Models 140, 141, and 142 Conductivity Sensors Instruction Sheet PN 51A-140/rev .E December 2010 SENSOR SPECIFICA TIONS For additional information, please visit our website at www .
Keep 1/4 inch (6 mm) clearance between electrodes and piping. The electrodes must be completely sub- merged in the process liquid, i.e., to the level of the threaded connection. See Figure 1 for recommended orientation. If the sensor is installed in a side stream with the sample draining to open atmosphere, bubbles may accumulate on the electrodes.
3 MODEL 140,141, and 142 sensors INST ALLA TION FIGURE 3. Inst alling model 141 sensor DWG . NO. REV . 40014204 B INST ALLA TION – 141 SENSOR 1. Install the sensor in a 3/4-inch NPT weldalet or in a 1-inch pipe tee. 2. Remove the plastic shipping cap from the sensor .
RTD RTD in RTD sense RTD return MODEL 140,141, and 142 sensors WIRING WIRING All 140 series sensors have a junction box mounted on the back of the sensor . Wiring connections in the junction box are shown in Figure 5. Figure 5. Sensor junction box wiring.
5 MODEL 140,141, and 142 sensors WIRING FIGURE 8. Model Xmt-C-10 Wiring (Panel) FIGURE 9. Model Xmt-C-1 1 Wiring (Pipe or Wall) RCV B RCV A RSHLD DRVB RTN SENSE RTD IN SHLD CLEAR WHITE/RED RED DRV A DSHLD WHITE CLEAR GRA Y ORANGE CLEAR TB1 TB2 FIGURE 10.
RETRACTING AND INSERTING THE MODEL 140 SENSOR RETRACTING THE SENSOR 1. Push in on the sensor junction box and slowly loosen the sensor compression fitting nut by reversing the sensor tightening procedure illustrated in Figure 2.
MODEL 140,141, and 142 sensors WIRING 7 3. Slowly open the valve. 4. Insert the sensor up to the desired insertion depth and turn the sensor compression fitting nut until it is finger tight. 5. Position the entire sensor for easy access to the ball valve handle, sensor compression fitting nut and J-box terminal block.
MODEL 140,141, and 142 sensors CALIBRA TION 8 CALIBRA TION Model 140 sensors are not calibrated at the factory . The cell const ant on the label is a nominal value only .
9 MODEL 140,141, and 142 sensors MAINTENANCE MAINTENANCE CLEANING THE SENSOR Use a warm detergent solution and a soft brush or pipe cleaner to remove oil and scale. Isopropyl alcohol can also be used to remove oily films. Avoid using strong mineral acids to clean conductivity sensors.
10 MODEL 140,141, and 142 sensors ACCESSORIES PN Description 23724-00 Ball valve kit for Model 140 sensor 23730-00 Process compression fitting kit (3/4-inch NPT) for Model 140 sensor 23731-00 Process .
11 TROUBLESHOOTING PROBLEM PROBABLE CAUSE SOLUTION Off-scale reading Wiring is wrong. V erify wiring. T emperature element is open or shorted. Check temperature element for open or short circuits. See Figure 14. Sensor is not in the process stream. Be sure the sensor is completely submerged in the process stream.
Credit Cards for U.S. Purchases Only . The right people, the right answers, right now . ON-LINE ORDERING NOW A V AILABLE ON OUR WEB SITE http://www .raihome.com Specifications subject to change without notice. 8 Emerson Process Management 2400 Barranca Parkway Irvine, CA 92606 USA T el: (949) 757-8500 Fax: (949) 474-7250 http://www .
Un punto importante, dopo l’acquisto del dispositivo (o anche prima di acquisto) è quello di leggere il manuale. Dobbiamo farlo per diversi motivi semplici:
Se non hai ancora comprato il Emerson 140 è un buon momento per familiarizzare con i dati di base del prodotto. Prime consultare le pagine iniziali del manuale d’uso, che si trova al di sopra. Dovresti trovare lì i dati tecnici più importanti del Emerson 140 - in questo modo è possibile verificare se l’apparecchio soddisfa le tue esigenze. Esplorando le pagine segenti del manuali d’uso Emerson 140 imparerai tutte le caratteristiche del prodotto e le informazioni sul suo funzionamento. Le informazioni sul Emerson 140 ti aiuteranno sicuramente a prendere una decisione relativa all’acquisto.
In una situazione in cui hai già il Emerson 140, ma non hai ancora letto il manuale d’uso, dovresti farlo per le ragioni sopra descritte. Saprai quindi se hai correttamente usato le funzioni disponibili, e se hai commesso errori che possono ridurre la durata di vita del Emerson 140.
Tuttavia, uno dei ruoli più importanti per l’utente svolti dal manuale d’uso è quello di aiutare a risolvere i problemi con il Emerson 140. Quasi sempre, ci troverai Troubleshooting, cioè i guasti più frequenti e malfunzionamenti del dispositivo Emerson 140 insieme con le istruzioni su come risolverli. Anche se non si riesci a risolvere il problema, il manuale d’uso ti mostrerà il percorso di ulteriori procedimenti – il contatto con il centro servizio clienti o il servizio più vicino.