pH/ORP Sensors Instruction Manual PN 51-396/397/398/rev .L November 2010 MODEL 396 and MODEL 398 MODEL 396VP and MODEL 398VP MODEL 397 Models 396/396VP/397/398/398VP.
CAUTION SENSOR/PROCESS APPLICA TION COMP A TIBILITY The wetted sensor materials may not be com- p atible with process com position and operat- ing conditions. Application compat ibility is entirely the responsibility of the user . DANGER HAZARDOUS AREA INST ALLA TION This sensor is not Intrinsically Safe.
i MODEL 396/396VP/397/398/398VP T ABLE OF CONTENTS TUpH MODELS 396/396VP/397/398/398VP pH/ORP SENSORS T ABLE OF CONTENTS Section Title Page 1.0 DESCRIPTION AND SPECIFICA TIONS ......................................................... 1 1.1 TUpH™ Features and Applications .
MODEL 396/396VP/397/398/398VP T ABLE OF CONTENTS LIST OF FIGURES Number Title Page 1-1 Cross Section Diagram of the TUpH Reference T echnology ............. 2 2-1 Flow Through/Insertion Installation for Models 396, 396VP , 398, & 398VP ...........
SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.1 TUpH™ FEA TURES AND APPLICA TIONS. Rosemount Analytical has achieved a new industry standard for the life expect ancy of pH sensors with the TUpH Sensors.
2 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTIONS AND SPECIFICA TIONS FIGURE 1-1. Cross Section Diagram of the TUpH Reference T echnology All TUpH sensors are designed with a large area reference junction, helical reference pathway , and an AccuGlass pH glass bulb.
3 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.2 PERFORMANCE & PHYSICAL SPECIFICA TIONS SPECIFICA TIONS MODELS 396 &396VP MODELS 398 & 398VP MODEL 397 QUIK-L.
4 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1.3 ORDERING INFORMA TION MODEL 396 TUpH INSERTION/SUBMERSION ST AINLESS STEEL pH SENSOR CODE ANAL YZER/TC COMP A TIBILITY (.
MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS The Model 398 Sensor , housed in a titanium tube and made with a T efzel reference junction, can be used with a 1 inch MNPT process connector (purchased sep arately).
6 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS The TUpH Model 397 pH Sensor is housed in a polypropylene body and is designed to be used with the Quik-Loc Kit. The sensor includes a large general purpose (GPHT) glass pH electrode and a large area polypropylene reference junction with gel filled reference electrolyte.
7 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS 1. Process Connector Accessories (required for all first time installations with 1-inch process connection threads) Choose one: PN 23166-00, 316 SST , 1 in. x 1 in. NPT process connector , with EPDM o-ring PN 23166-01, T itanium, 1 in.
8 MODEL 396/396VP/397/398/398VP SECTION 1.0 DESCRIPTION AND SPECIFICA TIONS ADDITIONAL ACCESSORIES FOR MODEL 396, 396VP , 398 AND 398VP TUpH SENSORS P ART NUMBER DESCRIPTION 33046-00 Ferrule 1 in., split 316 SS 3321 1-00 Adapter retrofit for PN 915240-04 9550167 EPDM O-ring for process connector (PN 23166-00) 9310100 Ferrule, 1 in.
9 2.2.1 Flow Through and Insertion Mounting for Models 396, 396VP , 398, and 398VP . The Models 396, 396VP , 398, and 398VP Sensors can be used with a 1 inch MNPT process connector at the front of the sensor for mounting into a 1-1/2 inch tee or the process.
10 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION Wrap the pipe threads of the T win-Kam Kamloc coupler with T eflon tape before placing it into the process pipe. The coupler can be connected to any 1 in. process connection and must be mounted within 80° of vertical, with the elec- trode facing down.
11 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-2. Dimensional Drawing For Models 396 and 398 DWG . NO. REV . 40039601 J MILLIMETER INCH.
12 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION INCH MILLIMETER Metal Process Connector PN 23166-xx (xx = 00 for 316 SST and xx = 01 for tit a- nium) can be used for insertion or sub- mersion mounting of Models 396VP or 398VP sensors in 1-inch fittings.
13 DWG . NO. REV . 40039602 D MILLIMETER INCH DWG . NO. REV . 40039603 C DWG . NO. REV . 40039604 D MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION P .N. 23166-00, 23166-01, OR 9510066 FIGURE 2-4. Submersion Inst allations for Models 396, 396VP , 398, and 398VP .
14 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-5. Dimensional Drawings For Model 397 in the Quik-Loc Unit DWG . NO. REV . 40039701 C MILLIMETER INCH DWG . NO. REV . 40039704 B NOTE: The EP gasket (see drawing #40039702) provided with the coupler should be periodically inspected.
15 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION 2.3 ELECTRICAL INST ALLA TION. The Models 396, 396VP , 397, 398, and 398VP are for use with a remote preamplifier . Each sensor comes with either a special 15 ft low noise coax cable or a V ariopol (VP) con- nector , which is used with a mating V ariopol cable.
16 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION DWG . NO. REV . 40039610 D FIGURE 2-9. Wiring Det ails For Model 396-54 (50), 397-54 (50), and 398-54 (50). For Use With Junction Box (PN 23309-04 (03)) and Remote Preamplifier DWG . NO. REV .
