pH 10 buffer , the cell volt age is V 10 . Note that V 7 is not 0 mV as would be expected in an ideal sensor , but is slightly different.
119 MODEL 3081 pH/ORP SECTION 13.0 pH MEASUREMENTS junction potentials in the buffers are assumed to be equal and are exaggerated for clarity . If the liquid junction potential in the sample differs from the buffers, a measurement error result s. Figure 13-8 illus- trates how the error comes about.
120 MODEL 3081 pH/ORP SECTION 14.0 ORP MEASUREMENTS SECTION 14.0 ORP MEASUREMENTS 14.1 General 14.2 Measuring Electrode 14.3 Reference Electrode 14.4 Liquid Junction Potential 14.5 Relating Cell V oltage to ORP 14.6 ORP , Concentration, and pH 14.7 Interpreting ORP Measurement s 14.
121 14.2 MEASURING ELECTRODE Figure 14-2 shows a typical ORP measuring elec- trode. The electrode consist s of a band or disc of platinum attached to the base of a sealed glass tube. A platinum wire welded to the band connects it to the lead wire. For a noble metal electrode to develop a st able potential, a redox couple must be present.
122 Liquid junction potentials exist whenever dissimilar elec- trolyte solutions come into contact. The magnitude of the potential depends on the difference between the mobility of the ions. Although liquid junction potentials cannot be elim- inated, they can be made small and relatively constant.
123 is described by the following equation, called the Nernst equation: E = E° - (2) In the Nernst equation, E is the electrode potential and E° is the standard electrode potential, both in millivol.
124 The Nernst equation for reaction 3 is: E = E°- (4) Note that the hydrogen ion factor in the concentration ratio is raised to the fourteenth power . The t able shows the expected effect of changing pH on the measured ORP at 25°C. pH changes ORP changes by from 2.
125 MODEL 3081 pH/ORP SECTION 14.0 ORP MEASUREMENTS The ORP of the iron (II) - iron (III) standard when meas- ured with a platinum electrode against a saturated sil- ver-silver chloride reference is 476 ± 20 mV at 25°C.
126 15.1 OVERVIEW OF HART COMMUNICA TION HART (highway addressable remote transducer) is a digital communication system in which two frequencies are superimposed on the 4 to 20 mA output signal from the transmitter . A 1200 Hz sine wave represents the digit 1, and a 2400 Hz sine wave represents the digit 0.
127 MODEL 3081 pH/ORP SECTION 15.0 THEOR Y - REMOTE COMMUNICA TIONS If your communicator does not recognize the Model 3081 pH/ORP transmitter , the device description library may need updating. Call the manufacturer of your HART communication device for updates.
128 TERM DEFINITION Acid When dissolved in water acids increase the hydrogen ion concentration. Pure water at 25°C contains 1 x 10 -7 moles per liter of hydrogen ions (H + ) and an equal concentration of hydroxide ions (OH - ).
129 MODEL 3081 pH/ORP SECTION 16.0 GLOSSAR Y Combination electrode In a combination electrode, the measuring electrode and reference electrode are combined in a single body . Of ten the temperature element is included in the body as well. Common A point in a circuit against which voltages are measured.
130 MODEL 3081 pH/ORP SECTION 16.0 GLOSSAR Y Fault A fault is a system disabling condition. Measurement data displayed during a fault condition are probably in error and should be regarded with great suspicion. The Model 3081 pH/ORP transmitter displays fault messages to aid in trouble-shooting.
131 MODEL 3081 pH/ORP SECTION 16.0 GLOSSAR Y Hold function During hold the Model 3081 pH/ORP transmitter generates a pre-programmed output current or remains at the last value. Placing the transmitter in hold avoids false alarms and unwanted operation of chemical dosing pumps while the sensor is being calibrated or cleaned.
132 MODEL 3081 pH/ORP SECTION 16.0 GLOSSAR Y pH Although pH is often defined as the negative common logarithm of the hydrogen ion activity , pH is best defined by describing how it is measured. The measure- ment requires three steps. Assemble an electrochemical cell consisting of a glass electrode, a reference electrode, and the solution.
133 MODEL 3081 pH/ORP SECTION 16.0 GLOSSAR Y Solution ground A solution ground is a metal post or ring incorporated into the sensor body and mak- ing contact with the process stream. Glass and reference impedances are meas- ured by applying a voltage pulse between the electrode and the solution ground.
134 MODEL 3081 pH/ORP SECTION 17.0 RETURN OF MA TERIAL SECTION 17.0 RETURN OF MA TERIAL 17.1 GENERAL. T o expedite the repair and return of instrument s, proper communication between the customer and the factory is important. Call 1-949-757-8500 for a Return Materials Authorization (RMA) number .
W ARRANTY Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defect s in workman- ship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller . Consumables, pH electrodes, membranes, liquid junctions, electrolyte, O-rings, etc.
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Emerson Process Management Rosemount Analytical Inc. 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Emerson Process Management 3081 pH/ORP è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Emerson Process Management 3081 pH/ORP - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Emerson Process Management 3081 pH/ORP imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Emerson Process Management 3081 pH/ORP ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Emerson Process Management 3081 pH/ORP, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Emerson Process Management 3081 pH/ORP.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Emerson Process Management 3081 pH/ORP. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Emerson Process Management 3081 pH/ORP insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.