Instruction Manual 226 LIQ_MAN_ABR_226 April 2014 Submer sion/Inser tion T oroidal Sensor For add itional in form ation, ple as e visit our we bsite a t www .ros emoun tanalytical.com CAUTION SENSOR/PROCESS AP PLICA TION C OMP A TIBILIT Y The wetted sensor mat erials may no t be compatible with process composition and oper ating conditions.
2 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Figure 2. Inser tion Installation Showing Mounting Hardware Wiring Kee p sensor wiring aw ay fr om ac c onductors and high curren t de mand ing equipme nt. Do not cut cable . Cu tting the cab le ma y vo id the w arranty .
3 Instruction Manual 226 LIQ_MAN_ABR_226 April 2014 Figure 4. Wiring for 226-54 and 226-56 sensors to 1056 and 56 analyzers Figure 5. Wiring for 226-54 and 226-56 sensors to 54eC analyzer.
4 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Figure 6. Wiring for 226-54 and 226-56 sensors to Xmt - T panel m ount transmitt er Figure 7. Wiring for 226-54 and 226-56 sensors to Xmt - T pi pe/.
5 Instruction Manual 226 LIQ_MAN_ABR_226 April 2014 Figure 8. Wiring for 226-54 and 226-56 sensors to 1066 transmitter RCV B RCV A RSHLD DRV B 226-54 RT N SENSE RT D IN SHLD CLEAR WHITE GREEN BLACK GREEN BLACK WHITE DRV A DSHLD WHEN INCH AND METRIC DIMS ARE GIVEN INCH MILLIMETER WIRING MODE L 226-54 SENSOR TO MODE L 1066 Rosemount Analytical, Inc.
6 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Figure 10. Wiring sensors through a r emot e junction box Wire sensors point to point. For wiring at the analyzer end, refer to the appropriate analyzer wiring diagram. F or interconnect - ing cable 23294-00, use the 226-54 wiring diagram.
7 Instruction Manual 226 LIQ_MAN_ABR_226 April 2014 Calibration The nominal ce ll c onstant of the 226 se nsor is 1.2/cm. The error in the c e ll c onstant is about ±10%, so c onductivit y read ings made using the nomin al c ell constant will h av e an error o f a t le ast ±10%.
8 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 T roubleshooting Problem Probable Cause Solution Off-scale re ading Wiring is wrong. Ve rify and correc t wiring. RTD is ope n or shor ted. Check the RTD f or open or short circuits See Figure 13. Sensor is no t in proce ss stream .
9 Instruction Manual 226 LIQ_MAN_ABR_226 April 2014 Figure 13. Resistance check. Disc onn ect leads from analyzer before measuring resistances. gree n black white black green - in clear - common white.
10 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Notes.
11 Instruction Manual 226 LIQ_MAN_ABR_226 April 2014 Notes.
226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Re v. E Credit Cards for U .S . Purch ases Only . 8 Emerson Process Management 2400 Barranca Park wa y Ir vine , C A 92606 USA T el : (949) 757-8500 F ax: (949) 474-7250 rosemo untanalytical.com © Ro semoun t Analytical Inc.
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Emerson LIQ_MAN_ABR_226 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Emerson LIQ_MAN_ABR_226 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Emerson LIQ_MAN_ABR_226 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Emerson LIQ_MAN_ABR_226 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Emerson LIQ_MAN_ABR_226, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Emerson LIQ_MAN_ABR_226.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Emerson LIQ_MAN_ABR_226. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Emerson LIQ_MAN_ABR_226 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.