17 DWG . NO. REV . 40038920 G MILLIMETER INCH MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-10. Wiring Det ails for Model 396-54, 397-54, and 398-54 (Pt-100 RTD).
18 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-1 1. Wiring Details For Models 396, 397, and 398. For Use W ith Model 1 181 pH. DWG .
19 FIGURE 2-12. Wiring Det ails For Models 396-54, 397-54, and 398-54. For Use With Models 1054ApH-54, 1054BpH-54, and 2054pH-54 DWG . NO. REV . 40039606 D MILLIMETER INCH MODEL 396/396VP/397/398/398VP SECTION 2.
20 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-13. Wiring Det ails For Models 396-54, 397-54, and 398-54. For Use With Models 2081 pH-05 and 1054 pH-54 DWG .
21 DWG . NO. REV . 40039605 F MILLIMETER INCH MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-14. Wiring Det ails For Models 396-54 , 397-54, and 398-54.
22 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-15. Wiring Det ails For Models 396-54 , 397-54, and 398-54. For Use With SoluComp Model SCL-P-014.
23 FIGURE 2-16. Wiring Models 396-54, 397-54, and 398-54 to Model 2700 Preamplifier (PN 23054-03). MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-17. Wiring Models 396-( )-54-62-(71), 397-02-10-54-62, and 398-( )-54-62 to Model 1055-01-10-22-32 FIGURE 2-18.
24 See below for wire functions for Models 396VP and 398VP . The V ariopol connector cable part numbers are PN 23645- 06 (with BNC) and PN 23645-07. FIGURE 2-19. Wire Functions and Pin Connections for V ariopol connector cable PN 23645-06 (with BNC) PN 23645-07 FIGURE 2-21.
FIGURE 2-24. Wiring Model 396VP and 398VP to Model 2081 FIGURE 2-25. Wiring Model 396VP and 398VP to Model 3081 and 4081 FIGURE 2-27. Wiring Model 396VP and 398VP to Model 3081, 4081, and 5081 through a Remote Junction Box FIGURE 2-26. Wiring Model 396VP and 398VP to Model 2081 through a Remote Junction Box FIGURE 2-23.
26 FIGURE 2-28. Wiring Model 396VP and 398VP to Model 1054 FIGURE 2-30. Wiring Model 396VP and 398VP to Model 1054 through a Remote Junction Box FIGURE 2-29.
FIGURE 2-33. Wiring Model 396VP/398VP to Models 54/54e FIGURE 2-32. Wiring Model 396VP and 398VP to Models 54 and 54e through a Remote Junction Box FIGURE 2-35. Wiring Model 396VP and 398VP to Model 2700 FIGURE 2-34. Wiring Model 396VP and 398VP to Model SCL-(P/Q) MODEL 396/396VP/397/398/398VP SECTION 2.
28 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-36. Wiring Model 396VP-54 to Model 1055-22-32.
MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION 29 FIGURE 2-37. Wiring Models 396-54-6, 397-54-62, & 398-54-62 to Model Xmt-P-XX-10 FIGURE 2-38.
30 MODEL 396/396VP/397/398/398VP SECTION 2.0 INST ALLA TION FIGURE 2-39. Wiring Models 396/397/398 to Model 5081-P FIGURE 2-40. Wiring Models 396-54-62/398-54-62 to Model 1055-01-10-22-32 Model 398-54.
MODEL 396/396VP/397/398/398VP SECTION 3.0 ST ART UP AND CALIBRA TION SECTION 3.0 ST ART UP AND CALIBRA TION 3.1 SENSOR PREP ARA TION. Shake down the sensor to remove any air bubbles that may be present at the tip of the pH glass bulb. In most cases, the pH sensor can simply be installed as shipped and readings with an accuracy of ±0.
32 MODEL 396/396VP/397/398/398VP SECTION 4.0 MAINTENANCE 4.0 MAINTENANCE. The Model 396, 396VP , 397, 398, and 398VP Sensors are disposable type sensors and therefore require minimum maintenance. Every sensor should be kept clean and free of debris and sediment at all times.
33 MODEL 396/396VP/397/398/398VP SECTION 5.0 TROUBLESHOOTING T rouble Probable Cause Remedy Meter reads off scale. (Display T .C. element shorted. Check T .C. element as instructed reads overrange.) in Section 4.2 and replace sensor if defective. Sensor not in process or sample Make sure sensor is in process with stream is low .
34 MODEL 396/396VP/397/398/398VP SECTION 6.0 RETURN OF MA TERIAL SECTION 6.0 RETURN OF MA TERIAL 6.1 GENERAL. T o expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for rep air , call 1-949-757-8500 for a Return Materials Authorization (RMA) number .
W ARRANTY Seller warrants that the firmware will execute the programming instructions provided by Seller , and that the Goods manufactured or Services provided by Seller will be free from defects in materials or workmanship under normal use and care until the expira- tion of the applicable warranty period.
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. 8 Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Emerson 396 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Emerson 396 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Emerson 396 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Emerson 396 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Emerson 396, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Emerson 396.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Emerson 396. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Emerson 396 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.